~b V
    CLr~PK’~
    CWFICE
    Acoustic Associates,
    Ltd.
    FEB
    102003
    iIIIIIIIllIIIIIIuIiuIIluuiuIIiuIiiiIiuiiiIuuiiIIIIu
    1111111
    S
    lATE OF
    ILLINO
    Specialists inHearing andAcoustics
    1-Oliutlon
    Control Board
    Thomas
    Thunder, MA, FAAA,
    INCE
    -
    Principal
    Roger Harmon, BSEE, PB
    -
    Acoustical Engineer
    Regina Dziewior, MA,
    CCC
    -
    Audiologist
    Peter Straus, AAS
    -
    Field Technician
    Illinois Pollution Control Board
    J.R. Thompson Center
    Suite 11-500
    -~
    100 West Randolph
    C
    Chicago, Illinois
    60601
    RE:
    Comments on the first-notice of the Noise RuleUpdate
    Proposed Amendments to
    35
    ILL. ADMCODE 900 and 903
    R03-8 NOISE RULEMAKING
    I appreciate the opportunity to comment on the subject rulemaking. Here are my comments:
    1.
    Add
    a
    definition for “reference time.”
    I suggest the following definition:
    “the
    time period over which the Leq average is measured or
    determined.” In 1987, the Board amended 900.103 to require that noise measurements be based on
    Leq averaging andthat the referencetime forthis averaging be 1-hour. If the Board intended this to
    be a
    1-hour measurement time, then with some sound level meters, measurements for all nine octave
    bands could take nine hours. By definingreference time as suggested here, the 1-hour Leq could be
    projectedbased on shorter measurement periods whichwould greatly reduce the time required forthe
    vast majority of situations.
    For example, ifthe noise were steady (as defined inthe regulation), there would be no need to
    measure fora full hour. In such cases, a 1-minute Leq would yield the same results as a 1-hour Leq. If
    the noise were intermittent, then a measurementtaken while the source was
    on would be sufficient
    to projecta 1-hour Leq when the on/offproportion (duty cycle) is known or assumed. Only in cases
    where the noise were non-steady or the source had no repetitive cycle, such as
    loading dock
    operations, would a full 1-hour measurementbe required.
    Finally, ifacomplainant was willing to assume ~
    contribution from the noise source for the
    unmeasured duration of the 1-hour reference time, the Boardshould not insist on afull 60 minutes of
    measurement to support anumerical violation of the regulation, For example, if 30 minutes of
    representative sampling was all that a complainant could obtain, for whatever reason, then the
    projected 1-hour Leq with ~
    assumed contribution from the sourcefor the remaining 30 minutes
    might still exceed the State’s limits. In this regard, the definition suggested above for “reference time”
    should be amended with: “For purposes of showing
    noncompliance with these regulations, 0
    dB
    of contribution can
    be assumed for
    the unmeasured portion of the reference time.”
    305
    E. Northwest Highway, Palatine, illinois
    60067
    Office:
    847-359-1068
    Fax:
    847-359-1207
    Website:
    wvw.AcousticAssociates.corn
    E-mail:
    tthunder@concentric.net
    February 7, 2003
    Dear Board Members:
    Acoustic Associates, Ltd.
    Page
    1

    Illinois Pollution Control Board
    Noise Rulemaking
    February 7,2003
    2.
    The American National Standard
    ANSI S12.9-l993fPart 3, “Short-Term Measurements
    with an
    Observer Present” should be added as a referencestandard under Section 900.101.
    ANSI
    S
    12.9-1993/Part
    1
    hasalready been added in the Notice.’ However, since this standard only
    deals with definitions for standard quantities, I recommend its companion standard,
    Part
    3, also be
    added sincethis is the most relevant American National Sthndard dealing with the type of
    measurements and procedures related to this Regulation.
    3.
    In the definition for
    “Background
    Sound Level”
    delete “A-weighted sound.”
    The regulation is written to measure the octave band Leq, not the A-weightedLeq. Better yet, the
    definition should be made congruent with the definition under ANSI S 12.9. Specifically I recommend
    the following language: “all-encompassing sound associated with a given environment-without
    contributions from the source or sources ofinterest. When the background noise consists of discrete,
    intrusive events (e.g., dog barking, carhorns, train horns, etc.), then~either:
    1) the data from these
    events must be subtracted from the sample, or2) a statistical analysis must be performed to determine
    the level exceeded 90
    ofthetime.” The Board should note that ANSI S
    12.9/Part
    3, as referenced in
    item ~2, describes the actual task required to correct for the background noise.
    It may, therefore, wish
    to refer to this standard to addressthe issue ofbackground noise corrections.
    4.
    In the definition ofLeq,
    add “...the measure
    is a time-averaged sound level during
    a stated time-
    interval,
    T.”
    This addition is necessary to be consistent with currentANSI terminology per ANSI S 1.1.
    5.
    The definition “Period ofobservation” should be replaced with “Measurement period.”
    This change is consistent with current
    ANSI terminology per ANSI S 12.9. An observation period
    would be the time a field person observed source operations:andbackground noise conditions. But the
    measurement period refers to the actual sampling duration.-Furthermore,I recommend a-minimum
    measurement period of 10 minutes. As explained above, the 1-hour Leq could be reliably projected
    from a 10-minute measurement ifthe sound was steady
    and
    continuous.
    And
    ifthe sound was
    intermittent,
    but
    steady when the source was operating during the 10-minute measurement, the
    1-hour
    Leq could still be projected with a
    known
    or assumed on/off proportion. Only forfluctuating sound
    would a full 60-minute measurement period be needed. In any case, a complainant ought to be able to
    assume 0 dB contribution for the remaining 50 minutes ofthe reference time to show noncompliance
    with the numerical portions ofthe State’s limits. Of course, for a respondent to show compliance for a
    non-steady noise, a full 60-minute measurement duration would be required.
    Sincere~~~
    ~
    Thomas Thunder, AuD,
    FAAA
    Licensed Audiologist
    State ofIllinois
    Board Certified
    -
    Institute ofNoise Control Engineering
    cc: Howard Chinn
    AcousticAssociates, Ltd.
    Page 2

    Back to top