1. Page 1
    2. Page 2
    3. Page 3
    4. Page 4
    5. Page 5
    6. Page 6
    7. Page 7
    8. Page 8
      1. Page 1
      2. Page 2
      3. Page 3
      4. Page 4
      5. Page 5
      6. Page 6
      7. Page 7
      8. Page 8
      9. Page 9
      10. Page 10
      11. Page 11
      12. Page 12
      13. Page 13
      14. Page 14
      15. Page 15
      16. Page 16
      17. Page 17
      18. Page 18
      19. Page 19
      20. Page 20
      21. Page 21
      22. Page 22
      23. Page 23
      24. Page 24
      25. Page 25
      26. Page 26
      27. Page 27
      28. Page 28
      29. Page 29
      30. Page 30
      31. Page 31
      32. Page 32
      33. Page 33
      34. Page 34

 
Here's Waters 6508 method that was approved for drinking water. Aren't
you limited to methods that are listed as approved for drinking water?
Based on the info at the top of the method, I'm guessing this may now be
an ASTM method. It was evaluated under the ATP program, so EPA was
given the method prior to the ASTM process. I don't know this for sure
and it will be something I investigate as we begin putting together our
next Expedited Methods Approval FR action. (If it is the same method,
we'll probably list it in Appendix A.)
(See attached file: Waters Method D 6508, Rev
2_EPA-HQ-OW-2003-0070-0063.pdf)
As for the other method, it was approved prior to the 2007 methods rule.
I don't have a copy of it, because I wasn't involved in the earlier
methods rules. However, I have asked our ATP coordinator to see if it
is in the ATP file. When I hear back from him, I'll let you know.
Hope this helps,
Pat
"Mike
McCambridge"
<mccambridge@ipc?
To
b.state.il.us>?
Pat Fair/Cl/USEPA/US@EPA
CC
07/08/2008 03:29
PM
?
?
Subject
Re: Waters Methods
Whatever you can do for me when you get back. I have continued to look
into this today. I am convinced that Method 6500 added to SW-846 in
Update IV in the end of 2007 is the method you have called "Method
D6508" from Waters. See 73 Fed. Reg. 486 (Jan. 3, 2008);
http://www.epa.gov/SW-846/pdfs/6500.pdf. If this is true, I will likely
cite the SW-846 version of the method, since it is much easier to obtain
than the method from Waters. As described, Waters initially acted like
I spoke a foreign language when I asked for "Method D6508." As for
From:
To:
Date:
Subject:
Mike,
<Fair.Pat@epamail.epa.gov>
MCCAMBM@ipcb.state.il.us
7/10/2008 7:46:26 AM
Re: Waters Methods
RECEIVED
CLERK'S OFFICE
AUG 0
2008
STATE
OF ILLINOIS
Pollution Control Board

 
Method B-1011, it seems to distill down to me needing the title to the
document in which the method appears.
Talk to you when you return.
Michael J. McCambridge
Attorney
Illinois Pollution Control Board
312-814-6924
>>> <Fair.Pat@epannail.epa.gov> 7/8/2008 2:18 PM >>>
Mike,
I'm working off site today, so I don't have access to the references I
need to answer your questions. I should have copies of the methods that
were added to 40 CFR 141 as part of the 2007 methods update rule. If
these Waters methods are prior to that, I might not be able to help you.
Unfortunately, I don't know who might have
.
them other than Waters.
I'll see what I can find tomorrow and get back to you.
Pat
"Mike McCambridge" <mccambridge@ipcb.state.il.us> wrote:
To: Pat Fair/CUUSEPA/US@EPA
From: "Mike McCambridge" <mccambridge@ipcb.state.il.us>
Date: 07/08/2008 01:57PM
Subject: Waters Methods
I have tried to obtain copies of the two Waters methods referenced in 40
C.F.R. 141.23(k)(1) for fluoride and nitrite/nitrate using the contact
information included in the rule. At first, the Waters rep could not
locate anything based on the EPA descriptions included in the rule.
This morning I received two documents that purport to be the methods.
The documents raise questions that you might answer for me.
The copy of Method B-1011 sent me by Waters is nearly identical to one
that I found on the USEPA website. The only difference between the two
is that the method from the USEPA website is headed "Waters." The
document it appears to include pages 13 through 17 from some reference.
It is undated, which means that I cannot use it for an incorporation by
reference. Do you have a dated copy of Method B-1011 or a fuller copy
of the posted reference that would include the date? It appears that
the method is just one cited out of a fuller reference, and I should
cite to that fuller reference by its own title. I will also approach
Waters with this request.
Your rule cites "Waters Method D6508, Rev. 2," entitled "Test Method for
Determination of Dissolved Inorganic Anions in Aqueous Matrices Using
Capillary Ion Electrophoresis and Chromate Electrolyte." Waters sent me
a document marked "Method 6500," "revision 0," and dated February 2007,"
and entitled "Dissolved Inorganic Anions In Aqueous Matrices By
Capillary Ion Electrophoresis." That document appears to be Method 6500
from SW-846. Is "Waters Method D6508, Rev. 2" the same as Method 6500,

 
rev. 0 from SW-846? If so, why did USEPA cite this as "D6508"? If not,
can you forward me a copy of Method D6508 or give me enough information
to identify the method to Waters, that I might obtain a copy of the
right method?
Michael J. McCambridge
Attorney
Illinois Pollution Control Board
312-814-6924

 
Page 1 of 5
Mike McCambridge - Re: Waters Methods
From:
<Fair.Pat epamail.epa.gov>
To: "Mike
McCambridge" <mccambridge@ipcb.state.il.us>
Date:
7/10/2008 9:31 PM
Subject:
Re: Waters Methods
Mike,
I haven't done a one-to-one check of the ASTM method against the Waters method, so I can't say
for sure that they are the same. My comment was meant to let you know that I would do that
BEFORE we issue the next set of method approvals. If they are the same or only have insignificant
differences, then we will include the ASTM method as an approved method. Legally, it won't be an
approved drinking water method until we publish a notice in the Federal Register.
It's my opinion that if the Waters methods aren't easily available from Waters, then you can easily
justify not including them in your state regulations. Our ATP coordinator wasn't able to find a copy
of the nitrate/nitrite method in his files. However, he is still checking on it.
I have the GA Tech method. I can email it to you on Monday. If you need it before then, you can
go to the e-docket for the 2007 Methods Update Rule. I know the method is in the docket,
because I put it there and it is available for download through the docket site.
I will see if I can find out how we should be referencing the GA Tech method. I thought our
information was correct when we went final on the rule.
Hope this helps.
Pat
"Mike McCambridge" <mccambridge@ipcb.state.il.us> wrote:
To: Pat Fair/Cl/USEPA/US@EPA
From: "Mike McCambridge" <mccambridge@ipcb.state.il.us>
Date: 07/10/2008 06:07PM
Subject: Re: Waters Methods
Thank you. That nails it down. I will cite it as an ASTM method.
I have another method problem. I have been trying
:
to obtain a copy
of that Ra-226/Ra-228 method by gamma-ray spectometry developed by
Georgia Insitute of Technology. The "Environmental Resources
Center" has been disbanned or something, so that the number at 40
C.F.R. 141.74 is no longer valid. It may have become the
Environmental Radiation Center or something. I have placed several
calls and e-mails with Bernd Kahn and the Center in an attmpt to
locate the method, but no luck so far.
Can you help on this one too?
file://C:\Documents and Settings\McCambM\Local Settings\Temp\GW100001.HTM 8/4/2008

 
Page 2 of 5
Michael J. McCambridge
Attorney
Illinois Pollution Control Board
312-814-6924
>>> <Fair.Pat@epamail.epa.gov> 7/10/2008 7:43 AM >>>
Mike,
Here's Waters 6508 method that was approved for drinking water.
Aren't
you limited to methods that are listed as approved for drinking
water?
Based on the info at the top of the method, I'm guessing this may
now be
an ASTM method. It was evaluated under the ATP program, so EPA was
given the method prior to the ASTM process. I don't know this for
sure
and it will be something I investigate as we begin putting together
our
next Expedited Methods Approval FR action. (If it is the same
method,
we'll probably list it in Appendix A.)
(See attached file: Waters Method D 6508, Rev
2 EPA-HQ-0W-2003-0070-0063.pdf)
As for the other method, it was approved prior to the 2007 methods
rule
I don't have a copy of it, because I wasn't involved in the earlier
methods rules. However, I have asked our ATP coordinator to see if
it
is in the ATP file. When I hear back from him, I'll let you know.
Hope this helps,
Pat
To
CC
"Mike
McCambridge"
<mccambridge@ipc
b.state.il.us>
Pat Fair/Cl/USEPA/US@EPA
file://C:\Documents and Settings\MeCambM\Local Settings\Temp\GW100001.HTM
8/4/2008

 
Page 3 of 5
07/08/2008 03:29
PM
Subject
Re: Waters Methods
Whatever you can do for me when you get back. I have continued to
look
into this today. I am convinced that Method 6500 added to SW-846 in
Update IV in the end of 2007 is the method you have called "Method
D6508" from Waters. See 73 Fed. Reg. 486 (Jan. 3, 2008);
http://www.epa.gov/SW-846/pdfs/6500.pdf. If this is true, I will
likely
cite the SW-846 version of the method, since it is much easier to
obtain
than the method from Waters. As described, Waters initially acted
like
I spoke a foreign language when I asked for "Method D6508." As for
Method B-1011, it seems to distill down to me needing the title to
the
document in which the method appears.
Talk to you when you return.
Michael J. McCambridge
Attorney
Illinois Pollution Control Board
312-814-6924
>>> <Fair.Pat@epamail.epa.gov> 7/8/2008 2:18 PM >>>
Mike,
I'm working off site today, so I don't have access to the references
I
file://C: \Documents and Settings\McCambM\Local Settings\Temp\GW}00001.HTM 8/4/2008

 
Page 4 of 5
need to answer your questions. I should have copies of the methods
that
were added to 40 CFR 141 as part of the 2007 methods update rule.
If
these Waters methods are prior to that, I might not be able to help
you.
Unfortunately, I don't know who might have them other than Waters.
I'll see what I can find tomorrow and get back to you.
Pat
"Mike McCambridge" <mccambridge@ipcb.state.il.us> wrote:
To: Pat Fair/Cl/USEPA/US@EPA
From: "Mike McCambridge" <mccambridge@ipcb.state.il.us>
Date: 07/08/2008 01:57PM
Subject: Waters Methods
I have tried to obtain copies of the two Waters methods referenced
in 40
C.F.R. 141.23(k)(1) for fluoride and nitrite/nitrate using the
contact
information included in the rule. At first, the Waters rep could
not
locate anything based on the EPA descriptions included in the rule.
This morning I received two documents that purport to be the
methods.
The documents raise questions that you might answer for me.
The copy of Method B-1011 sent me by Waters is nearly identical to
one
that I found on the USEPA website. The only difference between the
two
is that the method from the USEPA website is headed "Waters." The
document it appears to include pages 13 through 17 from some
reference.
It is undated, which means that I cannot use it for an incorporation
by
reference. Do you have a dated copy of Method B-1011 or a fuller
copy
of the posted reference that would include the date? It appears
that
the method is just one cited out of a fuller reference, and I should
cite to that fuller reference by its own title. I will also
approach
Waters with this request.
Your rule cites "Waters Method D6508, Rev. 2," entitled "Test Method
file://C:\Documents and Settings\MeCambM\Local Settings\Temp\GW}00001.HTM 8/4/2008

 
Page 5 of 5
for
Determination of Dissolved Inorganic Anions in Aqueous Matrices
Using
Capillary Ion Electrophoresis and Chromate Electrolyte." Waters
sent me
a document marked "Method 6500," "revision 0," and dated February
2007,"
and entitled "Dissolved Inorganic Anions In Aqueous Matrices By
Capillary Ion Electrophoresis." That document appears to be Method
6500
from SW-846. Is "Waters Method D6508, Rev. 2" the same as Method
6500,
rev. 0 from SW-846? If so, why did USEPA cite this as "D6508"? If
not,
can you forward me a copy of Method D6508 or give me enough
information
to identify the method to Waters, that I might obtain a copy of the
right method?
Michael J. McCambridge
Attorney
Illinois Pollution Control Board
312-814-6924
file://C:\Documents and Settings\McCambM\Local Setfings\Temp\GW100001.HTM
?
8/4/2008

 
00---09,1b--001°--1966,3
This ASTM D6508, Rev2 method document has been reviewed
by EPA
Office of Drinking
Water and Wastewater for EPA Tier 3 approval. Added updated QC criteria based upon
statistical analysis
by Dyncorp.
ATP Case #:
Draft #:
Date:
Method Author:
Telephone:
FAX:
N00-0002 and D00-0002
Second draft with EPA Modifications: ASTM D6508, Rev
2
December 2000
Jim Krol
508/482-2131
508/482-3625
Test Method for
Determination of Dissolved Inorganic Anions
in Aqueous Matrices Using
Capillary Ion Electrophoresis
1 Scope
and Chromate Electrolyte
1.1 This test method covers the determination of the inorganic anions fluoride,
bromide, chloride, nitrite, nitrate, ortho-phosphate, and sulfate in drinking water,
wastewater, and other aqueous matrices using capillary ion electrophoresis
(CIE) with indirect UV detection. See Fig. 1 through 6.
1.2 The test method uses a chromate-based electrolyte and indirect UV detection
at 254 nm. It is applicable for the determination of inorganic anions in the
range of 0.2 to 50 mg/L except for fluoride whose range is 0.2 to 25 mg/L.
1.3 It is the responsibility of the user to ensure the validity of this test method for
other anion concentrations and untested aqueous matrices.
Note 1: The highest accepted anion concentration submitted for P&B extend the anion
concentration range for the following anions; Chloride to 93 mg/L, Sulfate to 90 mg/L,
Nitrate to 72 mg/L, and ortho-phosphate to 58 mg/L.
1.4 This method does not purport to address all of the safety problems, if any,
associated with its use. It is the responsibility of the user of this standard to
establish appropriate safety and health practices and determine the
applicability of regulatory limitations prior to use. For specific hazard
statements, see sec. 9.

 
2
2 Referenced Documents
2.1 ASTM Standards
D 1066 Practice for Sampling Steam'
D
1129 Terminology Relating to Water'
D 1193 Specification for Reagent Water'
D
2777 Practice for Determination of Precision and Bias of Applicable
Methods of Committee D-19 on Water'
D 3370 Practices for Sampling Water'
D
3856 Guide for Good Laboratory Practices in Laboratories Engaged in
Sampling and Analysis of Water'
D
5810 Standard Practice of Spiking Samples'
D
5847 Standard Practice for Writing Quality Control Specifications for
Standard Test Methods for Water Analysis'
D
5905 Standard Specification for Substitute Wastewater'
F 488 Test Method for Total Bacterial Count in Water2
2.2 EPA 40 CFR Ch.1 (7-1-92 Edition), Pt 136, App. B, page 565 — 567: Definition
and Procedure for the Determination of the Method Detection Limit-Revision 1.11.
2.3 Draft Protocol for EPA Approval of New Methods for Organic and Inorganic
Analytes in Wastewater and Drinking Water, dated Mar 1999, EPA-821-B-98-003.
3 Terminology
3.1 Definitions - For definitions of terms used in this test method, refer to Terminology
D1129.
3.2 Description of Terms Specific to This Test Standard:
3.2.1 Capillary Ion Electrophoresis -- an electrophoretic technique in which an UV
absorbing electrolyte is placed in a 50 pm to 75 pm fused silica capillary.
Voltage is applied across the capillary causing electrolyte and anions to
migrate towards the anode and through the capillary's UV detector window.
Anions are separated based upon the their differential rates of migration in the
electrical field. Anion detection and quantitation are based upon the principles
of indirect UV detection.
3.2.2 Electrolyte -- combination of a UV absorbing salt and an electroosmotic flow
modifier placed inside the capillary, used as a carrier for the analytes, and for
detection and quantitation. The UV absorbing portion of the salt must be anionic
and have an electrophoretic mobility similar to the analyte anions of interest.
3.2.3 Electroosmotic Flow (EOF) -- the direction and velocity of electrolyte solution
flow within the capillary under an applied electrical potential (voltage); the
velocity and direction of flow is determined by electrolyte chemistry, capillary
wall chemistry, and applied voltage.
3.2.4 Electroosmotic Flow Modifier (OFM) -- a cationic quaternary amine in the electrolyte
that dynamically coats the negatively charged silica wall giving it a net positive
charge. This reverses the direction of the electrolyte's natural electroosmotic flow
and directs it towards the anode and detector. This modifier augments anion
migration and enhances speed of analysis. Its concentration secondarily effects
anion selectivity and resolution. See Fig. 7.
1)
Annual Book of ASTM Standards, Vol. 11.01
2)
Annual Book of ASTM Standards, Vol. 11.02

 
3
3.2.5 Electrophoretic Mobility -- the specific velocity of a charged analyte in the
electrolyte under specific electroosmotic flow conditions. The mobility of an
analyte is directly related to the analyte's equivalent ionic conductance and
applied voltage, and is the primary mechanism of separation.
3.2.6 Electropherogram -- a graphical presentation of UV detector response versus
time of analysis; the x axis is migration time which is used to qualitatively identify
the anion, and the y axis is UV response which can be converted to time
corrected peak area for quantitation.
3.2.7 Hydrostatic Sampling -- a sample introduction technique in which the capillary
with electrolyte is immersed in the sample, and both are elevated to a specific
height, typically 10 cm, above the receiving electrolyte reservoir for a preset
amount of time, typically less than 60 s. Nanolitres of sample are siphoned into
the capillary by differential head pressure and gravity.
3.2.8 Indirect UV Detection -- a form of UV detection in which the analyte displaces an
equivalent net charge amount of the highly UV absorbing component of the
electrolyte causing a net decrease in background absorbance. The magnitude
of the decreased absorbance is directly proportional to analyte concentration.
Detector output polarity is reversed in order to obtain a positive mV response.
3.2.9.Midpoint of Peak Width -- CIE peaks are typically asymmetrical with the peak
apex shifting with increasing concentration, and peak apex may not be
indicative of true analyte migration time. Midpoint of peak width is the midpoint
between the analyte peak's start and stop integration, or the peak center of
gravity.
3.2.10 Migration Time -- the time required for a specific analyte to migrate through the
capillary to the detector. The migration time in capillary ion electrophoresis is
analogous to retention time in chromatography.
3.2.11 Time Corrected Peak Area -- normalized peak area; peak area divided by
migration time. CE principles state that peak area is dependent upon migration
time, i.e. for the same concentration of analyte, as migration time increases
(decreases) peak area increases (decreases). Time corrected peak area
accounts for these changes.
4 Summary of Test Method
4.1 Capillary ion electrophoresis, see Fig. 7 through Fig. 10, is a free zone
electrophoretic technique optimized for the determination of anions with molecular
weight less than 200. The anions migrate and are separated according to their
mobility in the electrolyte when an electrical field is applied through the open
tubular fused silica capillary. The electrolyte's electroosmotic low modifier
dynamically coats the inner wall of the capillary changing the surface to a net
positive charge. This reversal of wall charge reverses the natural EOF. The
modified EOF in combination with a negative power supply augments the mobility
of the analyte anions towards the anode and detector achieving rapid analysis
times. Cations migrate in the opposite direction towards the cathode and are
removed from the sample during analysis. Water and other neutral species move
toward the detector at the same rate as the EOF. The neutral species migrate
slower than the analyte anions and do not interfere with anion analysis. See Fig.
7 and 8.

 
4
4.2 Sample is introduced into the capillary using hydrostatic sampling. The inlet of the
capillary containing electrolyte is immersed in the sample and the height of the
sample raised 10 cm for 30 s where low nanolitre volumes are siphoned into the
capillary. After sample loading, the capillary is immediately immersed back into
the electrolyte. The voltage is applied initiating the separation process.
4.3 Anion detection is based upon the principles of indirect UV detection. The UV
absorbing electrolyte anion is displaced charge-for-charge by the separated
analyte anion. The analyte anion zone has a net decrease in background
absorbance. This decrease in UV absorbance is quantitatively proportional to
analyte anion concentration. See Fig. 9. Detector output polarity is reversed to
provide positive mV response to the data system, and to make the negative
absorbance peaks appear positive.
4.4 The analysis is complete once the last anion of interest is detected. The capillary
is automatically vacuum purged by the system of any remaining sample, and
replenished with fresh electrolyte. The system is now ready for the next analysis.
5 Significance and Use
5.1 Capillary ion electrophoresis provides a simultaneous separation and
determination of several inorganic anions using nanolitres of sample in a single
injection. All anions present in the sample matrix will be visualized yielding an
anionic profile of the sample.
5.2 Analysis time is less than 5 min with sufficient sensitivity for drinking water, and
wastewater applications. Time between samplings is less than 7 minutes allowing
for high sample throughput.
5.3 Minimal sample preparation is necessary for drinking water and wastewater
matrices. Typically only a dilution with water is needed.
5.4 This test method is intended as an alternative to other multi-analyte methods and
various wet chemistries for the determination of inorganic anions in water and
wastewater. Compared to other multi-analyte methods the major benefits of CIE
are speed of analysis, simplicity, and reduced reagent consumption and operating
costs.
6 Interferences
6.1 Analyte identification, quantitation, and possible comigration occur when one
anion is in significant excess to other anions in the sample matrix. For two
adjacent peaks, reliable quantitation can be achieved when the concentration
differential is less than 100:1. As the resolution between two anion peaks
increase so does the tolerated concentration differential. In samples containing
1000
mg/L CI,
1 mg/L SO
4
can be resolved and quantitated, however, the high CI
will interfere with Br and NO
2
quantitation.
6.2 Dissolved carbonate, detected as HCO3
1
, is an anion present in all aqueous
samples, especially alkaline samples. Carbonate concentrations greater than 500
mg/L will interfere with
PO4
quantitation.
6.3 Monovalent organic acids, except for formate, and neutral organics commonly
found
in wastewater migrate later in the electropherogram, after carbonate, and
do not interfere. Formate, a common organic acid found in environmental
samples, migrates shortly after fluoride but before phosphate. Formate
'concentrations greater than 5 mg/L will interfere with fluoride identification and
quantitation. Inclusion of 2 mg/L formate into the Mixed Anion Working Solution
aids in fluoride and formate identification and quantitation.

 
5
6.4 Divalent organic acids usually found in wastewater migrate after phosphate. At
high concentrations, greater than 10 mg/L, they may interfere with phosphate
identification and quantitation.
6.5 Chlorate also migrates after phosphate and at concentrations greater than 10
mg/L will interfere with phosphate identification and quantitation. Inclusion of 5
mg/L chlorate into the Mixed Anion Working Solution aids in phosphate and
chlorate identification and quantitation.
6.6 As analyte concentration increases, analyte peak shape becomes asymmetrical.
If adjacent analyte peaks are not baseline resolved, the data system will drop a
perpendicular between them to the baseline. This causes a decrease in peak
area for both analyte peaks and a low bias for analyte amounts. For optimal
quantitation, insure that adjacent peaks are fully resolved, if they are not, dilute
the sample 1:1 with water.
6.7 Samples containing high levels of TOC, total organic carbon, may effect the
observed analyte migration times. The TOC binds to the capillary surface
decreasing the EOF and increasing analyte migration times. Refer to Figure 7.
However, the change in EOF does not effect analyte selectivity. Analytes are
identified using normalized analyte migration times with respect to a reference
peak, chloride, always the first peak in the electropherogram. The surface can be
regenerated with a 5 minute wash with 500 mM Na0H.
7
.Apparatus
7.1 Capillary Ion Electrophoresis System -- the system consists of the following
components, as shown in Fig. 10, or equivalent:3
7.1.1
High Voltage Power Supply --
capable of generating voltage (potential)
between 0 and minus 30 kV relative to ground with the capability working in a
constant current mode.
7.1.2
Covered Sample Carousel --
to prevent environmental contamination of the
samples and electrolytes during a multi-sample batch analysis.
7.1.3
Sample Introduction Mechanism –
capable of hydrostatic sampling technique,
using gravity, positive pressure, or equivalent.
7.1.4
Capillary Purge Mechanism --
to purge the capillary after every analysis with
fresh electrolyte to eliminate any interference from the previous sample matrix,
and to clean the capillary with other reagents, such as sodium hydroxide.
7.1.5 UV Detector --
having the capability of monitoring 254 nm, or equivalent, with a
time constant of 0.3 s.
7.1.6
Fused Silica Capillary --
a 75 pm (inner diameter) x 375 pm (outer diameter) x
60 cm (length) having a polymer coating for flexibility, and a non-coated
section to act as the cell window for UV detection.''
7.1.7
Constant Temperature Compartment --
to keep the samples, capillary, and
electrolytes at constant temperature.
3) Available from Waters, 34 Maple St., Milford, Ma., 01757, 800/252-4752.

 
6
7.2 Data System -- computer system that can acquire data at 20 points per second
minimum, express migration time in minutes to 3 decimal places, use midpoint of
the analyte peak width, or center of gravity, to determine the analyte migration
time, use normalized migration times with respect to a reference peak for
qualitative identification, use time corrected peak area response for analyte
quantitation, and express results in concentration units.3
Note 2: It is recommended that integrators or standard chromatographic data processing not be used with
this test method.
7.3 Anion Exchange Cartridges in the Hydroxide form.4
7.4 Plastic Syringe -- 20 mL, Disposable.
7.5 Vacuum Filtration Apparatus -- capable for filtering 100 mL of reagent through a
0.45 gm aqueous filter.
8 Reagents and Materials
8.1 Purity of Reagents: -- Unless otherwise indicated, it is intended that all reagents
shall conform to the reagent grade specification of the Analytical Reagents of the
American Chemical Society, where such specifications are available. Other
grades may be used, provided it is first ascertained that the reagent is of sufficient
high purity to permit its use without lessening the performance or accuracy of the
determination. Reagent chemicals shall be used for all tests.
Note 3: Calibration and detection limits of this method are biased by the purity of the reagents.
8.2 Purity of Water:-- Unless otherwise indicated, references to water shall be
understood to mean Type I reagent water conforming or exceeding specification
D1193. Freshly drawn water should be used fcr preparation of all stock and
working standards, electrolytes, and solutions. Performance and detection limits
of this method are limited by the purity of reagent water, especially TOC.
8.3 Reagent Blank: Reagent Water or any other solution used to preserve or dilute
the sample.
8.4 Individual Anion Solution, Stock:
Note 4: It is suggested that certified individual
1000
mg/L anion standards be purchased for use
with this test method.
Note 5: All weights given are for anhydrous or dried salts. Must account for reagent purity to
calculate true value concentration. Certify against NIST traceable standards.
8.4.1 Bromide Solution, Standard (1.0 mL = 1.00
MCI
Bromide):
Dry approximately 0.5 g of sodium bromide (NaBr) for 6 h at 150°C and cool in
a desiccator. Dissolve 0.128 g of the dry salt in a 100 mL volumetric flask with
water, and fill to mark with water.
4)
Available from Alltech Associates, P/N 30254, 2051 Waukegan Rd, Deerfield IL, 60015, 847/948-
8600.
5)
Reagent Chemicals, American Chemical Society Specifications,
Am. Chem. Soc., Washington, DC
For suggestions on the testing of reagents not listed by the American Chemical Society, see
Analar
Standards for Laboratory Chemicals,
BDH Ltd., Poole, Dorset. U.K., and the United States
Pharmacopeia and National Formulary, U.S.
Pharmacopoeia Convention, Inc. (USPC), Rockville,
Md.
6)
Although the reagent water may exceed
D1193 specification, the reagent water needs to be
periodically tested for bacterial contamination. Bacteria and their waste products may adversely
affect system performance. As a guide, ASTM type IA water specifies a total bacteria count of 10
colonies/L. Refer to Test Method F 488 for analysis procedure.

 
7
8.4.2 Chloride Solution, Standard (1.0 mL = 1.00 ma Chloride):
Dry approximately 0.5 g of sodium chloride (NaCI) for 1 h at 100°C and cool in
a desiccator. Dissolve 0.165 g of the dry salt in a 100 mL a volumetric flask
with water, and fill to mark with water.
8.4.3 Fluoride Solution, Standard (1.0 mL = 1.00 mq Fluoride):
Dry approximately 0.5 g of sodium fluoride (NaF) for 1 h at 100°C and cool in,a
desiccator. Dissolve 0.221 g of the dry salt in a 100 mL volumetric flask with
water, and fill to mark with water.
8.4.4 Formate Solution, Standard (1.0 mL = 1.00 mg Formate):
Dissolve 0.151 g of sodium formate in a 100 mL volumetric flask with water, and fill
to mark with water.
8.4.5 Nitrate Solution, Standard (1.0 mL = 1.00
ma
Nitrate):
Dry approximately 0.5 g of sodium nitrate (NaNO3)
for 48 h at 105°C and cool
in a desiccator. Dissolve 0.137 g of the dry salt in a 100 mL volumetric flask
with water, and fill to mark with water.
8.4.6 Nitrite Solution, Standard (1.0 mL = 1.00 mg Nitrite):
Dry approximately 0.5 g of sodium nitrite (NaNO2)
for 24 h in a desiccator
containing concentrated sulfuric acid. Dissolve 0.150 g of the dry salt in a 100
mL volumetric flask with water, and fill to mark with water. Store in a sterilized
glass bottle. Refrigerate and prepare monthly.
Note 6: Nitrite is easily oxidized, especially in the presence of moisture. Use only fresh
reagent.
Note 7: Prepare sterile bottles for storing nitrite solutions by heating for 1 h at 170
0C
in an air
oven.
8.4.7 Ortho-Phosphate Solution, Standard (1.0 mL = 1.00 ma o-Phosphate):
Dissolve 0.150 g of anhydrous dibasic sodium phosphate (Na
2HPO4)
in a 100
mL volumetric flask with water, and fill to mark with water.
8.4.8 Sulfate Solution, Standard (1.0 mL = 1.00 mg Sulfate):
Dry approximately 0.5 g of anhydrous sodium sulfate (Na
2SO4)
for 1 h at
110°C and cool in a desiccator. Dissolve 0.148 g of the dry salt in a 100 mL
volumetric flask with water, and fill to mark with water
8.5 Mixed Anion Solution, Working: Prepare a 0.2 mg/L and at least 3 different
working standards concentrations for the analyte anions of interest bracketing the
desired range of analysis, typically between 0.2 and 50 mg/L, and add 2 mg/L
formate to all standards. Add an appropriate aliquot of Individual Anion Stock
Solution (8.4) to a pre-rinsed 100 mL volumetric flask, and dilute to 100 mL with
water.
Note 8: Use 100 p.1_ of Individual Anion Stock Solution (8.4) per 100 mL for 1 mg/L anion.
Note 9: Anions of no interest may be omitted.
Note 10: The mid-range Mixed Anion Solution, Working may be used for the determination of
migration times and resolution described in 12.1.
8.6 Calibration Verification Solution (CVS): A solution formulated by the laboratory of
mixed analytes of known concentration prepared in water. The CVS solution must
be prepared from a different source to the calibration standards.
8.7 Performance Evaluation Solution (PES): A solution formulated by an independent
source of mixed analytes of known concentration prepared in water. Ideally, the
PES solution should be purchased from an independent source.

 
8
8.8 Quality Control Solution (QCS): A solution of known analyte concentrations added
to a synthetic sample matrix such as Substitute Wastewater that sufficiently
challenges the Test Method.
8.9 Buffer Solution (100 mM CHES / 1 mM Calcium Gluconate): Dissolve 20.73 g of
CHES (2-[N-Cyclohexylamino]-Ethane Sulfonic Acid) and 0.43 g of Calcium
Gluconate in a 1 L volumetric flask with water, and dilute to 1 L with water. This
concentrate may be stored in a capped glass or plastic container for up to 1year.
8.10 Chromate Concentrate Solution (100 mM Sodium Chromate): Dissolve 23.41 g of
sodium chromate tetrahydrate (Na
2Cr04•4 H
2
0) in a 1 L volumetric flask with
water, and dilute to 1 L with water. This concentrate may be stored in a capped
glass or plastic container for up to 1 year.
8.11 OFM Concentrate Solution (100 mM Tetradecyltrimethyl Ammonium Bromide):
Dissolve 33.65 g of Tetradecyltrimethyl Ammonium Bromide (TTABr) in a 1 L
volumetric flask with water, and dilute to 1 L with water. Store this solution in a
capped glass or plastic container for up to 1 year.
Note 11: TTABr needs to be converted to the hydroxide form
7(TTAOH)
for use with this test method.
TTAOH is commercially available as 100 mM TTAOH which is an equivalent substitute.
8.12 Sodium Hydroxide Solution (500 mM Sodium Hydroxide)-- Dissolve 20 g
- of sodium
hydroxide (NaOH) in a 1 L
plastic volumetric flask with water, and dilute to 1 L with
water.
8.13 Electrolyte Solution, Working (4.7 mM Chromate / 4 mM TTAOH / 10 mM CHES /
0.1 mM Calcium Gluconate) : Wash the anion exchange cartridge in the hydroxide
form (7.3) using the 20 mL plastic syringe (7.4) with 10 mL of 500 mM NaOH
(8.12) followed by
10
mL of water. Discard the washings. Slowly pass 4 mL of the
100 mM TTABr Solution (8.11) through the cartridge into a 100 mL volumetric
flask. Rinse the cartridge with 20 mL of water, adding the washing to the
volumetric flask.
Note 12: The above procedure is used to convert the TTABr to TTAOH, which is used in the
electrolyte. If using commercially available 100 mM TTAOH, the above conversion step is not
necessary; substitute 0.5 mL of 100 mM TTAOH and continue below
Into the 100 mL volumetric flask add 4.7 mL of Chromate Concentrate Solution
(8.10) and 10 mL of Buffer solution (8.9). Mix and dilute to 100 mL with water.
The natural pH of the electrolyte should be 9 ± 0.1. Filter and degas using the
vacuum filtration apparatus. Store the any remaining electrolyte in a capped
glass or plastic container at ambient temperature. The electrolyte is stable for 1
year.
7)
Available from Waters Corp. as lonSelect 100mM OFM Hydroxide Concentrate, 100 mM
TTAOH, P/N 49387.
8)
Availiable from Waters Corp. as lonSelect High Mobility Anion Electrolyte, P/N 49385.

 
9
9 Precautions
9.1 Chemicals used in this test method are typical of many useful laboratory
chemicals, reagents and cleaning solutions, which can be hazardous if not
handled properly. Refer to Guide D 3856.
9.2 It is the responsibility of the user to prepare, handle, and dispose of chemical
solutions in accordance with all applicable federal, state, and local regulations.
9.3
Warning --
This capillary electrophoresis method uses high voltage as a means
for separating the analyte anions, and can be hazardous if not used properly.
Use only those instruments that have all proper safety features.
10 Sampling
10.1 Collect samples in accordance with Practice D 3370.
10.2 Rinse samples containers with sample and discard to eliminate any
contamination from the container. Fill to overflowing and cap to exclude air.
10.3 Analyze samples as soon as possible after collection. For nitrite, nitrate, and
phosphate refrigerate the sample at 4°C after collection. Warm to room
temperature before dilution and analysis.
10.4 At the lab, filter samples containing suspended solids through a pre-rinsed 0.45
pm aqueous compatible membrane filter before analysis.
10.5 If sample dilution is required to remain within the scope of this Test Method,
dilute with water only.
11 Preparation of Apparatus
11.1 Set up the CE and data system according to the manufacturer's instructions.
11 2 Program the CE system to maintain a constant temperature of 25° ± 0.5°C; or
5°C above ambient laboratory temperature. Fill the electrolyte reservoirs with
fresh chromate electrolyte working solution (8.13), and allow 10 min for thermal
equilibration.
11.3 Condition a new capillary (7.1.6) with 500 mM NaOH Solution (8.12) for 5 min
followed by water for 5 min. Purge the capillary with electrolyte (8.13) for 3 min.
11.4 Apply 15 kV of voltage and test for current. The current should be 14 ± 1 IA. If
no current is observed, then there is a bubble and/or blockage in the capillary.
Degas the chromate electrolyte working solution and retry. If still no current,
replace the capillary.
11.5 Set the UV detector to 254 nm detection, or equivalent. Zero the detector to
0.000 absorbance. UV offset is less then 0.1 AU.
11.6 Program the CE system for constant current of 14 gA.
11.7 Program the CE system for a hydrostatic sampling of 30 s. Approximately 37nL
of sample is siphoned into the capillary. Different sampling times may be used
provided that the samples and standards are analyzed identically.
11.8 Program the CE system for a 1 min purge with the chromate electrolyte working
• solution between each analysis. Using a 15 psi vacuum purge mechanism, one
60 cm capillary volume can be displaced in 30 s.

 
10
11.9 Program the data system for an acquisition rate of at least 20 points per s.
Program the data system to identify analyte peaks based upon normalized
migration time using CI as the reference peak, and to quantitate analyte peak
response using time corrected peak area.
Note 13: Under the analysis conditions CI is always the first peak in the electropherogram, and
can be used as a migration time reference peak.
12 Calibration
12.1 Determination of Migration Times-- Calibrate Daily. The migration time of an
anion is dependent upon the electrolyte composition, pH, capillary surface and
length, applied voltage, the ionic strength of the sample, and temperature. For
every fresh electrolyte determine the analyte migration time, in min to the third
decimal place, of the mid-range mixed anion standard working solution (8.5),
described in Sec 11. Use the mid-point of analyte peak width as the determinant
of analyte migration time.
Note 14: Analyte peak apex may be used as the migration time determinant, but potential analyte
misidentification may result with asymmetrical peak shape at high analyte concentrations.
12.2 Analyze the blank (8.3), a 0.2 mg/L, and at least 3 working mg/L solutions (8.5),
using the set-up described in sec 11. For each anion concentration (X-axis) plot
time corrected peak area response (Y-axis). Determine the best linear
calibration line through the data points, or use the linear regression calibration
routine (1/X Weighting and Linear Through Zero) available in the data system.
Note 15: Do not use peak height for calibration. Peak area is directly related to migration time, i.e.
for the same analyte concentration, increasing migration time gives increasing peak area.
Note 16: EPA recommends calibration at the minimum concentration of 0.2 mg/L plus 3 additional
points.
The r2
(coefficient of determination) values should be greater than 0.995; typical
r2
values obtained from the interlaboratory collaborative are given in Table A2.
12.3 Calibrate daily and with each change in electrolyte, and validate by analyzing the
CVS solution (8.6) according to procedure in Sec16.4.
12.4 After validation of linear multiple point calibration, a single point calibration
solution can be used between 0.2 and 50 mg/L for recalibration provided the
quality control requirements in Sec 16.4 are met.
13 Procedure
13.1 Dilute the sample, if necessary with water, to remain within the scope (Sec 1.2,
1.3) and calibration of this test method. Refer to A1.5.1.
13.2 Analyze all blanks (8.3), standards (8.5), and samples as described in Sec 11
using the quality control criteria described in Sec 16.5 to 16.9. Refer to Fig. 1
through 5 for representative anion standard, detection limit standard, substitute
wastewater, drinking water, and wastewater electropherograms.
13.3 Analyze all blanks, calibration standards, samples, and quality control solutions
in singlicate. Perform at least one matrix spike analysis in duplicate as part of
the QC protocol, Sec 16.7. Optional: Duplicate analyses are preferred due to
short analysis times.
Note 17: Collaborative data was acquired, submitted and evaluated as the average of duplicate
samplings.

 
11
13.4 After 20 sample analyses, or batch, analyze the QCS solution (8.8). If
necessary, recalibrate using a single mixed anion standard working solution
(8.5), and replace analyte migration time.
Note 18: A change in analyte migration time of the mixed anion standard working solution by more
than +5% suggests that components in the previously analyzed sample matrices have
contaminated the capillary surface. Refer to
sec
6.7. Continue but wash the capillary with
NaOH solution (8.12) before the next change in electrolyte.
14 Calculation
14.1 Relate the time corrected peak area response for each analyte with the
calibration curve generated in section 12.2 to determine mg/L concentration of
analyte anion. If the sample was diluted prior to analysis, then multiply mg/L
anion by the dilution factor to obtain the original sample concentration, as
follows:
Original Sample mg/L Analyte = (A x SF) where;
A = analyte concentration determined from the calibration curve, in mg/L,
SF = scale or dilution factor.
15 Report Format
15.1 The sample analysis report should contain the sample name, analyte anion
name, migration time reported to 3 decimal places, migration time ratio, peak
area, time corrected peak area, sample dilution, and original solution analyte
concentration. Optional: Report analysis method parameters, date of sample
data acquisition, and date of result processing for documentation and validation
purposes.
16 Quality Control
16.1 Before this test method is applied to the analysis of unknown samples, the
analyst should establish quality control according to procedures recommended in
Practice D5847, and Guide D5810.
16.2 The laboratory using this test must perform an initial demonstration of laboratory
capability according to procedures outlined in Standard Practice D5847, and
Appendix C.
Note 19: Certified Performance Evaluation Solutions (PES) and QC Solutions (QCS and CVS) are
commercially available, and recommended.
16.3 Initial Demonstration of Performance: Analyze seven replicates of a Performance
Evaluation Solution (PES, 8.7). Calculate analyte concentration mean and
standard deviation of the seven replicates and compare to the precision and
Initial %Recovery for the analyte in reagent water given in Table 8.
16.3.1 Repeat the 7 replicate analysis protocol before using a freshly prepared QVS
solution (8.6) and QCS solution (8.8) for the first time. Calculate the standard
deviation and compare with previous results using the student t-test. If no
significant difference is noted then use the combined standard deviation to
determine the QC limits, for the QVS and QCS solutions.
16.4 Calibration Verification: After calibration, verify the calibration linearity and
acceptable instrument performance using a Calibration Verification Solution (8.6)
treated as an unknown. If the determined CVS concentrations (8.6) are not
within ± 3 standard deviations of the known true values as described in 16.3.1,
the calibration solutions may be out of control. Reanalyze, and if analyte
concentration still falls outside the acceptable limits, fresh calibration solutions
(8.5) are required. Successful CVS analyte concentration must be confirmed
after recalibration before continuing with the Test Method.

 
12
16.5 Analyze a reagent blank (8.3) with each batch to check for contamination
introduced by the laboratory or use of the Test Method.
16.6 Quality Control Solution: Analyze one QCS (8.8) after 20 samples, or batch. The
analyte concentrations for the QCS should fall within the lower limit (LL) and
upper limits (UL) given in Table 8.
16.7 Matrix Spike Recovery: One Matrix Spike (MS) must be analyzed in duplicate
with each batch of samples to test method recovery and relative %difference
between them. Spike a portion of one sample from each batch with a known
concentration of analyte, prepared in accordance with Guide D3856. The
cY0
recovery of the spike should fall within the MS/MSD lower and upper limits, and
the Relative %Difference given in Table 8 for the appropriate sample matrix. If it
does not, an interference may be present and the data for the set of similar
samples matrices must be qualified with a warning that the data are suspect, or
an alternate test method should be used. Refer to Guide D5810.
16.7.1 If the known analyte concentration is between 15 and 50 mg/L, then spike the
sample solution to increase analyte concentration by 50%.
16.7.2 If the known analyte concentration is between 2 mg/L and 15 mg/L, then
spike the sample solution to increase analyte concentration by 100%, but not
less than 2 mg/L.
16.7.3 If the known analyte concentration is less than 2 mg/L, then spike the sample
solution with 1 mg/L, 5 times the ML.
16.7.4 Calculate the percent recovery of the spike using the following formula:
% Recovery = 100 [A (Vs + V) - B Vs] / C V where
A = Analyte Concentration (mg/L) in Spiked Sample
B = Analyte Concentration (mg/L) in Unspiked Sample
C = Concentration (mg/L) of Analyte in Spiking Solution
Vs = Volume (mL) of Sample Used
V = Volume (mL) Added with Spike.
Evaluate performance according to Practice D5847.
16.8 Method Precision: One unknown sample should be analyzed in triplicate with
each batch to test method precision. Calculate the standard deviation and use
the F-Test to compare with the single operator precision given in Tables 1
through 7 for the equivalent analyte concentration and matrix type. Evaluate
performance according to Practice D5847.
16.9 The laboratory may perform additional quality control as desired or appropriate.
17 Precision and Bias
17.1 The precision and bias data presented in this test method meet the requirements
of Practice 2777-98, and are given in Tables 1 through 7. The full Research
Report, RR# D19-1165, can be obtained from ASTM Headquarters.
17.2 This test method interlaboratory collaborative was performed by 11 laboratories
using one operator each. Four Youden Pair spike concentrations for the 7
analytes anions yielding 8 analyte concentration levels. Test data was submitted
for 11 Reagent Waters, 11 Substitute Wastewaters, 15 Drinking Waters, and 13
Wastewater sample matrices.

 
13
17.3 All data given in this method was quantitated using non-weighted linear
calibration through zero, except where noted.
17.4 The precision, bias, and matrix recovery of this test method per anion analyte in
the 4 tested sample matrices are based upon the analyte true value, calculated
using weight, volume, and purity. True value spiking solution concentrations are
given in Table A4.
17.5 The bias and matrix recovery statements for less than 2 mg/L of chloride, sulfate,
and nitrate in naturally occurring sample matrices may be misleading due to
spiking of small analyte concentration into a high naturally occurring analyte
concentration observed with the matrix blank. The commonly occurring analyte
concentrations observed in the sample matrix blanks for the naturally occurring
tested matrices are given in Table A5.
17.6 The high nitrate bias and %recovery noted for the 0.5 mg/L NO
3
spike solution
are attributed to the spiking solution containing 50 mg/L nitrite and 0.5 mg/L
nitrate. Refer to Appendix Table A4, Solution 3. Some of the nitrite converted to
nitrate prior to analysis. Similar NOx conversion effect is observed with the 2
mg/L nitrate and 2 mg/L nitrite spike, Solution 7.
17.7 All collaborative participants used the premade Chromate electrolyte, (lOnSelect
High Mobility Anion Electrolyte, available from Waters Corp.) Ten laboratories
used a Waters CIA Analyzer with Millennium Data Processing Software, and one
laboratory used a Agilent CE System with Diode Array Detector that provided
equivalent results, although different sampling and detection conditions were
necessary for equivalent performance.
Note 20: Refer to reference B1.16 and Agilent (the former HP Company) website for
recommended operating conditions.
18 Key Words
Anion
Capillary Electrophoresis
Drinking Water
Ion Analysis
Reagent Water
Substitute Wastewater
Wastewater

 
14
Appendix A
Mandatory Information
A1.1 All data presented in the following Tables conform and exceed the requirements of
D2777-98. Data from eleven reagent waters, eleven substitute wastewater, fifteen
Drinking Water, and thirteen wastewater sample matrices, were tested using a set
of 4 Youden Pair concentrations for 7 analyte anions. All submitted individual data
points are the average of duplicate samplings.
A1.2 Calibration Linearity
A1.2.1 All laboratories used a provided set of 4 certified, mixed anion calibration
solutions in concentrations between 0.5 mg/L and 50 mg/L, formulated in
random concentrations given in Table A1. They were prepared from certified,
individual 1000 mg/L Stock Standards obtained from APG, Inc, Belpre, Ohio.
No dilution was necessary.
Table Al: Collaborative Calibration Standard, mg/L Concentrations
Analyte Anion
Standard 1
Standard 2
Standard 3
Standard 4
Chloride
50
25
0.5
10
Bromide
0.5
25
10
50
Nitrite
25
0.5
50
10
Sulfate
10
25
0.5
50
Nitrate
25
0.5
50
10
Fluoride
5
0.5
10
25
Phosphate
50
25
0.5
10
A1.2.2 A Linear Through Zero; no weighting regression was used to calculate the
calibration curve. The range coefficient of determination (r
2 ) values obtained
from the collaborative is shown in Table A2
Table A2: Expected Range of (r
2) Coefficient of Determination
Anion
?
/
Average, n=29
Lowest
Highest
Chloride
0.99987
0.99959
0.99997
Bromide
0.99953
0.99878
0.99996
Nitrite
0.99983
0.99961
0.99999
Sulfate
0.99976
0.99901
0.99999
Nitrate
0.99957
0.99840
0.99999
Fluoride
0.99972
0.99797
0.99999
Phosphate
0.99982
0.99942 0.99999
A1.2.3 EPA requires that 1/X weighting be used for calibration. The P & B data
were derived using unweighted calibration. Table A2a shows there is no
significant difference in r2
linearity between these 2 calibration routines.
Table A2a Coefficient of Determination r
2
from a Single Calibration
Analyte
Anion
No Weighted 1/x Weighted
Calibration?
Calibration
Chloride?
0.99994
Bromide
?
0.99942
Nitrite?
0.99975
Sulfate?
0.99971
Nitrate?
0.99975
Fluoride
?
0.99986
Phosphate
?
0.99999
0.99996
0.99923
0.99981
0.99974
0.99974
0.99967
0.99999

 
15
A1.3 Quality Control Solution Preparation
A1.3.1 The Quality Control Solution (QCS) was also used as the Calibration
Verification Solution (CVS).
A1.3.2 Quality Control Solution (QCS) was manufactured, analyzed using ion
chromatography, and certified by APG as 100X concentrate, to replicate typical
Drinking Water concentrations. Required 1:100 dilution with water before
analysis. The QCS analyte concentrations, required control limits, and
interlaboratory determined control limits based upon n# analyses are given in
Table A3.
Table A3: Quality Control Acceptance Limits
Analyte
Anion
True Value
mg/L
Certified
Value
mg/L
Required
99%
Confidence
Interval
Determined
QCS
Mean ± Std Dev,
n = 82
Chloride
Bromide
48.68
0.00
48.61 ± 0.12
0.00
43.99 - 52.96
0.00
47.64
±1.53
0.00
Nitrite
2.87
2.90 ± 0.07
2.39 - 3.26
2.88 ± 0.19
Sulfate
35.69
35.63
±
0.25
29.54 - 40.53
35.02
± 1.21
Nitrate
15.76
15.78 ± 0.15
12.80
-18.39
15.33 ±
4.35
Fluoride
1.69
1.68 ± 0.01
1.49
-1.87
1.67 ± 0.09
Phosphate
5.47
5.55
± 0.12
4.78
- 6.20
5.58
± 0.28
A1.3.3 A single day's QCS was reprocessed using a 1/X weighting linear
calibration and remained within the QC Acceptance Limits.
Table A3a QC Standard Results: Reprocessed Using 1/x Calibration
Analyte
Anion
No Weighted
Calibration
1/x Weighted?
QC Acceptance
Calibration?
99%Conf Interval
Chloride
48.64 ± 1.06
48.77 ± 1.07
43.99
- 52.96
Nitrite
2.93
± .03
2.82
± .03
2.39
- 3.26
Sulfate
34.49 ± .79
34.64 ± .79
29.54
- 40.53
Nitrate
15.28
± .15
15.23
± .18
12.80
-18.39
Fluoride
1.74 ± .02
1.63
± .02
1.49
-1.87
Phosphate
5.75 ± .15
5.77
± .15
4.78
- 6.20
A1.4 Youden Pair Spiking Solution Preparation
A1.4.1 Eight mixed anion, 100X concentrate, spiking solutions were prepared in
accordance with Sec 8.3 (Reagents and Materials) of the test method using
anhydrous sodium salts. The mg/L concentrations of the eight standards
followed the approved Youden Pair design - 0.5 & 0.7, 2 & 3, 15 & 20, 40 & 50
mg/I for all anions except fluoride, which is 0.5 & 0.7, 2 & 3, 7& 10, 20 &
25mg/L. The analyte true value concentrations were randomized among the
eight spiking solutions as described in Table A4.
A1.4.2 A ninth solution containing approximately 10 mg/L of each analyte was
diluted 1:50 with water, and was used for method detection limit calculations.

 
16
Table A4: True Value Youden Pair Spiking mg/L Concentrations
Anion / TV
1
2
3
4
5
6
7
89
Chloride
0.71
2.00
2.98
14.92
39.81
19.91
49.76
0.50
10.20
Bromide
2.00
3.01
14.93
39.81
19.91
49.77
0.70
0.51
10.49
Nitrite
2.98
39.61
19.81
14.86
49.52
0.50
2.00
0.70
9.94
Sulfate
39.60
49.51
0.49
0.70
1.98
2.98
14.86
19.81
10.23
Nitrate
14.92
19.19
39.87
49.78
0.50
0.70
2.00
2.98
10.35
Fluoride
2.00
0.71
0.50
3.00
9.99
6.99
19.98
24.99
10.40
Phosphate
49.51
39.60
19.90
0.50
2.98
1.99
0.69
14.86
10.48
These solutions, kept at ambient temperature, were analyzed before and during
the collaborative to monitor for accuracy and stability. The mg/L True Value in
was used to determine bias, matrix recovery, and the single operator and
interlaboratory precision in the P & B tables per the requirement of D 2777.
Solution 3 and 7 exhibited some conversion of nitrite to nitrate before analysis.
This conversion is evident in the bias and % Recovery for 0.5 mg/L and 2 mg/I
nitrite and nitrate.
A1.5 Sample Matrix Preparation
A1.5.1 All participating laboratories provided and tested reagent water, substitute
wastewater, naturally occurring drinking water, and naturally occurring
wastewater. Before matrix spiking with the Youden Pair solutions, the sample
matrix was evaluated, then appropriately diluted to give the highest anion
concentration below 50 mg/L. The diluted sample matrix was used to dilute
each Youden Pair spiking solution 1:100.
A1.5.2 Reagent Water was used as-is. Substitute wastewater was diluted 1:20
with water. Naturally occurring drinking water was used as-is or diluted 1:5 with
water. Naturally occurring wastewater was diluted between 1:3 and 1:20, except
one which required a 1:1000 dilution due to high chloride.
A1.5.3 Due to the anion content of the naturally occurring drinking water and "real"
wastewater matrices, some of the reported spike matrix results exceeded the
scope of this test method. Linearity and matrix recovery data obtained from the
collaborative indicated that these data are acceptable, and extended the useful
range of this test method.
A1.5.4 Due to the anion content of the naturally occurring sample matrices given
in Table A5, the low concentration bias and recovery may be misleading
because of spiking a low anion concentration increment into a large naturally
occurring concentration of the same anion.
Table A5: Blank Analvte Concentrations for Naturally Occurring Sample Matrices
Data in mg/L
Chloride
Sulfate
Nitrate
Drinking Water
0.7 to 41.9
0.5 to 33.6
0.2 to 6.5
Substitute Wastewater
20.5 to 25.5 3.2 to 4.0
Not Detected
"Real"
0.9 to 43.4
0.5 to 50.4
0.3 to 23.0
Wastewater

 
17
A1.6 Test Method Detection Limits:
A1.6.1 Spiking Solution #9, containing 10 mg/L of each analyte, was diluted 1:50 with
water and was used for detection limit calculations. Ten laboratories performed
seven replicate samplings, and the mean and standard deviation from each
laboratory was calculated. The mean time corrected peak area response for the
7 replicates was given the true value of the solution #9, and from a simple
proportion, the standard deviation was calculated as mg/L.
Std Dev, mg/L = (True Value Conc Sol'n #9, mo/L)(Response Std Dev)
Ave Response of Sol'n #9
A1.6.2 Method detection limits (MDL) were derived using "pooled" EPA protocol and
the student t-test at 6 degrees of freedom, as follows;
The method detection limit (MDL) .(3.14)(Std Dev, mg/L).
A1.6.3 The upper and lower confidence limits were calculated as;
95% Confidence Interval:
?
LCL (Lower Confidence Limit) = 0.64 x MDL
UCL (Upper Confidence Limit) = 2.20 x MDL
A1.6.4 Method Detection Limits are given in Table A6.
Table A6: Method Detection Limits
Anion
mg/L Solution?
Method Detection I 95% Confidence interval
Concentration?
MDL, mg/L?
mg/L
Chloride
0.204
0.075
0.048 to 0.165
Bromide
0.210
0.120
0.077 to 0.264
Nitrite
0.199
0.103
0.066 to 0.227
Sulfate
0.205
0.065
0.042 to 0.143
Nitrate
0.207
0.076
0.049 to 0.167
Fluoride
0.208
0.032
0.020 to 0.070
Phosphate
0.210
0.097
0.062 to 0.213

 
18
Table 1
Precision, Bias, and Matrix Recovery for Chloride
Matrix
?
# of
?
True?
Mean
?
Bias vs?
Recovery
?
Interlab
?
Interlab?
Single?
Analyst
Values?
Value?
Result
?
True?
vs True?
Std Dev
?
%RSD
?
Operator
?
%RSD
?
Value?
Value
?
S(t)
?
Std Dev, S(o)
Reagent
?
9?
0.50?
0.55?
0.05?
110.0?
0.11?
19.8
Water?
10
?
0.71?
0.69?
-0.02?
97.2?
0.08
?
11.5?
0.05?
7.5
?
10
?
2.00?
1.97
?
-0.03?
98.5?
0.14?
6.8
?
9
?
2.98?
2.97?
-0.01?
99.7?
0.11?
3.8
?
0.05?
2.1
?
10?
14.92?
14.76
?
-0.16?
98.9?
0.61?
4.2
?
10?
19.91?
19.81?
-0.10
?
99.5?
0.81?
4.1
?0.48?
2.8
?
10?
39.81
?
38.58?
-1.23?
96.9?
1.43?
3.7
?
10?
49.76
?
48.70?
-1.06?
97.9?
1.94?
4.0?
1.36?
3.1
Substitute?
9?
0.50
?
0.46?
-0.04?
92.0?
0.51?
111.1
Wastewater
?
9?
0.71?
0.43?
-0.28?
60.6?
0.69?
160.7
?
0.42?
93.8
9
?
2.00?
1.52?
-0.48?
76.0?
0.68?
45.0
?
9?
2.98?
2.58
?
-0.40
?
86.6?
0.63?
24.5?
0.50?
24.3
?
9?
14.92?
14.29?
-0.63?
95.8
?
1.02?
7.1
?
9
?
19.91
?
18.93?
-0.98?
95.1?
1.24?
6.6
?
0.60?
3.6
?
9?39.81
?
37.34?
-2.47?
93.8?
5.44?
14.6
?
9?
49.76?
47.54?
-2.22?
95.5?
3.13?
6.6
?
4.43?
10.4
Drinking
?
12?0.50
?
0.63?
0.13?
126.0
?
0.67?
106.1
Water?
12
?
0.71?
0.75?
0.04?
105.6?
0.34?
45.5?
0.40?
57.2
?
12?
2.00?
2.15?
0.15?
107.5
?
0.51?
23.6
?
12?
2.98?
2.95?
-0.03?
99.0?
0.39?
13.1?
0.47
?
18.5
?
12?
14.92?
14.54
?
-0.38?
97.5?
0.71?
4.9
?
12?
19.91?
19.09?
-0.82?
95.9?
1.11
?
5.8
?
0.37?
2.2
?
12?
39.81
?
38.38?
-1.43?
96.4?
1.56?
4.1
49.76
?
47.97?
-1.79?
96.4?
2.19?
4.6?
1.26?
3.9
°Real"
?
9
?0.50?
0.42?
-0.08?
84.0
?
0.34?
81.0
Wastewater
?
10?
0.71?
0.47?
-0.24?
66.2?
0.34?
72.6?
0.26?
59.3
?
10
?2.00?
1.56?
-0.44
?
78.0?
0.51?
32.7
?
9?
2.98?
2.78
?
-0.20?
93.3
?
0.19?
6.8?
0.37?
17.3
?
10
?
14.92
?
14.29?
-0.63?
95.8
?
0.63?
4.4
?
10?
19.91?
18.83?
-1.08?
94.6?
0.78?
4.1
?0.46?
2.8
?
9?39.81?
37.01
?
-2.80?
93.0?
2.78?
7.5
?
10?
49.76
?
48.24?
-1.52?
96.9
?
3.15?
6.5?
2.54?
6.0

 
19
Table 2
Precision, Bias, and Matrix Recovery for Bromide
Matrix
?
# of?
True?
Mean?
Bias vs
?
Recovery
?
Interlab
?
Interlab?
Single?
Analyst
Values?
Value
?
Result?
True
?
vs True
?
Std Dev
?
%RSD
?
Operator
?
%RSD
Value?
Value?
S(t)
?
Std Dev, S(o)
Reagent?
10
?
0.51?
0.60?
0.09?
117.6?
0.19
?
31.0
Water?
10?
0.70?
0.83?
0.13?
118.6
?
0.23?
28.2?
0.10?
14.6
?
10?
2.00?
2.06?
0.06?
103.0
?
0.14?
6.6
?
10
?3.01?
2.88?
-0.13?
95.7?
0.23?
7.9?
0.15?
6.3
?
10
?
14.93
?
15.00
?
0.07?
100.5
?
0.58?
3.9
?
10
?
19.91
?
19.32
?
-0.59?
97.0?
0.97?
5.0
?
0.75?
4.4
?
10
?
39.81?
39.66
?
-0.15?
99.6
?
1.24?
3.1
?
10?
49.77
?
50.04
?
0.27?
100.5?
2.94?
5.9?
1.61?3.6
Substitute
?
9?
0.51?
0.67?
0.16?
131.4?
0.19?
28.8
Wastewater
?
9?
0.70?
0.96
?
0.26
?
137.1?
0.21?
21.8?
0.08?
9.3
?
9
?
2.00?
2.14?
0.14?
107.0
?
0.22
?
10.2
?
9
?
3.01?
2.72?
-0.29?
90.4?
0.35?
12.8?
0.17
?
7.0
?
9?
14.93
?
14.70?
-0.23?
98.5
?
0.58?
3.9
?
9
?
19.91
?
18.91
?
-1.00
?
95.0?
2.62?
13.8
?
1.63
?
9.7
?
9?
39.81
?
38.76?
-1.05
?
97.4
?
1.11?
2.9
?
9
?
49.77
?
48.81?
-0.96?
98.1?
1.52?
3.1
?
0.48
?
1.1
Drinking?
13
?
0.51
?
0.58?
0.07?
113.7?
0.25?
43.4
Water
?
13?
0.70?
0.83?
0.13
?
118.6
?
0.22?
26.5?
0.14?
19.9
?
13
?
2.00?
1.98?
-0.02?
99.0?
0.25
?
12.5
?
13
?3.01?
2.56?
-0.45?
85.0?
0.25?
9.7?
0.15?
6.8
?
13
?
14.93
?
14.63?
-0.30?
98.0
?
0.50
?
3.4
?
13?
19.91
?
19.22
?
-0.69
?
96.5?
1.10?
5.7?
0.77?
4.6
?13
?
39.81?
38.97?
-0.84?
97.9?
1.99
?
5.1
?
13?
49.77
?
48.74?
-1.03?
97.9?
1.49?
3.1?1.13?
2.6
"Real"
?
11
?
0.51
?
0.59?
0.08?
115.7?
0.11?
19.3
Wastewater
?
12
?0.70?
0.78?
0.08?
111.4
?
0.19?
24.4?
0.10?
14.0
?
11
?2.00?
2.08?
0.08?
104.0
?
0.13?
6.3
?
12
?3.01?
2.70?
-0.31?
89.7?
0.41?
15.1?
0.27?
11.5
?
12
?
14.93
?
15.16?
0.23?
101.5?
0.90
?
6.0
?
11
?
19.91
?
19.46?
-0.45
?
97.7?
1.63?
8.4?
1.09?
6.3
?
12
?
39.81
?
40.24?
0.43?
101.1?
2.27?
5.7
?
12
?
49.77?
49.97
?
0.20
?
100.4?
2.52?
5.0?
0.91?
2.0

 
20
Table 3
Precision, Bias, and Matrix Recovery for Nitrite
Matrix
?
# of?
True?
Mean
?
Bias vs
?
Recovery?
Interiab?
Interlab?
Single
?
Analyst
Values
?
Value?
Result
?
True?
vs True
?
Std Dev?
%RSD?
Operator?
%RSD
Value?
Value?
S(t)?
Std Dev,
S(o)
Reagent?
9?
0.50?
0.62?
0.12?
124.0
?
0.16?
26.1
Water
?
9?
0.70?
0.72?
0.02
?
102.9
?
0.08?
10.5?
0.05?
7.1
?
10
?
2.00?
1.31?
-0.69?
65.5?
0.25?
19.2
?
10?
2.98?
3.11?
0.13?
104.4?
0.17?
5.4
?
0.13?
6.0
?
10
?
14.86
?
14.70
?
-0.16?
98.9?
0.47?
3.2
?
10
?
19.81?
19.88?
0.07?
100.4
?
0.70?
3.5?
0.27?
1.5
?
10?
39.61
?
39.90?
0.29?
100.7
?
0.88?
2.2
?
10?
49.52?
48.24
?
-1.28?
97.4?
1.34?
2.8
?
1.25?
2.8
Substitute?
9?
0.50?
0.37?
-0.13?
74.0?
0.22?
59.7
Wastewater
?
9?
0.70?
0.59?
-0.11?
84.3?
0.28?
48.1?
0.21?
43.2
?
10?
2.00?
1.25?
-0.75?
62.5?
0.38?
30.8
?
9?2.98?
2.62?
-0.36?
87.9?
0.82?
31.4?
0.43
?
22.1
?
9?
14.86?
14.40
?
-0.46?
96.9?
0.58?
4.0
?
10?
19.81
?
19.50?
-0.31?
98.4?
1.66?
8.5
?
0.81?4.8
?
10?
39.61
?
39.97?
0.36?
100.9
?
2.02
?
5.0
?
9?
49.52?
49.09?
-0.43
?
99.1?
3.03?
6.2
?
2.11?4.7
Drinking
?
11?
0.50
?
0.52
?
0.02?
104.0
?
0.08?
14.4
Water?
12?
0.70?
0.74?
0.04?
105.7?
0.17?
23.3
?
0.09?
13.5
?
12
?
2.00?
1.30
?
-0.70
?
65.0?
0.21?
15.9
?
12?
2.98?
2.97?
-0.01?
99.7?
0.14?
4.6
?
0.16?
7.4
?
11?
14.86?
14.60
?
-0.26?
98.3?
0.40?
2.8
?
11?
19.81
?
19.82?
0.01?
100.1
?
0.59?
3.0
?
0.26?
1.5
?
11?
39.61
?
39.35
?
-0.26?
99.3?
0.99?
2.5
?
12?
49.52
?
49.14?
-0.38?
99.2?
1.93
?
3.9?
0.64?
1.5
"Real"
?
9?0.50?
0.55?
0.05?
110.0?
0.13?
24.5
Wastewater
?
10
?0.70?
0.73
?
0.03
?
104.3?
0.24?
32.9?
0.07?
10.8
?
9?
2.00?
1.27?
-0.73?
63.5?
0.18?
14.2
?
10?
2.98?
2.99?
0.01
?
100.3?
0.19?
6.2
?
0.15?
7.0
?
10
?
14.86?
14.55?
-0.31?
97.9?
0.46?
3.1
?
10?
19.81
?
19.68
?
-0.13?
99.3?
0.71?3.6
?
0.38?
2.2
?
9?
39.61
?
39.21
?
-0.40?
99.0?
1.03?
2.6
?
9?
49.52?
47.27
?
-2.25?
95.5?
3.50?
7.4
?
2.40?
5.6

 
21
Table 4
Precision, Bias, and Matrix Hecovery for Sulfate
Matrix?
# of
?
True?
Mean
?
Bias vs
?
Recovery?
Interlab
?
Interlab?
Single?
Analyst
Values
?
Value?
Result?
True
?
vs True
?
Std Dev
?
%RSD
?
Operator?
%RSD
Value?
Value
?
S(t)?
Std Dev,
S(o)
Reagent
?
9
?
0.49?
0.49?
0.00
?
100.0
?
0.18?
37.5
Water?
10?
0.70?
0.71?
0.01?
101.4
?
0.20?
29.2?
0.05?
8.3
?
10?
1.98?
2.04?
0.06?
103.0
?
0.19?
9.7
?
10?
2.98?
3.09?
0.11?
103.7
?
0.24?
7.9?
0.06?
2.5
?
10
?
14.86
?
14.67?
-0.19?
98.7?
0.57?
4.0
?
10
?
19.81?
19.67?
-0.14?
99.3
?
0.73?
3.8?
0.44
?
2.6
?
10?
39.60?
39.66?
0.06?
100.2
?
0.92?
2.4
?
10
?
49.51?
49.27
?
-0.24
?
99.5
?
1.26?
2.6?
0.49?
1.1
Substitute
?
9?
0.49?
0.38
?
-0.11?
77.6?
0.25?
66.9
Wastewater?
9?
0.70?
0.51?
-0.19?
72.9?
0.08?
16.4
?
0.18?
39.3
?
9?
1.98?
1.83?
-0.15?
92.4?
0.29?
16.2
?
9?
2.98?
2.86?
-0.12?
96.0?
0.31?
11.2?
0.20?
8.6
9?14.86?
14.19
?
-0.67
?
95.5?
1.06?
7.7
?
9
?
19.81
?
19.23?
-0.58?
97.1?
0.97?
5.2
?
0.46?
2.8
?
9?39.60
?
38.45
?
-1.15?
97.1?
1.33?
3.6
?
9
?49.51
?
47.75
?
-1.76?
96.4?
1.43
?
3.1
?0.75?
1.8
Drinking
?
12
?
0.49?
0.41?
-0.08?
83.7?
0.21?
52.8
Water?
12
?
0.70
?
0.41?
-0.29?
58.6?
0.20?
50.3?
0.14
?
34.3
?
13
?1.98?
1.77?
-0.21?
89.4
?
0.53?
30.3
?
13?
2.98
?
2.68?
-0.30?
89.9?
0.42?
16.2?
0.27?
12.1
?
13?
14.86
?
14.25
?
-0.61?
95.9?
1.11?
8.0
?
12?
19.81
?
19.31?
-0.50?
97.5?
1.39?
7.4
?
1.48?
8.9
?
12?
39.60?
38.58
?
-1.02?
97.4
?
1.96?
5.2
?
13?
49.51?
48.43?
-1.08?
97.8?
2.04?
4.3?
1.44
?
3.3
'Real"
?
10?
0.49?
0.37?
-0.12?
75.5?
0.39?
106.4
Wastewater
?
11?
0.70?
0.16
?
-0.54?
22.9?
1.19?
765.2
?
0.47?
179.6
?
11?
1.98?
1.57?
-0.41?
79.3?
0.87?
55.4
?
11
?
2.98
?
2.53?
-0.45?
84.9?
0.64?
25.4?
0.24?
11.9
?
11
?
14.86
?
14.69?
-0.17?
98.9?
1.26
?
8.6
?
10
?
19.81
?
19.38
?
-0.43?
97.8?
0.90?
4.6?
0.57?
3.4
?
11
?
39.60?
38.74?
-0.86?
97.8?
1.71
?
4.4
?
10?
49.51
?
48.36
?
-1.15?
97.7?
1.51?3.1?0.47?
1.1

 
22
Table 5
Precision, Bias, and Matrix Recovery for Nitrate
Matrix?
# of?
True?
Mean
?
Bias vs
?
Recovery
?
interlab
?
Interlab
?
Single?
Analyst
Values?
Value
?
Result?
True?
vs True?
Std Dev?
%RSD?
Operator
?
%RSD
?
Value
?
Value
?
S(t)
?
Std Dev, S(o)
Reagent?
10?
0.50?
1.02?
0.52?
204.00?
0.08?
7.4
Water
?
10
?
0.69?
0.71?
0.02?
102.90?
0.08?
11.6?
0.06?
6.4
?
11
?1.99?
2.83
?
0.84?
142.21
?
0.23?
8.1
?
11?
2.97
?
2.89?
-0.08?
97.31
?
0.18?
6.4?
0.14?
5.0
?
11?
14.91?
14.77
?
-0.14
?
99.06
?
0.44?
3.0
?
11
?
19.18?
19.77?
0.59?
103.08?
0.64
?
3.2
?
0.24?
1.4
?
10?
39.86
?
39.09
?
-0.77?
98.07
?
1.43?
3.7
?
10?
49.77
?
48.93
?
-0.84?
98.31
?
1.72?
3.5?
0.62?
1.4
Substitute?
11?
0.50?
1.18?
0.68?
236.00?
0.41?
34.9
Wastewater?
10?
0.69?
0.55?
-0.14
?
79.71
?
0.30?
55.3?
0.42?
4.9
?
10?
1.99
?
2.70?
0.71?
135.68?
0.42?
15.4
?
10
?
2.97?
2.33
?
-0.64
?
78.45
?
1.10?
47.3?
0.39?
15.4
?
9
?
14.91
?
14.29
?
-0.62?
95.84?
0.78?
5.4
?
10
?
19.18
?
18.69
?
-0.49?
97.45?
1.46?
7.8?
025?
1.5
?
11?
39.86
?
37.70?
-2.16?
94.58?
1.93
?
5.1
?
11?
49.77?
47.78
?
-1.99?
96.00?
2.18?
4.6
?
1.62?
3.8
Drinking?
11?
0.50
?
1.06 .?
0.56?
212.00?
0.19?
18.1
Water
?
11?
0.69?
0.65?
-0.04?
94.20?
0.06?
8.7
?
0.12?
14.4
?
12?
1.99?
3.05?
1.06?
153.27?
0.39?
12.8
?
11?
2.97?
3.01?
0.04?
101.35?
0.22?
7.2
?
0.33
?
10.8
?
12
?
14.91
?
14.69
?
-0.22?
98.52?
0.62?
4.2
?
12?
19.18
?
20.05
?
0.87
?
104.54?
0.88?
4.4
?
0.46?
2.7
?
12?
39.86
?
39.31
?
-0.55?
98.62?
1.67?
4.3
?
12?
49.77?
48.93
?
-0.84?
98.31?
1.43?
2.9
?
0.78?
1.8
"Rear
?
11?
0.50
?
0.94?
0.44?
188.00?
0.80?
84.7
Wastewater
?
10?
0.69?
0.69?
0.00?
100.00?
0.09?
13.3?
0.39?
47.6
?
10
?
1.99?
3.00?
1.01?
150.75?
0.38?
12.7
?
10
?
2.97?
3.01?
0.04?
101.35?
0.20?
6.6?
0.23?
7.8
?
11?
14.91
?
14.52
?
-0.39?
97.38?
0.66
?
4.6
?
11?
19.18
?
19.26
?
0.08
?
100.42?
0.77?
4.0?
0.77?
4.6
?
11?
39.86
?
39.13?
-0.73?
98.17?
1.78
?
4.6
?
11?
49.77?
49.17
?
-0.60
?
98.79?
2.26?
4.6
?
0.93?
2.1

 
23
Table 6
Precision, Bias, andNigi1371<
ecovery for Fluoride
Matrix?
# of?
True
?
Mean
?
Bias vs
?
Recovery?
Interlab
?
Intertab?
Single
?
Analyst
Values?
Value
?
Result
?
True
?
vs True
?
Std Dev?
%RSD?
Operator?
%RSD
?
Value?
Value
?
S(t)?
Std Dev, S(o)
Reagent
?
10?
0.50?
0.51?
0.01?
102.00
?
11.00?
11.4
Water
?
10?
0.71?
0.73
?
0.02?
102.82
?
7.90?
8.1?
0.02?
2.9
?
10
?2.00?
2.05?
0.05
?
102.50
?
3.60?
3.7
?
10?
3.00?
2.96?
-0.04?
98.67
?
4.40?
4.6
?0.09?
3.4
?
10?
6.99?
7.02?
0.03?
100.43?
5.40?
5.6
?
10?
9.99?
9.79?
-0.20?
98.00
?
4.60?
4.8?0.13?
1.6
?
10
?
19.98
?
19.60
?
-0.38?
98.10?
3.80?
3.9
?
10?
24.99
?
24.51
?
-0.48?
98.08?
4.80?
4.9
?0.74?
3.4
Substitute?
10
?0.50?
0.50?
0.00?
100.00?
0.09?
18.0
Wastewater
?
10?
0.71?
0.71?
0.00?
100.00?
0.09?
12.0?
0.01
?
2.3
?
10?
2.00?
1.98?
-0.02?
99.00
?
0.12?
6.0
?
10?3.00?
2.94?
-0.06?
98.00?
0.10?
3.4?
0.06?
2.6
?
10?
6.99?
6.92?
-0.07?
99.00
?
0.28?
4.1
?
9?
9.99
?
9.94?
-0.05?
99.50?
0.46?
4.7?
0.28?
3.3
?
10?
19.98
?
19.67?
-0.31?
98.45
?
0.94?
4.8
?
10?
24.99
?
24.78?
-0.21?
99.16?
1.09?
4.4?
0.63?
2.8
Drinking?
13?
0.50
?
0.48?
-0.02?
96.00
?
0.06?
12.9
Water
?
13
?0.71?
0.68?
-0.03?
95.77?
0.06?
9.5?
0.02?
3.4
?
13?
2.00?
1.96?
-0.04?
98.00
?
0.08?
3.9
?
13?
3.00?
2.90?
-0.10?
96.67
?
0.10?
3.4
?
0.08?
3.5
?
13
?6.99?
6.91?
-0.08?
98.86?
0.25?
3.6
?
13?
9.99?
9.91?
-0.08?
99.20
?
0.37?
3.7?
0.18?
2.2
?
13
?
19.98?
19.94
?
-0.04?
99.80?
0.68?
3.4
?
12?
24.99
?
24.27?
-0.72?
97.12?
1.63?
6.7
?1.30?
5.9
"Rear
?
11
?0.50?
0.47?
-0.03?
94.00?
0.08?
16.9
Wastewater
?
11
?
0.71
?
0.68?
-0.03?
95.77
?
0.08?
11.7?
0.04?
7.6
?
11
?2.00?
1.96?
-0.04
?
98.00?
0.12?
6.3
?
11
?
3.00
?
2.93?
-0.07?
97.67?
0.18?
6.2
?
0.09?
3.5
?
11
?
6.99?
6.85
?
-0.14
?
98.00
?
0.26?
3.8
?
10?
9.99?
9.56?
-0.43?
95.70?
0.73?
7.7?
0.44?
5.3
?
11?
19.98?
20.06
?
0.08?
100.40
?
1.23?
6.1
?
11
?
24.99?
25.12?
0.13?
100.52?
1.34?
5.3?
0.32?
1.4

 
24
Table 7
Precision, Bias, and Matrix Recovery for o-Phosphate
Matrix
?
# of?
True
?
Mean?
Bias vs
?
Recovery?
Interlab?
Interlab
?
Single
?
Analyst
Values?
Value?
Result
?
True
?
vs True?
Std Dev?
%RSD?
Operator
?
%RSD
?
Value?
Value?
S(t)?
Std Dev, S(o)
Reagent
?
10
?
0.50?
0.41?
-0.09?
82.00
?
0.12
?
29.6
Water?
9?
0.69
?
0.51?
-0.18?
73.91
?
0.13?
26.6?
0.03?
7.2
10?
1.99?
1.88?
-0.11?
94.47?
0.16?
8.3
10?
2.98?
2.76?
-0.22?
92.62
?
0.14
?
4.9?
0.08?
3.2
10?
14.86?
14.93
?
0.07?
100.47?
0.64?
4.3
9
?
19.80?
19.76
?
-0.04?
99.80?
1.00?
5.1
?
0.85?
4.9
10?
39.60?
39.79
?
0.19?
100.48?
1.38?
3.5
10?
49.51
?
50.10?
0.59?
101.19
?
1.76?
3.5?
0.72?
1.6
Substitute
?
11
?
0.50?
0.49?
-0.01?
98.00?
0.15?
30.0
Wastewater
?
10
?
0.69
?
0.59?
-0.10?
85.51?
0.17?
28.8?
0.13?
24.4
11?
1.99?
1.92?
-0.07?
96.48?
0.28?
14.6
10?
2.98?
2.89?
-0.09?
96.98
?
0.22?
7.6
?
0.18?
7.5
11?
14.86
?
15.31
?
0.45?
103.03
?
1.74?
11.4
11
?
19.80?
19.78
?
-0.02?
99.90
?
1.16?
5.9?
0.84?
4.8
11
?
39.60?
39.58?
-0.02?
99.95
?
2.72?
6.9
11?
49.51
?
49.19?
-0.32?
99.35
?
3.98?
8.1
?
2.18?
4.9
Drinking
?
12?
0.50?
0.46?
-0.04?
92.00
?
0.14?
30.0
Water?
13?
0.69
?
0.55
?
-0.14?
79.71
?
0.20?
36.3?
0.07?
13.4
13
?
1.99?
1.89?
-0.10?
94.97?
0.22?
11.9
13?
2.98
?
2.87?
-0.11?
96.31?
0.24?
8.5
?
0.07?
2.8
12
?
14.86?
15.09
?
0.23?
101.55?
0.91?
6.1
13
?
19.80
?
20.28
?
0.48?
102.42?
0.96?
4.7?
1.06?
6.0
13?
39.60
?
40.37?
0.77?
101.94?
2.15?
5.3
13
?
49.51
?
50.75?
1.24?
102.50?
3.14?62?
1.03?
2.3
'Real°
?
11
?0.50?
0.43?
-0.07?
86.00?
0.17?
39.1
Wastewater?
11
?0.69?
0.53
?
-0.16?
76.81?
0.24?
46.5?
0.12?
25.8
11?
1.99?
1.72?
-0.27?
86.43
?
0.27?
15.8
11?
2.98?
2.52?
-0.46?
84.56
?
0.48?
19.2?
0.30?
14.0
11
?
14.86
?
14.93?
0.07?
100.47
?
0.91?
6.1
11?
19.80?
19.90?
0.10?
100.51?
1.35?
6.8
?
0.91?
5.2
11?
39.60
?
38.98?
-0.62
?
98.43?
1.45?
3.7
10?
49.51
?
48.26?
-1.25?
97.48?
1.80?
3.7?
0.82?
1.9

 
25
Table 8
QC Acceptance Criteria
Analyte
Matrix
Precision
% RSD
Average
%Recovery
Initial
LL - UL
Ongoing
LL - UL
MS/MSD
LL - UL
MS/MSD
RPD
Chloride
RW
6.30
98.5
90.8 - 106.2
88.7 - 108.3
89.4 - 107.5
12.0
DW
10.00
97.0
84.0 - 110.0
81.1 - 113.0
81.9 - 112.5
18.6
WW
7.00
92.8
83.0 -102.6
81.4 - 104.2
81.8 - 103.8
13.2
Bromide
RW
10.10
99.7
92.2 - 107.2
86.7 - 112.7
88.5 - 111.0
19.2
DW
12.70
95.8
85.9 -105.6
79.8 - 111.8
81.8 - 109.8
23.3
WW
14.40
99.2
87.2 - 111.2
80.1 -118.3
82.4 - 116.0
26.9
Nitrite
RW
6.40
100.6
95.1 - 106.0
91.9 - 109.2
92.8 - 108.3
12.1
DW
4.30
99.6
92.4 - 106.7
91.5 - 107.7
91.8 - 107.4
8.1
WW
4.90
98.9
91.3 - 106.5
90.2 - 107.6
90.5 - 107.3
9.2
Sulfate
RW
9.40
100.4
90.9 -109.9
86.9 - 113.9
88.2 - 112.6
17.9
DW
16.1
95.6
82.6 - 108.6
74.9 - 116.2
77.5 - 113.7
29.7
WW
19.60
95.3
78.9 - 111.7
70.1 - 120.5
72.6 - 118.0
36.9
Nitrate
RW
8.40
99.5
93.1 - 105.9
88.6 - 110.4
90.0 - 108.9
15.9
DW
9.40
100.2
93.0 -107.4
88.0 -112.4
89.4 - 111.0
17.4
WW
6.70
99.1
90.7 - 107.6
88.6 -109.7
89.2 - 109.1
12.4
Fluoride
RW
7.90
99.5
92.2 - 106.7
88.7 - 110.3
89.8 - 109.1
14.9
DW
4.86
98.3
91.9 - 104.8
90.5 - 106.2
90.9 - 105.7
9.0
WW
7.90
98.5
90.0 - 107.1
87.0 - 110.1
88.0 - 109.1
14.7
Phosphate
RW
10.60
98.2
91.9 - 104.5
85.4 - 111.0
87.4 - 109.0
20.1
DW
9.40
100.2
89.3 - 111.1
85.8 - 114.6
87.0 - 113.4
17.4
WW
16.90
94.6
81.5 - 107.7
73.5 - 115.8
76.1 - 113.1
31.5
All data determined as spike recovery from ASTM method validation and EPA Tier 3 Criteria
Reagent water (RW) data between 0.5 and 50 mg/L, except Fluoride 0.5 and 25 mg/L
consisting of 4 Youden Pairs
Drinking (DW) and Wastewater (WW) data between 2 and 50 mg/L except Fluoride 2 and 25 mg/L
consisting of 3 Youden Pairs
RSD = %Relative Standard Deviation; (std dev / mean)(100)
LL = Lower Limit of %Recovery
UL = Upper Limit of %Recovery
RPD = Relative % Difference between MSD

 
26
Appendix B
(Non-mandatory Information)
B.1 Suggested Background References
B1.1 EPA Method 6500, "Dissolved Inorganic Anions in Aqueous Matrices by
Capillary Ion Electrophoresis", SW846, Rev 0, January 1998.
B1.2 Method 4140, "Inorganic Anions by Capillary Ion Electrophoresis", Standard
Methods for the Examination of Water and Wastewater, 20 Edition, 1998, p 4-12
to 4-20.
B1.3 Krol, Benvenuti, and Romano, "Ion Analysis Methods for IC and CIA and
Practical Aspects of Capillary Ion Analysis Theory", Waters Corp, Lit Code WT-139,
1998.
B1.4 Jandik, P., Bonn, G., "Capillary Electrophoresis of Small Molecules and Ions",
VCH Publishers, 1993
B1.5 Romano, J., Krol, J, "Capillary Ion Electrophoresis, An Environmental Method
for the Determination of Anions in Water", J. of Chromatography, Vol. 640, 1993, p.
403.
B1.6 Romano, J., "Capillary Ion Analysis: A Method for Determining Ions in Water
and Solid Waste Leachates", Amer. Lab., May 1993, p. 48.
B1.7 Jones, W., "Method Development Approaches for Ion Electrophoresis", J. of
Chromatography, Vol. 640, 1993, p. 387.
B1.8 Jones, W., Jandik, P., "Various Approaches to Analysis of Difficult Sample
Matrices for Anions using Capillary Electrophoresis", J. of Chromatography, Vol.
608, 1992, p. 385.
B1.9 Bondoux, G., Jandik, P., Jones, W., "New Approaches to the Analysis of Low
Level of Anions in Water", J. of Chromatography, Vol. 602, 1992, p. 79.
B1.10 Jandik, P., Jones, W., Weston, A., Brown, P.,"Electrophoretic Capillary Ion
Analysis: Origins, Principles, and Applications", LC•GC, Vol. 9, Number 9, 1991, p.
634.
B1.11 Romano, J., Jackson, P., "Optimization of Inorganic Capillary Electrophoresis
for the Analysis of Anionic Solutes in Real Samples", J.
of
Chromatography, Vol.
546, 1991, p. 411.
B1.12 Jandik, P., Jones, W., "Optimization of Detection Sensitivity in the Capillary
Electrophoresis of Inorganic Anions", J
of
Chromatography, Vol. 546, 1991, p. 431.
B1.13 Jandik, P., Jones, W., "Controlled Changes of Selectivity in the Separation
of
Ions by Capillary Electrophoresis", J. of Chromatography, Vol. 546, 1991, p 445.
B1.14 Foret, R., et.al., "Indirect Photometric Detection in Capillary Zone
Electrophoresis", J. of Chromatography, Vol. 470, 1989, p. 299.
B1.15 Hjerte'n, S. et. al., "Carrier-free Zone Electrophoresis, Displacement
Electrophoresis and Isoelectric Focusing in an Electrophoresis Apparatus", J. of
Chromatography, Vol. 403, 1987, p. 47.
B1.16 Serwe, M., "New ASTM Standard: Recommended Operating Conditions for the
Agilent CE", Agilent Technologies Application Brief, Publication Number 5968-
8660E.

 
27
Appendix C
Capillary Ion Electrophoresis
Initial Demonstration of Performance
Single Operator
General Inorganic Anion
&
Organic Acid Analysis with Indirect UV Detection
Basis for EPA Method 6500, ASTM D6508, and Standard Methods 4140
The performance data given in this appendix was provided in the collaborative instruction
booklet to evaluate initial demonstration of performance required by the collaborative design.
E
ce)
2
4
5
6
7
11
PPM Standards
1
2
3
4
5
6
7
8
9
10
11
Chloride
Bromide
Nitrite
Sulfate
Nitrate
Oxalate
Fluoride
Formate
Phosphate
Bicarbonate
Acetate
= 2
= 4
= 4
= 4
= 4
= 5
= 1
= 5
= 4
= 5
10
■.)
I?
'
?
I
? I
3.000
?
3.500
Minutes
?
4.000
?
4.500
Analysis Conditions:
Electrolyte:
Capillary:
Temperature:
Power Supply:
Voltage:
Current:
Sampling:
Detection:
Time Constant:
Sampling Rate:
Analyte MT:
Quantitation:
lonSelect High Mobility Anion Electrolyte, P/N 49385
75 gm (id) x 375 gm (od) x 60 cm (length)
25°C (5°C Above Ambient)
Negative
15 kV
14 ± 1 pA (Use Constant Current for Analysis)
Hydrostatic for 30 Seconds
Indirect UV at 254 nm, Hg Lamp, 185 or 254 nm Window
0.3 Seconds, or less
20 Data Points per Second
Mid-Point of Analyte Peak Width at Baseline
Time Corrected Peak Area (Peak Area / MT)

 
Millennium Data Processing Method:
28
CIE Processing
Integration
Calibration
Method using Mid-Point of Peak Width for Migration Time
Peak Width = 2.25 - 3.00 Threshold = 100 ± 25
Min Area = 100
?
Min Height = 50
Inhibit lntg. = 0 to 3 min
Averaging?
= None?MT Window = 2%
Update MT?
= Average Standards
Peak Match
?
= First for Chloride
(CI is always first in the pherogram, use as a ref peak)
CI MT Window = 10%
Other Analytes = Closest
Quantitate By = Time Corrected Peak Area
Fit Type
?
= Linear Through Zero
Report
Analyte Name
Analyte Migration Time
Analyte Migration Time Ratio (respect to CI Ref Peak)
Peak Area
Time Corrected Peak Area
Amounts
Use fresh electrolyte daily; recalibrate with every change in electrolyte.
Clear previous calibration (in Quick Set Page) before recalibration.
Do Not use analyte peak height for quantitation due to asymmetrical peak shapes.
Method Validation:
The single operator performance given below using the ASTM validation design is intended as
a basis to evaluate Initial Demonstration of Performance.
Individual Youden Pair Standard, in ppm
12345678
CI
0.7
2.0
3.0
15.0
40.0
20.0
50.0
0.5
Br
2.0
3.0
15.0
40.0
20.0
50.0
0.7
0.5
NO2
3.0
40.0
20.0
15.0
50.0
0.5
2.0
0.7
SO4
40.0
50.0
0.5
0.7
2.0
3.0
15.0
20.0
NOs
15.0
20.0
40.0
50.0
0.5
0.7
2.0
3.0
F
2.0
0.7
0.5
3.0
10.0
7.0
20.0
25.0
PO4
50.0
40.0
20.0
0.5
3.0
2.0
0.7
15.0
O
5,
C

 
CI
SO4
CI?
R2= 0.9996
SO. R 2 = 0.9998
Br?
R2= 0.9995
Br
3 Data Points per Concentration
Using Validation Standards
Method Linearity:
10
7
.5
r.n
5
.c0
2.5
NO a
3 Data Points per Concentration
Using Validation Standards
0
0
20
ppm Anion
30
40
50
NO 3
R2
= 0.9992
29
0
?
10
?
20?
30
?
40
?
50
ppm Anion
10 —
F?
R2= 0.9985
?
F
PO4
PO4?
R2= 0.9996
7.5 —
2.5
3 Data Points per Concentration
Using Validation Standards
0
?
0
?
10
?
2 Op
pm Anion
0
?
40
?
50
10 —
?
7.5 —
?
NO 2
?
R2= 0.9996

 
SO4
CI
PO4
30
Method Detection Limits:
100 ppb Anion Standard
E
I
?
I?
I
3.200
?
3.400
?
3.600
?
3.800?
4.000
M in ute s
Seven replicates of the above 100 ppb anion standard were used to calculate time corrected
peak area precision. Using EPA and Standard Methods protocols, the detection limits, as
ppb, for these analytes are:
Chloride = 46
Nitrate = 84
Bromide = 90
Fluoride = 20
Nitrite = 72
Phosphate = 41
Sulfate = 32
This method has been validated between 0.1 to 50 ppm. Quantitation below 0.1 ppm is not
advised.

 
31
Migration Time Reproducibility:
Use mid-point of analyte peak width at the baseline as the analyte migration time determinant.
Data given as average absolute migration time for each validation standard analyzed in
triplicate.
Analyte
CI
Br
NO2
SO4
NO3
F
PO4
1
3.132
3.226
3.275
3.405
3.502
3161
3.906
2
3.147
3.239
3.298
3.431
3.517
3.779
3.931
as
03
-0
3.138
3.231
3.283
3.411
3.497
3.771
3.925
cco
?
4
3.158
3.244
3.307
3.434
3.510
3.781
3.963
u)co?
5
3.184
3.271
3.331
3.435
3.551
3.787
3.981
fs
6
3.171
3.260
3.312
3.418
3.537
3.776
3.964
Vs'
>
?
7.
3.191
3.272
3.315
3.437
3.544
3.773
3.978
8
3.152
3.248
3.294
3.418
3.526
3.739
3.954
Std Dev
0.021
0.015
0.018
0.012
0.20
0.015
0.027
%RSD
0.67%
0.46%
0.55%
0.36%
0.56%
0.40%
0.68%
Average Standard Deviation = 0.018 min = 1.1 sec
Average %RSD of Analyte Migration Time = 0.53%
Quantitation Precision:
Time Corrected Peak Area Precision, given as %RSD, based upon 3 samplings per
concentration.
Analyte
CI
Br
NO2
SO4
NO3
F
PO4
0.1
12.36
18.89
16.19
13.25
23.13
9.82
14.00
0.5
10.51
20.00
3.90
2.25
2.18
2.03
7.71
c
....m.?
0.7
1.23
13.36
2.01
2.95
0.37
2.72
4.41
Li
a'
0.32
3.76
4.14
1.79
2.17
0.73
1.91
0
c
3
Q
0.63
1.80
1.72
1.70
0.58
0.98
2.70
0
L 15
0.43
0.27
0.48
0.07
0.36
0.15
1.37
0...
20
0.45
0.66
0.17
0.13
0.88
0.16
0.81
40
0.36
0.56
0.36
0.46
0.58
0.47
50
0.45
0.51
0.48
0.16
0.46
0.46

 
32
Quantitation Accuracy:
Used a Certified Performance Evaluation Standard diluted 1:100 with DI water. Amounts
based upon multi-point calibration curve prepared from certified standards.
Analyte
CI
NO2
SO4
NO3
F
PO4
Performance
True
Evaluation
Standard
Value
in ppm
43.00
1.77
37.20
15.37
2.69
6.29
Official
Measured
43.30
1.77
37.00
15.42
2.75
6.38
Anion
Mean
Methods
Measured
3.09
0.07
2.24
1.15
0.26
0.21
Wet Chem & IC
Std
Dev
CIA Using
Chromate
Ave CIA
n=18
43.34
1.64
37.11
14.41
2.64
6.34
Electrolyte
CIA/Mean
1.003
0.927
1.003
0.935
0.959
0.993
CIA/TrueValue
1.008
0.927
0.996
0.938
0.981
1.008
A CIA/True Value, or Mean = 1.000 indicates perfect agreement between CIA and official
anion methods.
Method Recovery:
A Certified Performance Evaluation Standard (PES) was diluted 1:100 with Typical Drinking
Water (DW). Amounts based upon multi-point calibration curve prepared from certified
standards.
Analyte
CI
NO2
SO4
NO3
F
PO4
Drinking Water
n=3, as ppm
24.72 ±
0.18
Not
Detected
7.99 ±
0.07
0.36 +
0.05
Not
Detected
Not
Detected
Amount %RSD
0.73%
0.91%
13.3%
Performance
Evaluation Std
43.00
1.77
37.20
15.37
2.69
6.29
DW + PES
n=3; as ppm
66.57±
0.34
1.74±
0.03
45.19±
0.17
15.42±
0.12
2.62±
0.07
5.55±
0.31
Amount %RSD
0.51%
1.85%
0.38%
0.79%
2.69%
5.52%
% Recovery
97.9%
98.3%
100.2%
98.1%
97.4%
88.2%

 
1
2
Li
L.
Anions in mg/L. No Dlution
1 Chloride = 93.3
2 Nitrite = 0.46
3 Sulfate =
60.3
4 Nitrate =
498
5 Carbonate = Natural
•'?
I?
•?"?
•?
3 ODD?
3'500
Minutes
4.000
'
•?
I?
'
4.500
'
•?
•?
?
•?"?
I
?
Minutes
4.000?
4.500
• I?
3.500
33
Fig. 1 Electropherogram of Mixed Anion Working Solution
and Added Common Organic Acids
Anion Standard in me&
1 Chloride = 2 7
Fluoride
/
4
?
I
2 Bromide = 4 8
Formate = 5
3 Mlle =4 9
Phosphate = 4
8?
4 Sulfate = 4 10
Carbonate
5 Nitrate = 4 11
Acetate = 5
9?
11
Fig. 2 Electropherogram of 0.2 mg/L Anions
Used to Determine MDL
6
1 Chloride
5
Nitrate
2 Brorride
6
Fluoride
3 Nitrite
7 Phosphate
4 Sulfate
4
3
5
3
E
4
C0.
0
7
2
\ON
,/ \A-vkilL
\41,-..,,Av■Yetv
I • " •?
3 000?
3.500?
4.000?
4.500
Minutes
1?1?1?1?
1?1?
1?1
?
I?I?
1?1?
f
?
i
3
030
3.200
3.400 3.600
3.800
Minutes
0)
Fig. 3 Electropherogram of Substitute Wastewater
fiknions n moil–
120 Dilution
1 Chloride =24.2
2 Sulfate = 3.77
3 Phosphate = 0.89
4 Carbonate = Nahral
2
3
Fig. 4 Electropherogram of Drinking Water
Anions In moil-, No Didion
1 Chloride = 20.2
2 Sulfate = 7.5
3 Nitrate =
1.6
4 RtnriCe
=
ace
5 Carbonate = Natural
3
3.000
11
?1?1
3.500
?
4/.003
Knutes
3.000
• I
?
•?
•?
3.500?
Minlite3
• I?•
4.003
'?
I?
4.500
Fig. 5 Electropherogram of Municipal Wastewater
?
Fig. 6 Electropherogram of Industrial Wastewater
Treatment Plant Discharge
3
E
0
Anions
in
moil_ Dilution
1 Chloride = 2.0
2 Nitrite =
1.6
3 Sulfate =34.7
4 Nitrate =
1&5
5 Formate < 0.05
6 Phosphate = 12.3
7 Carbonate = Nahral
4
L.
3.000

 
?
INorpmc
?
avnlent .
ANC,
:'5
Or
.
Adds,
?
' -
Ci; Br„
:
r
?
Oxyrn.r.?tals
?
NO:,,
S
C4
?
'
'
F
,
PO4,
?
No., ,,l_?
CIO C..10.
?
SO ; S
20 3?
':FanTlate,
Migration Time
High Mobility?
Low Mobility MT > 7 mini,
Anions
?
Anions
.4
MT= 0
O.?
1'
0.
?
?
9'
0.
?
51.
0
Y-1 9' 9'
N+? N? N
.
?
O.
?
O_
?
OH
N?
N
Anode
Matyte Ion (A) displaces electrolyte ion (e)
charge for charge or-transfer ratio causing
a net decrease n battleground absotbance.
The charge in absorbance is directly
related to Am:1sta concentration.
75
pm x 60 en7"..'..\\
Slim Capillary
CapPiary with Polyirnide
Coating Removal,
CNI Window
UV Detector
at 254 nm
Vane= Purge
rE
lectrolyranism
Constant
Temperature
Compartment,
25.30° C
!ta
& Samples
ndard: •
High Voltage Power Supply 0 to -30 kV
34
Fig. 7 Pictorial Diagram of Anion Mobility and
?
Fig. 8 Selectivity Diagram of Anion Mobility Using
ElectroOsomotic Flow Modifier
?
Capillary Ion Electrophoresis
HO
Mobility
Anion Q --••■■•••••••÷
Low Mobility Anion?
Cicotnto
•iE
?
-
Nurtnnis
(
?
&
W4or
* )
1"›
Donn:lion
Side
N1.14'.1""t4
Cathodo
0
miscues
Sid.
Fig. 9 Pictorial Diagram of Indirect UV Detection
e eeeeeAA Aeeeeeeeee
e
eeeeAAA Aeeeeeeee
e eeeeeAAA Aeeeeeeee
e
eeeeAAA Aeeeeeeee
Fig. 10 General Hardware Schematic of a
Capillary Ion Electrophoresis System

Back to top