ILLINOIS POLLUTION CONTROL BOARD
    August 24, 2000
    IN THE MATTER OF:
    SAFE DRINKING WATER UPDATE, USEPA
    REGULATIONS
    (July 1, 1999, through December 31, 1999)
    )
    )
    )
    )
    )
    R00-10
    (Identical-in-Substance Rulemaking -
    Public Water Supplies)
    Adopted Rule. Final Order.
    OPINION AND ORDER OF THE BOARD (by R.C. Flemal):
    BACKGROUND
    Under Section 17.5 of the Environmental Protection Act (Act) (415 ILCS 5/17.5 (1998)),
    the Board today adopts amendments to the Illinois regulations that are “identical in substance”
    (IIS) to the National Primary Drinking Water regulations (NPDWRs) adopted by the United States
    Environmental Protection Agency (USEPA). These regulations implement sections 1412(b),
    1414(c), 1417(a), and 1445(a) of the Safe Drinking Water Act (SDWA), 42 U.S.C. §§ 300g-1(b),
    300g-3(c), 300g-6(a), and 300j-4(a)(1996). The nominal timeframe of this docket includes
    SDWA amendments that the USEPA adopted in the period from July 1, 1999, through December
    31, 1999. The federal SDWA regulations are found at 40 C.F.R. §§ 141 and 142. The USEPA
    took two actions during the nominal timeframe period that could necessitate Board action, but, as
    explained below, in response to public comment the Board is incorporating only one federal action
    into the Board’s rules. The final rule adopted today includes the addition of analytical methods
    approved for determining compliance with the drinking water regulations.
    Section 17.5 of the Act (415 ILCS 5/17.5 (1998)) provides for quick adoption of
    regulations that are “identical in substance” to federal regulations that the USEPA adopts to
    implement sections 1412(b), 1414(c), 1417(a), and 1445(a) of the SDWA. Section 17.5 of the Act
    also provides that Title VII of the Act (415 ILCS 5/1
    et seq
    . (1998)) and Section 5 of the
    Administrative Procedure Act (APA) (5 ILCS 100/5-35 & 5-40 (1998)) do not apply to the
    Board’s adoption of IIS regulations. Accordingly, this rulemaking is not subject to first or second-
    notice review by the Joint Committee on Administrative Rules (JCAR). However, to facilitate
    receipt of public comments, the Board’s June 8, 2000 proposal for public comment was published
    in the
    Illinois Register
    . 24 Ill. Reg. 8728 (June 30, 2000).
    Section 7.2(b) of the Act (415 ILCS 5/7.2 (1998)) requires the Board to complete its IIS rulemaking actions
    within one year of the date of the earliest federal amendments involved in a docket. In this docket, the earliest
    federal action requiring Board action occurred on September 17, 1999. Thus, the Board must complete its
    rulemaking activity in this docket prior to September 17, 2000.
    FEDERAL ACTIONS CONSIDERED IN THIS RULEMAKING

    2
    This rulemaking incorporates certain federal amendments that occurred during the period of July 1, 1999,
    through December 31, 1999. The SDWA regulations are found at 40 C.F.R. §§ 141-142. The USEPA amended the
    federal SDWA regulations two times during that period: September 17, 1999, and December 1, 1999. This
    rulemaking considers both of these actions. As explained below, only the December 1, 1999 action requires
    corresponding Board action.
    The federal amendments incorporated into this rulemaking are summarized as follows:
    Federal Action
    64 Fed. Reg. 50556
    (September 17, 1999)
    64 Fed. Reg. 67450
    (December 1, 1999)
    Summary
    Revisions to the Unregulated Contaminant Monitoring
    Regulation for Public Water.
    The USEPA adopted
    amendments requiring public water supplies to
    monitor for unregulated drinking water
    contaminants. The USEPA adopted the
    amendments under section 1445(a) of SDWA.
    National Primary and Secondary Drinking Water
    Regulations: Analytical Methods for Chemical and
    Microbiological Contaminants and Revisions to Laboratory
    Certification Requirements.
    The USEPA amended the
    rules setting forth the analytical methods
    approved for use in demonstrating compliance
    with the SDWA requirements.
    Public Comments Received
    The Board adopted a proposal for public comment in this matter by order dated June 8,
    2000. The “Notices of Proposed Amendments” appeared in the June 30, 2000 issue of the
    Illinois
    Register
    at 24 Ill. Reg. 8728. The Board accepted public comments on the proposal for a 45-day
    period following publication in the
    Illinois Register.
    The comment period expired on August 15,
    2000.
    The Board received four comments during the public comment (PC) period. Those
    comments are as follows:
    PC 1
    Identical First Notice Line Numbered Version of proposed amendments from JCAR
    (received June 29, 2000).
    PC 2
    July 24, 2000 memo from Deborah Connelly, JCAR (received July 25, 2000).
    PC 3
    August 9, 2000 letter from David S. Horak, Safe Drinking Water Branch, USEPA,
    Region 5 (received August 15, 2000).
    PC 4
    August 14, 2000 letter from Lou Allyn Byus, Assistant Manager - Field Operations,
    Division of Water Supplies, Illinois EPA (received August 18, 2000).

    3
    These public comments are discussed in more detail later in this opinion.
    Response to JCAR Additional Questions
    In addition to PC 1, the Board received public comment from JCAR in PC 2, which
    included two inquiries from JCAR. It is more efficient to address these questions here, rather than
    later. The first question addressed incorporating by reference certain sections of the unregulated
    contaminant monitoring regulations, as proposed by the Board. As the Board is not adopting the
    proposed program in response to Illinois Environmental Protection Agency (Agency) and USEPA
    comments that the proposed rules were not federally required to be adopted as IIS rules, the issue
    is no longer relevant.
    The second question in PC 2 addressed a section of the Board’s June 8, 2000 proposal for
    public comment, appearing at page 4. The section reads as follows:
    Board Amendments not Directly Federally Driven
    The Board is updating all C.F.R. citations in the affected sections of this rulemaking to the
    1999 version. The Board is also amending Section 611.310(a) and the Section 611.310(c)
    Board Note to correct any ambiguity in the regulations for the purpose of avoiding
    confusion to the regulated community. Section 611.310 is applicable to all public water
    systems. The Board is also making numerous technical changes suggested by JCAR, the
    USEPA, and the Agency.
    JCAR questioned the Board’s authority to make such changes. JCAR’s inquiry points out
    that the Board’s caption of this portion of the opinion is not precisely accurate, and is in need of
    revision.
    In every IIS rulemaking, the Board makes various necessary changes to any Section of the
    Administrative Code that may be open as a result of USEPA actions taken during the update
    period. The Board updates references to the most recent version of the C.F.R. and various statutes
    and ensures that the regulations are consistent with the requirements of the C.F.R. The Board also
    corrects any spelling or punctuation errors that may be present. The Board performs these
    functions to provide the regulated community with cost effective and accurate regulations. JCAR
    also suggests changes to the regulations that are not a result of the most recent federal amendments.
    Section 17.5 of the Act mandates that the Board adopt regulations that are “identical in
    substance” to the Primary Drinking Water regulations. 415 ILCS 5/17.5 (1998). If the Board
    could not make corrections to the regulations of the type described above to correct any observed
    error, omission, or lack of conformity with rules of the Secretary of State, the Illinois regulations
    would not become or remain “identical in substance” to the federal regulations.
    Section 7.2(a) of the Act contemplates the necessity of such changes in the third sentence,
    in which it directs that “the only changes that may be made by the Board to the federal regulations
    are those changes that in no way change the scope or meaning of any portion of the regulations.”

    4
    415 ILCS 5/7.2(a) (1998). The Board finds that the changes described in the quoted passage of the
    opinion above are well within its statutory authority, and the Board will continue to make them in
    its IIS rulemakings. However, acknowledging the misnomer,
    i.e.
    , “Board Amendments not
    Federally Driven” identified by JCAR, the Board will discontinue use of the inaccurate descriptor.
    Beginning with this opinion, and in those hereafter, the Board will label the portion of the opinion
    describing such changes as “Additional Technical Changes Authorized by Section 7.2(a).”
    DISCUSSION
    September 17, 1999: Revisions to the Unregulated Contaminant Monitoring Regulation for Public
    Water
    Federal Action
    The Safe Drinking Water Act (SDWA)(42 USC §§ 300f
    et seq
    . (1998)), as amended in
    1996, requires the USEPA to establish criteria for a program to monitor for unregulated
    contaminants and, by August, 1999 to publish a list of contaminants to be monitored. The USEPA
    promulgated the Unregulated Contaminant Monitoring Regulation (UMCR) for Public Water
    Systems (PWSs), which significantly revises the existing regulations for unregulated contaminant
    monitoring. The amendments include a list of contaminants to monitor for, procedures for
    selecting a nationwide representative sample of small systems that will be required to monitor,
    and the frequency and schedule for monitoring, the sampling points, and the approved analytical
    methods for testing. The USEPA adopted the amendments under Section 1445(a) of SDWA (42
    USC §§ 300j-4 (1998)).
    Amendments to Board Rules
    The Board proposed incorporating the bulk of these amendments into its rules by adding Sections 611.511,
    611.512, and 611.Appendix I. However, as a result of the public comments of the USEPA and the Agency, the Board
    is not including these amendments into this final rulemaking. The adoption of the unregulated contaminant
    monitoring program is not necessary for State primacy. For a more detailed discussion on the reasoning behind the
    omission of the unregulated contaminant monitoring program, see the “Public Comments” section below.
    December 1, 1999: Revisions to National Primary and Secondary Drinking Water Regulations:
    Analytical Methods for Chemical and Microbiological Contaminants and Revisions to Laboratory
    Certification Requirements.
    Federal Action
    The USEPA approved the use of updated versions of 25 American Society for Testing and
    Materials (ASTM), 54 Standard Methods for Examination of Water and Wastewater Methods
    (Standard Methods), and 13 USEPA analytical methods (USEPA Methods) for compliance
    determinations of chemical contaminants in drinking water. At the same time, the USEPA
    withdrew approval of 13 previous versions of the USEPA Methods. The USEPA also approved
    the use of a new medium and 2 new methods for simultaneous determination of total coliforms and
    E. coli
    . The USEPA also approved new methods of determination for lead, magnesium, and acid
    herbicides. The USEPA made several technical corrections to existing regulations.

    5
    Amendments to Board Rules
    The Board is amending its regulations for coliform sampling in Section 611.526, and
    methods incorporated by reference in Section 611.102. In Section 611.526 microbiological
    monitoring analytical methods, the Board adds tests set forth under the Standard Methods 19th
    edition. The Board also adds the E*Colite® Test (Charm Sciences, Inc.) and the m-ColiBlue24®
    Test (Hatch Company).
    The Board is amending regulations incorporated by reference for inorganic chemical
    sampling and analytical requirements in Section 611.102. The Board is amending its regulations
    for analytical methods or organic chemical contaminants in Section 611.645 and regulations
    incorporated by reference in Section 611.102. In Section 611.645 the Board is adding analytical
    methods that the USEPA added. The Board is also amending its regulations for volatile organic
    chemicals in Section 611.646 and regulations incorporated by reference in Section 611.102. The
    Board amends Section 611.531 to include the Standard Methods 19th edition.
    The Board amends Section 611.490 to require measurements for alkalinity, calcium,
    conductivity, disinfectant residual, orthophosphate, and silica to be performed by a certified
    operator under 35 Ill. Adm. Code 603.103.
    Additional Technical Changes Authorized by Section 7.2(a)
    The Board is updating all C.F.R. citations in the affected sections of this rulemaking to the
    1999 version. The Board is also amending Section 611.310(a) and the Section 611.310(c) Board
    Note to correct any ambiguity in the regulations for the purpose of avoiding confusion to the
    regulated community. Section 611.310 is applicable to all public water systems. The Board is
    also making numerous technical changes suggested by JCAR, the USEPA, and the Agency.
    In PC 1, JCAR submitted a line numbered version of the proposed rule, as it appeared in
    the Illinois Register. Changes in the proposed rule that were made by JCAR prior to publication
    in the Illinois Register were indicated in an accompanying green cover sheet. JCAR highlighted
    various minor technicalities, formatting discrepancies, and misspellings that were corrected by
    JCAR prior to publication. Also, JCAR indicated that it had struck “μg” and had added “mg” four
    times in Section 611.526(f). These changes to the regulations were made in error. The Board
    does not strike the four instances of “μg” currently on file in this final rule.
    In PC 2, JCAR submitted additional suggested changes in a line numbered version of the
    proposed rule. The Board made numerous minor technical changes suggested by JCAR in this
    final rule. The changes include adding and removing punctuation, spacing, capitalization, and
    spelling. The more substantive changes are indicated in the following table of revisions.
    In PC 3, the USEPA submitted comments concerning monitoring and analytical
    requirements. The Board has incorporated these suggestions. The USEPA also stated that since
    revisions to the unregulated contaminant monitoring regulations are not required for primacy, it did
    not review the related proposed amendments.

    6
    In PC 4, the Agency requested that the proposed amendments implementing the unregulated
    contaminant monitoring program be omitted from the final rule. The Agency stated that Illinois is
    not required to adopt these rules in order to maintain primary enforcement authority for the SDWA.
    The Board proposed implementing the unregulated contaminant monitoring program by adding
    Sections 611.511, 611.512, and 611.Appendix I. The Board has omitted these sections in this final
    rule at the request of the Agency. The Agency also noted inconsistencies with respect to certain
    analytical methods. The Board has corrected any inconsistencies.
    Revisions to the Text of the Proposed Amendments in Final Adoption
    Section Revised
    Source(s) of
    Revision(s)
    Revision(s)
    611.102(b)
    JCAR
    Deleted “(See Environetics, Inc.)”
    611.102(b)
    JCAR
    Changed “Water Works” to “Waterworks”
    611.102(b)
    JCAR
    Added parenthesis to “(ATI Orion)”
    611.102(b)
    JCAR
    Capitalized “Complexometric”
    611.102(b)
    JCAR
    Deleted “ASTM Method D2459-72, “Standard Test
    Method for Gamma Spectrometry in Water,” approved
    July 28, 1972, discontinued 1988.”
    611.102(b)
    JCAR
    Did not amend “1995” in the Board Note to “Technical
    Notes on Drinking Water Methods”, EPA-600/R-94-173
    611.310
    JCAR, Board
    Changed “Section 611.641 et seq.” to “Subpart O of this
    Part”
    611.511
    Agency
    Removed proposed Section
    611.512
    Agency
    Removed proposed Section
    611.526(c)(1)
    JCAR
    Changed “false negative” to “false-negative”
    611.526(f)
    Board
    Did not strike “μg” and add “mg” (four times)
    611.531(a)(1)
    JCAR
    Changed “analysis” to “analyses”
    611.531(a)(2)(iii)
    JCAR
    Capitalized “Autoanalysis Colilert System”
    611.531(b)
    JCAR
    Added “19th ed.”
    611.611(a)(13)
    JCAR
    Deleted redundant “Method”

    7
    611.611(a)(16)
    Agency,
    USEPA
    Changed “D3559-90 D” to “D3559-95D”
    611.611(a)(17)
    Agency,
    USEPA
    Changed “D1688-90 C” to “D1688-95 C”
    611.611(a)(19)
    Agency,
    USEPA
    Changed “D1125-91 A” to “D1125-95 A”
    611.611(a)(22)
    Agency,
    USEPA
    Changed “4110” to “4110 B”
    611.611(a)(23)
    Agency,
    USEPA
    Changed “D859-88” to “D859-95”
    611.611(b)(2)
    Agency,
    USEPA
    Added (C): “Samples must be analyzed as soon after
    collection as possible, but in any event within 6
    months.”
    611.611(b)(11)
    Agency,
    USEPA
    Changed “28 days” to “14 days”
    611.611(c)(1)
    JCAR, Board
    Changed “35 Ill. Adm. Code 183.125(c)”, which has
    been repealed, to “35 Ill. Adm. Code Part 186”
    611.611 Board Note
    JCAR, Board
    Deleted the Board Note
    611.645
    USEPA
    Changed TTHM Method “551” to “551.1”
    611.646(d)
    JCAR, Board
    Deleted redundant “compliance period starting in the”
    611.646(k)(2)
    JCAR
    Added “shall be required”
    611.646(k)(2)(A)
    JCAR
    Changed “annual” to “annually”
    611.Appendix I
    Agency
    Removed proposed Section
    ORDER
    The Board will cause the amendments to be published in the
    Illinois Register.
    TITLE 35: ENVIRONMENTAL PROTECTION
    SUBTITLE F: PUBLIC WATER SUPPLIES
    CHAPTER I: POLLUTION CONTROL BOARD
    PART 611

    8
    PRIMARY DRINKING WATER STANDARDS
    SUBPART A: GENERAL
    Section
    611.100
    Purpose, Scope and Applicability
    611.101
    Definitions
    611.102
    Incorporations by Reference
    611.103
    Severability
    611.107
    Agency Inspection of PWS Facilities
    611.108
    Delegation to Local Government
    611.109
    Enforcement
    611.110
    Special Exception Permits
    611.111
    Relief Equivalent to SDWA Section 1415(a) Variances
    611.112
    Relief Equivalent to SDWA Section 1416 Exemptions
    611.113
    Alternative Treatment Techniques
    611.114
    Siting requirements
    611.115
    Source Water Quantity
    611.120
    Effective dates
    611.121
    Maximum Contaminant Levels and Finished Water Quality
    611.125
    Fluoridation Requirement
    611.126
    Prohibition on Use of Lead
    611.130
    Special Requirements for Certain Variances and Adjusted Standards
    611.131
    Relief Equivalent to SDWA Section 1415(e) Small System Variance
    611.160
    Composite Correction Program
    SUBPART B: FILTRATION AND DISINFECTION
    Section
    611.201
    Requiring a Demonstration
    611.202
    Procedures for Agency Determinations
    611.211
    Filtration Required
    611.212
    Groundwater under Direct Influence of Surface Water
    611.213
    No Method of HPC Analysis
    611.220
    General Requirements
    611.230
    Filtration Effective Dates
    611.231
    Source Water Quality Conditions
    611.232
    Site-specific Conditions
    611.233
    Treatment Technique Violations
    611.240
    Disinfection
    611.241
    Unfiltered PWSs
    611.242
    Filtered PWSs
    611.250
    Filtration
    611.261
    Unfiltered PWSs: Reporting and Recordkeeping
    611.262
    Filtered PWSs: Reporting and Recordkeeping
    611.271
    Protection during Repair Work
    611.272
    Disinfection following Repair

    9
    SUBPART C: USE OF NON-CENTRALIZED TREATMENT DEVICES
    Section
    611.280
    Point-of-Entry Devices
    611.290
    Use of Point-of-Use Devices or Bottled Water
    SUBPART D: TREATMENT TECHNIQUES
    Section
    611.295
    General Requirements
    611.296
    Acrylamide and Epichlorohydrin
    611.297
    Corrosion Control
    SUBPART F: MAXIMUM CONTAMINANT LEVELS (MCLs) AND MAXIMUM
    RESIDUAL DISINFECTANT LEVELS (MRDLs)
    Section
    611.300
    Old MCLs for Inorganic Chemicals
    611.301
    Revised MCLs for Inorganic Chemicals
    611.310
    Old Maximum Contaminant Levels (MCLs) for Organic Chemicals
    611.311
    Revised MCLs for Organic Contaminants
    611.312
    Maximum Contaminant Levels (MCLs) for Disinfection Byproducts (DBPs)
    611.313
    Maximum Residual Disinfectant Levels (MRDLs)
    611.320
    Turbidity
    611.325
    Microbiological Contaminants
    611.330
    Radium and Gross Alpha Particle Activity
    611.331
    Beta Particle and Photon Radioactivity
    SUBPART G: LEAD AND COPPER
    Section
    611.350
    General Requirements
    611.351
    Applicability of Corrosion Control
    611.352
    Corrosion Control Treatment
    611.353
    Source Water Treatment
    611.354
    Lead Service Line Replacement
    611.355
    Public Education and Supplemental Monitoring
    611.356
    Tap Water Monitoring for Lead and Copper
    611.357
    Monitoring for Water Quality Parameters
    611.358
    Monitoring for Lead and Copper in Source Water
    611.359
    Analytical Methods
    611.360
    Reporting
    611.361
    Recordkeeping
    SUBPART I: DISINFECTANT RESIDUALS, DISINFECTION BYPRODUCTS,
    AND DISINFECTION BYPRODUCT PRECURSORS
    Section
    611.380
    General Requirements
    611.381
    Analytical Requirements
    611.382
    Monitoring Requirements

    10
    611.383
    Compliance Requirements
    611.384
    Reporting and Recordkeeping Requirements
    611.385
    Treatment Technique for Control of Disinfection Byproduct (DBP) Precursors
    SUBPART K: GENERAL MONITORING AND ANALYTICAL
    REQUIREMENTS
    Section
    611.480
    Alternative Analytical Techniques
    611.490
    Certified Laboratories
    611.491
    Laboratory Testing Equipment
    611.500
    Consecutive PWSs
    611.510
    Special Monitoring for Unregulated Contaminants
    SUBPART L: MICROBIOLOGICAL MONITORING AND ANALYTICAL
    REQUIREMENTS
    Section
    611.521
    Routine Coliform Monitoring
    611.522
    Repeat Coliform Monitoring
    611.523
    Invalidation of Total Coliform Samples
    611.524
    Sanitary Surveys
    611.525
    Fecal Coliform and E. Coli Testing
    611.526
    Analytical Methodology
    611.527
    Response to Violation
    611.531
    Analytical Requirements
    611.532
    Unfiltered PWSs
    611.533
    Filtered PWSs
    SUBPART M: TURBIDITY MONITORING AND ANALYTICAL
    REQUIREMENTS
    Section
    611.560
    Turbidity
    SUBPART N: INORGANIC MONITORING AND ANALYTICAL
    REQUIREMENTS
    Section
    611.591
    Violation of State MCL
    611.592
    Frequency of State Monitoring
    611.600
    Applicability
    611.601
    Monitoring Frequency
    611.602
    Asbestos Monitoring Frequency
    611.603
    Inorganic Monitoring Frequency
    611.604
    Nitrate Monitoring
    611.605
    Nitrite Monitoring
    611.606
    Confirmation Samples
    611.607
    More Frequent Monitoring and Confirmation Sampling
    611.608
    Additional Optional Monitoring

    11
    611.609
    Determining Compliance
    611.610
    Inorganic Monitoring Times
    611.611
    Inorganic Analysis
    611.612
    Monitoring Requirements for Old Inorganic MCLs
    611.630
    Special Monitoring for Sodium
    611.631
    Special Monitoring for Inorganic Chemicals
    SUBPART O: ORGANIC MONITORING AND ANALYTICAL
    REQUIREMENTS
    Section
    611.640
    Definitions
    611.641
    Old MCLs
    611.645
    Analytical Methods for Organic Chemical Contaminants
    611.646
    Phase I, Phase II, and Phase V Volatile Organic Contaminants
    611.647
    Sampling for Phase I Volatile Organic Contaminants (Repealed)
    611.648
    Phase II, Phase IIB, and Phase V Synthetic Organic Contaminants
    611.650
    Monitoring for 36 Contaminants (Repealed)
    611.657
    Analytical Methods for 36 Contaminants (Repealed)
    611.658
    Special Monitoring for Organic Chemicals
    SUBPART P: THM MONITORING AND ANALYTICAL REQUIREMENTS
    Section
    611.680
    Sampling, Analytical and other Requirements
    611.683
    Reduced Monitoring Frequency
    611.684
    Averaging
    611.685
    Analytical Methods
    611.686
    Modification to System
    611.687
    Sampling for THM Potential
    611.688
    Applicability Dates
    SUBPART Q: RADIOLOGICAL MONITORING AND ANALYTICAL
    REQUIREMENTS
    Section
    611.720
    Analytical Methods
    611.731
    Gross Alpha
    611.732
    Manmade Radioactivity
    SUBPART R: ENHANCED FILTRATION AND DISINFECTION
    Section
    611.740
    General Requirements
    611.741
    Standards for Avoiding Filtration
    611.742
    Disinfection Profiling and Benchmarking
    611.743
    Filtration
    611.744
    Filtration Sampling Requirements
    611.745
    Reporting and Recordkeeping Requirements

    12
    SUBPART T: REPORTING, PUBLIC NOTIFICATION AND
    RECORDKEEPING
    Section
    611.830
    Applicability
    611.831
    Monthly Operating Report
    611.832
    Notice by Agency
    611.833
    Cross Connection Reporting
    611.840
    Reporting
    611.851
    Reporting MCL, MRDL, and other Violations
    611.852
    Reporting other Violations
    611.853
    Notice to New Billing Units
    611.854
    General Content of Public Notice
    611.855
    Mandatory Health Effects Language
    611.856
    Fluoride Notice
    611.858
    Fluoride Secondary Standard
    611.860
    Record Maintenance
    611.870
    List of 36 Contaminants
    SUBPART U: CONSUMER CONFIDENCE REPORTS
    Section
    611.881
    Purpose and Applicability of this Subpart
    611.882
    Compliance Dates
    611.883
    Content of the Reports
    611.884
    Required Additional Health Information
    611.885
    Report Delivery and Recordkeeping
    611.Appendix A
    Mandatory Health Effects Information
    611.Appendix B
    Percent Inactivation of G. Lamblia Cysts
    611.Appendix C
    Common Names of Organic Chemicals
    611.Appendix D
    Defined Substrate Method for the Simultaneous Detection of Total
    Coliforms and Eschericia Coli from Drinking Water
    611.Appendix E
    Mandatory Lead Public Education Information
    611.Appendix F
    Converting Maximum Contaminant Level (MCL) Compliance Values for
    Consumer Confidence Reports
    611.Appendix G
    Regulated Contaminants
    611.Appendix H
    Health Effects Language
    611.Table A
    Total Coliform Monitoring Frequency
    611.Table B
    Fecal or Total Coliform Density Measurements
    611.Table C
    Frequency of RDC Measurement
    611.Table D
    Number of Lead and Copper Monitoring Sites
    611.Table E
    Lead and Copper Monitoring Start Dates
    611.Table F
    Number of Water Quality Parameter Sampling Sites
    611.Table G
    Summary of Monitoring Requirements for Water Quality Parameters
    1
    611.Table Z
    Federal Effective Dates

    13
    AUTHORITY: Implementing Sections 7.2, 17, and 17.5 and authorized by Section 27 of the
    Environmental Protection Act [415 ILCS 5/7.2, 17, 17.5, and 27].
    SOURCE: Adopted in R88-26 at 14 Ill. Reg. 16517, effective September 20, 1990; amended in
    R90-21 at 14 Ill. Reg. 20448, effective December 11, 1990; amended in R90-13 at 15 Ill. Reg.
    1562, effective January 22, 1991; amended in R91-3 at 16 Ill. Reg. 19010, effective December 1,
    1992; amended in R92-3 at 17 Ill. Reg. 7796, effective May 18, 1993; amended in R93-1 at 17 Ill.
    Reg. 12650, effective July 23, 1993; amended in R94-4 at 18 Ill. Reg. 12291, effective July 28,
    1994; amended in R94-23 at 19 Ill. Reg. 8613, effective June 20, 1995; amended in R95-17 at 20
    Ill. Reg. 14493, effective October 22, 1996; amended in R98-2 at 22 Ill. Reg. 5020, effective
    March 5, 1998; amended in R99-6 at 23 Ill. Reg. 2756, effective February 17, 1999; amended in
    R99-12 at 23 Ill. Reg. 10348, effective August 11, 1999; amended in R00-8 at 23 Ill. Reg. 14715,
    effective December 8, 1999; amended in R00-10 at 24 Ill. Reg. ______, effective_______.
    SUBPART A: GENERAL
    Section 611.102
    Incorporations by Reference
    a)
    Abbreviations and short-name listing of references. The following names and
    abbreviated names, presented in alphabetical order, are used in this Part to refer to
    materials incorporated by reference:
    “Amco-AEPA-1 Polymer” is available from Advanced Polymer Systems.
    “ASTM Method” means a method published by and available from the
    American Society for Testing and Materials (ASTM).
    “Colisure Test” means “Colisure Presence/Absence Test for Detection and
    Identification of Coliform Bacteria and Escherichia Coli in Drinking
    Water”, available from Millipore Corporation, Technical Services
    Department.
    “Dioxin and Furan Method 1613” means “Tetra- through Octa- Chlorinated
    Dioxins and Furans by Isotope-Dilution HRGC/HRMS”, available from
    NTIS.
    “GLI Method 2” means GLI Method 2, “Turbidity”, Nov. 2, 1992, available
    from Great Lakes Instruments, Inc.
    “Guidance Manual for Compliance with the Filtration and Disinfection
    Requirements for Public Water Systems Using Surface Water Sources”,
    available from USEPA Science and Technology Branch.
    “HASL Procedure Manual” means HASL Procedure Manual, HASL 300,
    available from ERDA Health and Safety Laboratory.

    14
    “Maximum Permissible Body Burdens and Maximum Permissible
    Concentrations of Radionuclides in Air and in Water for Occupational
    Exposure”, NCRP Report Number 22, available from NCRP.
    “NCRP” means “National Council on Radiation Protection”.
    “NTIS” means “National Technical Information Service”.
    “New Jersey Radium Method” means “Determination of Radium 228 in
    Drinking Water”, available from the New Jersey Department of
    Environmental Protection.
    “New York Radium Method” means “Determination of Ra-226 and Ra-228
    (Ra-02)”, available from the New York Department of Public Health.
    “ONGP-MUG Test” (meaning “minimal medium ortho-nitrophenyl-beta-d-
    galactopyranoside-4-methyl-umbelliferyl-beta-d-glucuronide test”), also
    called the “Autoanalysis Colilert System”, is Method 9223, available in
    “Standard Methods for the Examination of Water and Wastewater”, 18th
    ed., from American Public Health Association.
    “Procedures for Radiochemical Analysis of Nuclear Reactor Aqueous
    Solutions”, available from NTIS.
    “Radiochemical Methods” means “Interim Radiochemical Methodology for
    Drinking Water”, available from NTIS.
    “Standard Methods”, means “Standard Methods for the Examination of
    Water and Wastewater”, available from the American Public Health
    Association or the American Waterworks Association.
    “Technical Bulletin 601” means “Technical Bulletin 601, Standard Method
    of Testing for Nitrate in Drinking Water”, July 1994, available from
    Analytical Technology, Inc.
    “Technicon Methods” means “Fluoride in Water and Wastewater”,
    available from Technicon.
    “USDOE Manual” means “EML Procedures Manual”, available from the
    United State Department of Energy.
    “USEPA Asbestos Methods-100.1” means Method 100.1, “Analytical
    Method for Determination of Asbestos Fibers in Water”, available from
    NTIS.

    15
    “USEPA Asbestos Methods-100.2” means Method 100.2, “Determination
    of Asbestos Structures over 10-mm in Length in Drinking Water”, available
    from NTIS.
    “USEPA Environmental Inorganics Methods” means “Methods for the
    Determination of Inorganic Substances in Environmental Samples”,
    available from NTIS; “Methods for the Determination of Inorganic
    Substances in Environmental Samples”, August 1993, for Method 300.0;
    “Determination of Inorganic Anions in Drinking Water by Ion
    Chromatography, Revision 1.0”, 1997, for Method 300.1.
    “USEPA Environmental Metals Methods” means “Methods for the
    Determination of Metals in Environmental Samples”, available from NTIS.
    “USEPA Organic Methods” means “Methods for the Determination of
    Organic Compounds in Drinking Water”, July 1991, for Methods 502.2,
    505, 507, 508, 508A, 515.1, and 531.1; “Methods for the Determination of
    Organic Compounds in Drinking Water--Supplement I”, July 1990, for
    Methods 506, 547, 550, 550.1, and 551; and “Methods for the
    Determination of Organic Compounds in Drinking Water--Supplement II”,
    August 1992, for Methods 515.2, 524.2, 548.1, 549.1, 552.1, and 555,
    available from NTIS. Methods 504.1, 508.1, and 525.2 are available from
    EPA EMSL; “Methods for the Determination of Organic Compounds” in
    Drinking Water-Supplement II, August 1992, for Method 552.1; “Methods
    for the Determination of Organic Compounds in Drinking
    Water-Supplement III”, August 1995, for Methods 502.2, 524.2, 551.1, and
    552.2.
    “USEPA Interim Radiochemical Methods” means “Interim Radiochemical
    Methodology for Drinking Water”, EPA 600/4-75-008 (revised), March
    1976. Available from NTIS.
    “USEPA Radioactivity Methods” means “Prescribed Procedures for
    Measurement of Radioactivity in Drinking Water”, EPA 600/4-80-032,
    August 1980. Available from NTIS.
    “USEPA Radiochemical Analyses” means “Radiochemical Analytical
    Procedures for Analysis of Environmental Samples”, March 1979.
    Available from NTIS.
    “USEPA Radiochemistry Methods” means “Radiochemistry Procedures
    Manual”, EPA 520/5-84-006, December 1987. Available from NTIS.
    “USEPA Technical Notes” means “Technical Notes on Drinking Water
    Methods”, available from NTIS.

    16
    “USGS Methods” means “Methods of Analysis by the U.S. Geological
    Survey National Water Quality Laboratory--Determination of Inorganic and
    Organic Constituents in Water and Fluvial Sediments”, available from
    NTIS and USGS.
    “Waters Method B-1011” means “Waters Test Method for the
    Determination of Nitrite/Nitrate in Water Using Single Column Ion
    Chromatography”, available from Millipore Corporation, Waters
    Chromatography Division.
    b)
    The Board incorporates the following publications by reference:
    Access Analytical Systems, Inc. (See Environetics, Inc.)
    Advanced Polymer Systems, 3696 Haven Avenue, Redwood City, CA
    94063 415-366-2626:
    Amco-AEPA-1 Polymer. See 40 CFR 141.22(a) (19981999).
    Also, as referenced in ASTM D1889.
    American Public Health Association, 1015 Fifteenth Street NW,
    Washington, DC 20005 800-645-5476:
    “Standard Methods for the Examination of Water and Wastewater”,
    17th Edition, 1989 (referred to as “Standard Methods, 17th ed.”).
    “Standard Methods for the Examination of Water and Wastewater”,
    18th Edition, 1992, including “Supplement to the 18th Edition of
    Standard Methods for the Examination of Water and Wastewater”,
    1994 (collectively referred to as “Standard Methods, 18th ed.”).
    See the methods listed separately for the same references under
    American Water WorksWaterworks Association.
    “Standard Methods for the Examination of Water and Wastewater”,
    19th Edition, 1995 (referred to as “Standard Methods, 19th ed.”).
    American Waterworks Association et al., 6666 West Quincy Ave., Denver,
    CO 80235 303-794-7711:
    Standard Methods for the Examination of Water and Wastewater,
    13th Edition, 1971 (referred to as “Standard Methods, 13th ed.”).
    Method 302, Gross Alpha and Gross Beta Radioactivity in
    Water (Total, Suspended and Dissolved).

    17
    Method 303, Total Radioactive Strontium and Strontium 90
    in Water.
    Method 304, Radium in Water by Precipitation.
    Method 305, Radium 226 by Radon in Water (Soluble,
    Suspended and Total).
    Method 306, Tritium in Water.
    Standard Methods for the Examination of Water and Wastewater,
    18th Edition, 1992 (referred to as “Standard Methods, 18th ed.”):
    Method 2130 B, Turbidity, Nephelometric Method.
    Method 2320 B, Alkalinity, Titration Method.
    Method 2510 B, Conductivity, Laboratory Method.
    Method 2550, Temperature, Laboratory and Field Methods.
    Method 3111 B, Metals by Flame Atomic Absorption
    Spectrometry, Direct Air-Acetylene Flame Method.
    Method 3111 D, Metals by Flame Atomic Absorption
    Spectrometry, Direct Nitrous Oxide-Acetylene Flame
    Method.
    Method 3112 B, Metals by Cold-Vapor Atomic Absorption
    Spectrometry, Cold-Vapor Atomic Absorption
    Spectrometric Method.
    Method 3113 B, Metals by Electrothermal Atomic
    Absorption Spectrometry, Electrothermal Atomic
    Absorption Spectrometric Method.
    Method 3114 B, Metals by Hydride Generation/Atomic
    Absorption Spectrometry, Manual Hydride
    Generation/Atomic Absorption Spectrometric Method.
    Method 3120 B, Metals by Plasma Emission Spectroscopy,
    Inductively Coupled Plasma (ICP) Method.
    Method 3500-Ca D, Calcium, EDTA Titrimetric Method.

    18
    Method 4110 B, Determination of Anions by Ion
    Chromatography, Ion Chromatography with Chemical
    Suppression of Eluent Conductivity.
    Method 4500-CN
    -
    C, Cyanide, Total Cyanide after
    Distillation.
    Method 4500-CN
    -
    E, Cyanide, Colorimetric Method.
    Method 4500-CN
    -
    F, Cyanide, Cyanide-Selective Electrode
    Method.
    Method 4500-CN
    -
    G, Cyanide, Cyanides Amenable to
    Chlorination after Distillation.
    Method 4500-ClO
    2
    C, Chlorine Dioxide, Amperometric
    Method I.
    Method 4500-F
    -
    B, Fluoride, Preliminary Distillation Step.
    Method 4500-F
    -
    C, Fluoride, Ion-Selective Electrode
    Method.
    Method 4500-F
    -
    D, Fluoride, SPADNS Method.
    Method 4500-F
    -
    E, Fluoride, Complexone Method.
    Method 4500-H
    +
    B, pH Value, Electrometric Method.
    Method 4500-NO
    2
    -
    B, Nitrogen (Nitrite), Colorimetric
    Method.
    Method 4500-NO
    3
    -
    D, Nitrogen (Nitrate), Nitrate Electrode
    Method.
    Method 4500-NO
    3
    -
    E, Nitrogen (Nitrate), Cadmium
    Reduction Method.
    Method 4500-NO
    3
    -
    F, Nitrogen (Nitrate), Automated
    Cadmium Reduction Method.
    Method 4500-O
    3
    B, Ozone (Residual) (Proposed), Indigo
    Colorimetric Method.
    Method 4500-P E, Phosphorus, Ascorbic Acid Method.

    19
    Method 4500-P F, Phosphorus, Automated Ascorbic Acid
    Reduction Method.
    Method 4500-Si D, Silica, Molybdosilicate Method.
    Method 4500-Si E, Silica, Heteropoly Blue Method.
    Method 4500-Si F, Silica, Automated Method for
    Molybdate-Reactive Silica.
    Method 4500-SO
    4
    2-
    C, Sulfate, Gravimetric Method with
    Ignition of Residue.
    Method 4500-SO
    4
    2-
    D, Sulfate, Gravimetric Method with
    Drying of Residue.
    Method 4500-SO
    4
    2-
    F, Sulfate, Automated Methylthymol
    Blue Method.
    Method 6610, Carbamate Pesticide Method.
    Method 6651, Glyphosate Herbicide (Proposed).
    Method 7110 B, Gross Alpha and Beta Radioactivity (Total,
    Suspended, and Dissolved), Evaporation Method for Gross
    Alpha-Beta.
    Method 7110 C, Gross Alpha and Beta Radioactivity (Total,
    Suspended, and Dissolved), Coprecipitation Method for
    Gross Alpha Radioactivity in Drinking Water (Proposed).
    Method 7500-Cs B, Radioactive Cesium, Precipitation
    Method.
    Method 7500-3H, B, Tritium, Liquid Scintillation
    Spectrometric Method.
    Method 7500-I B, Radioactive Iodine, Precipitation Method.
    Method 7500-I C, Radioactive Iodine, Ion-Exchange
    Method.
    Method 7500-I D, Radioactive Iodine, Distillation Method.
    Method 7500-Ra B, Radium, Precipitation Method.

    20
    Method 7500-Ra C, Radium, Emanation Method.
    Method 7500-Ra D, Radium, Sequential Precipitation
    Method (Proposed).
    Method 7500-U B, Uranium, Radiochemical Method
    (Proposed).
    Method 7500-U C, Uranium, Isotopic Method (Proposed).
    Method 9215 B, Heterotrophic Plate Count, Pour Plate
    Method.
    Method 9221 A, Multiple-Tube Fermentation Technique for
    Members of the Coliform Group, Introduction.
    Method 9221 B, Multiple-Tube Fermentation Technique for
    Members of the Coliform Group, Standard Total Coliform
    Fermentation Technique.
    Method 9221 C, Multiple-Tube Fermentation Technique for
    Members of the Coliform Group, Estimation of Bacterial
    Density.
    Method 9221 D, Multiple-Tube Fermentation Technique for
    Members of the Coliform Group, Presence-Absence (P-A)
    Coliform Test.
    Method 9222 A, Membrane Filter Technique for Members
    of the Coliform Group, Introduction.
    Method 9222 B, Membrane Filter Technique for Members
    of the Coliform Group, Standard Total Coliform Membrane
    Filter Procedure.
    Method 9222 C, Membrane Filter Technique for Members
    of the Coliform Group, Delayed-Incubation Total Coliform
    Procedure.
    Method 9223, Chromogenic Substrate Coliform Test
    (Proposed).
    Standard Methods for the Examination of Water and Wastewater,
    19th Edition, 1995 (referred to as “Standard Methods, 19th ed.”):
    Method 7120-B, Gamma Spectrometric Method.

    21
    Method 7500-U C, Uranium, Isotopic Method.
    Method 4500-Cl D, Chlorine (Residual), Amperometric
    Titration Method.
    Method 4500-Cl E, Chlorine (Residual), Low-Level
    Amperometric Titration Method.
    Method 4500-Cl F, Chlorine (Residual), DPD Ferrous
    Titrimetric Method.
    Method 4500-Cl G, Chlorine (Residual), DPD Colorimetric
    Method.
    Method 4500-Cl H, Chlorine (Residual), Syringaldazine
    (FACTS) Method.
    Method 4500-Cl I, Chlorine (Residual), Iodometric
    Electrode Technique.
    Method 4500-ClO
    2
    D, Chlorine Dioxide, DPD Method.
    Method 4500-ClO
    2
    E, Chlorine Dioxide, Amperometric
    Method II.
    Method 6251 B, Disinfection Byproducts: Haloacetic Acids
    and Trichlorophenol, Micro Liquid-Liquid Extraction Gas
    Chromatographic Method.
    Method 5910 B, UV Absorbing Organic Constituents,
    Ultraviolet Absorption Method.
    Supplement to the 19th Edition of Standard Methods for the
    Examination of Water and Wastewater, American Public Health
    Association, 1996:
    Method 5310 B, TOC, Combustion-Infrared Method.
    Method 5310 C, TOC, Persulfate-Ultraviolet Oxidation
    Method.
    Method 5310 D, TOC, Wet-Oxidation Method.
    Analytical Technology, Inc. (ATI Orion), 529 Main Street, Boston, MA
    02129:

    22
    Technical Bulletin 601, “Standard Method of Testing for Nitrate in
    Drinking Water”, July, 1994, PN 221890-001 (referred to as
    “Technical Bulletin 601”).
    ASTM. American Society for Testing and Materials, 1976 Race Street,
    Philadelphia, PA 19103 215-299-5585:
    ASTM Method D511-93 A and B, “Standard Test Methods for
    Calcium and Magnesium in Water”, “Test Method A--
    Ccomplexometric Titration” & “Test Method B--Atomic Absorption
    Spectrophotometric”, approved 1993.
    ASTM Method D515-88 A, “Standard Test Methods for Phosphorus
    in Water”, “Test Method A--Colorimetric Ascorbic Acid
    Reduction”, approved August 19, 1988.
    ASTM Method D859-88, “Standard Test Method for Silica in
    Water”, approved August 19, 1988.
    ASTM Method D1067-92 B, “Standard Test Methods for Acidity or
    Alkalinity in Water”, “Test Method B--Electrometric or Color-
    Change Titration”, approved May 15, 1992.
    ASTM Method D1125-91 A, “Standard Test Methods for Electrical
    Conductivity and Resistivity of Water”, “Test Method A--Field and
    Routine Laboratory Measurement of Static (Non-Flowing)
    Samples”, approved June 15, 1991.
    ASTM Method D1179-93 B, “Standard Test Methods for Fluoride
    in Water”, “Test Method B--Ion Selective Electrode”, approved
    1993.
    ASTM Method D1293-84, “Standard Test Methods for pH of
    Water”, “Test Method A--Precise Laboratory Measurement” &
    “Test Method B--Routine or Continuous Measurement”, approved
    October 26, 1984.
    ASTM Method D1688-90 A or C, “Standard Test Methods for
    Copper in Water”, “Test Method A--Atomic Absorption, Direct” &
    “Test Method C--Atomic Absorption, Graphite Furnace”, approved
    March 15, 1990.
    ASTM Method D2036-91 A or B, “Standard Test Methods for
    Cyanide in Water”, “Test Method A--Total Cyanides after

    23
    Distillation” & “Test Method B--Cyanides Amenable to
    Chlorination by Difference”, approved September 15, 1991.
    ASTM Method D2459-72, “Standard Test Method for Gamma
    Spectrometry in Water,” approved July 28, 1972, discontinued
    1988.
    ASTM Method D2460-90, “Standard Test Method for
    Radionuclides of Radium in Water”, approved 1990.
    ASTM Method D2907-91, “Standard Test Methods for
    Microquantities of Uranium in Water by Fluorometry”, “Test
    Method A--Direct Fluorometric” & “Test Method B—Extraction”,
    approved June 15, 1991.
    ASTM Method D2972-93 B or C, “Standard Test Methods for
    Arsenic in Water”, “Test Method B--Atomic Absorption, Hydride
    Generation” & “Test Method C--Atomic Absorption, Graphite
    Furnace”, approved 1993.
    ASTM Method D3223-91, “Standard Test Method for Total
    Mercury in Water”, approved September 23, 1991.
    ASTM Method D3454-91, “Standard Test Method for Radium-226
    in Water”, approved 1991.
    ASTM Method D3559-90 D, “Standard Test Methods for Lead in
    Water”, “Test Method D--Atomic Absorption, Graphite Furnace”,
    approved August 6, 1990.
    ASTM Method D3645-93 B, “Standard Test Methods for Beryllium
    in Water”, “Method B--Atomic Absorption, Graphite Furnace”,
    approved 1993.
    ASTM Method D3649-91, “Standard Test Method for High-
    Resolution Gamma-Ray Spectrometry of Water”, approved 1991.
    ASTM Method D3697-92, “Standard Test Method for Antimony in
    Water”, approved June 15, 1992.
    ASTM Method D3859-93 A, “Standard Test Methods for Selenium
    in Water”, “Method A--Atomic Absorption, Hydride Method”,
    approved 1993.
    ASTM Method D3867-90 A and B, “Standard Test Methods for
    Nitrite-Nitrate in Water”, “Test Method A--Automated Cadmium

    24
    Reduction” & “Test Method B--Manual Cadmium Reduction”,
    approved January 10, 1990.
    ASTM Method D3972-90, “Standard Test Method for Isotopic
    Uranium in Water by Radiochemistry”, approved 1990.
    ASTM Method D4107-91, “Standard Test Method for Tritium in
    Drinking Water”, approved 1991.
    ASTM Method D4327-91, “Standard Test Method for Anions in
    Water by Ion Chromatography”, approved October 15, 1991.
    ASTM Method D4785-88, “Standard Test Method for Low-Level
    Iodine-131 in Water”, approved 1988.
    ASTM Method D5174-91, “Standard Test Method for Trace
    Uranium in Water by Pulsed-Laser Phosphorimetry”, approved
    1991.
    ASTM Method D 1253-86, “Standard Test Method for Residual
    Chlorine in Water,”, reapproved 1992.
    ERDA Health and Safety Laboratory, New York, NY:
    HASL Procedure Manual, HASL 300, 1973. See 40 CFR
    141.25(b)(2) (19981999).
    Great Lakes Instruments, Inc., 8855 North 55th Street, Milwaukee, WI
    53223:
    GLI Method 2, “Turbidity”, Nov. 2, 1992.
    Millipore Corporation, Technical Services Department, 80 Ashby Road,
    Milford, MA 01730 800-654-5476:
    Colisure Presence/Absence Test for Detection and Identification of
    Coliform Bacteria and Escherichia Coli in Drinking Water,
    February 28, 1994 (referred to as “Colisure Test”).
    Millipore Corporation, Waters Chromatography Division, 34 Maple St.,
    Milford, MA 01757 800-252-4752:
    Waters Test Method for the Determination of Nitrite/Nitrate in
    Water Using Single Column Ion Chromatography, Method B-1011
    (referred to as “Waters Method B-1011”).

    25
    NCRP. National Council on Radiation Protection, 7910 Woodmont Ave.,
    Bethesda, MD 301-657-2652:
    “Maximum Permissible Body Burdens and Maximum Permissible
    Concentrations of Radionuclides in Air and in Water for
    Occupational Exposure”, NCRP Report Number 22, June 5, 1959.
    NSF. National Sanitation Foundation International, 3475 Plymouth Road,
    PO Box 130140, Ann Arbor, Michigan 48113-0140, 734-769-8010:
    NSF Standard 61, section 9, November 1998.
    NTIS. National Technical Information Service, U.S. Department of
    Commerce, 5285 Port Royal Road, Springfield, VA 22161, 703- 487-4600
    or 800-553-6847:
    “Interim Radiochemical Methodology for Drinking Water”, EPA
    600/4-75-008 (revised), March 1976 (referred to as “USEPA
    Interim Radiochemical Methods”). (Pages 1, 4, 6, 9, 13, 16, 24, 29,
    34)
    Method 100.1, “Analytical Method for Determination of Asbestos
    Fibers in Water”, EPA-600/4-83-043, September, 1983, Doc. No.
    PB83-260471 (referred to as “USEPA Asbestos Methods-100.1”).
    Method 100.2, “Determination of Asbestos Structures over 10-mm
    in Length in Drinking Water”, EPA-600/4-83-043, June, 1994, Doc.
    No. PB94-201902 (Rreferred to as “USEPA Asbestos Methods-
    100.2”).
    “Methods for Chemical Analysis of Water and Wastes”, March,
    1983, Doc. No. PB84-128677 (referred to as “USEPA Inorganic
    Methods”). (Methods 150.1, 150.2, and 245.2, which formerly
    appeared in this reference, are available from USEPA EMSL.)
    “Methods for the Determination of Metals in Environmental
    Samples”, June, 1991, Doc. No. PB91-231498 (referred to as
    “USEPA Environmental Metals Methods”).
    “Methods for the Determination of Organic Compounds in Drinking
    Water”, December, 1988, revised July, 1991, EPA-600/4-88/039
    (referred to as “USEPA Organic Methods”). (For methods 502.2,
    505, 507, 508, 508A, 515.1, and 531.1.)
    “Methods for the Determination of Organic Compounds in Drinking
    Water--Supplement I”, July, 1990, EPA-600-4-90-020 (referred to

    26
    as “USEPA Organic Methods”). (For methods 506, 547, 550,
    550.1, and 551.)
    “Methods for the Determination of Organic Compounds in Drinking
    Water--Supplement II”, August, 1992, EPA-600/R-92-129 (referred
    to as “USEPA Organic Methods”). (For methods 515.2, 524.2,
    548.1, 549.1, 552.1, and 555.)
    “Prescribed Procedures for Measurement of Radioactivity in
    Drinking Water”, EPA 600/4-80-032, August 1980 (referred to as
    “USEPA Radioactivity Methods”). (Methods 900, 901, 901.1, 902,
    903, 903.1, 904, 905, 906, 908, 908.1)
    “Procedures for Radiochemical Analysis of Nuclear Reactor
    Aqueous Solutions”, H.L. Krieger and S. Gold, EPA-R4-73-014,
    May, 1973, Doc. No. PB222-154/7BA.
    “Radiochemical Analytical Procedures for Analysis of
    Environmental Samples”, March, 1979, Doc. No. EMSL LV 053917
    (referred to as “USEPA Radiochemical Analyses”). (Pages 1, 19,
    33, 65, 87, 92)
    “Radiochemistry Procedures Manual”, EPA-520/5-84-006,
    December, 1987, Doc. No. PB-84-215581 (referred to as “USEPA
    Radiochemistry Methods”). (Methods 00-01, 00-02, 00-07, H-02,
    Ra-03, Ra-04, Ra-05, Sr-04)
    “Technical Notes on Drinking Water Methods”, EPA-600/R-94-
    173, October, 1994, Doc. No. PB-104766 (referred to as “USEPA
    Technical Notes”).
    BOARD NOTE: USEPA made the following assertion with regard
    to this reference at 40 CFR 141.23(k)(1) and 141.24(e) and (n)(11)
    (1995): “This document contains other analytical test procedures
    and approved analytical methods that remain available for
    compliance monitoring until July 1, 1996.”
    “Tetra- through Octa-Chlorinated Dioxins and Furans by Isotope
    Dilution HRGC/HRMS”, October, 1994, EPA-821-B-94-005
    (referred to as “Dioxin and Furan Method 1613”).
    New Jersey Department of Environment, Division of Environmental
    Quality, Bureau of Radiation and Inorganic Analytical Services, 9 Ewing
    Street, Trenton, NJ 08625:
    “Determination of Radium 228 in Drinking Water”, August 1990.

    27
    New York Department of Health, Radiological Sciences Institute, Center
    for Laboratories and Research, Empire State Plaza, Albany, NY 12201:
    “Determination of Ra-226 and Ra-228 (Ra-02)”, January 1980,
    Revised June 1982.
    Technicon Industrial Systems, Tarrytown, NY 10591:
    “Fluoride in Water and Wastewater”, Industrial Method #129-71W,
    December, 1972 (referred to as “Technicon Methods: Method
    #129-71W”). See 40 CFR 141.23(k)(1), footnote 11 (19951999).
    “Fluoride in Water and Wastewater”, #380-75WE, February, 1976
    (referred to as “Technicon Methods: Method #380-75WE”). See
    40 CFR 141.23(k)(1), footnote 11 (19951999).
    United States Department of Energy, available at the Environmental
    Measurements Laboratory, U.S. Department of Energy, 376 Hudson Street,
    New York, NY 10014-3621:
    “EML Procedures Manual”, 27th Edition, Volume 1, 1990.
    United States Environmental Protection Agency, EMSL, Cincinnati, OH
    45268 513-569-7586:
    “Interim Radiochemical Methodology for Drinking Water”, EPA-
    600/4-75-008 (referred to as “Radiochemical Methods”).
    (Revised) March, 1976.
    “Methods for the Determination of Organic Compounds in Finished
    Drinking Water and Raw Source Water” (referred to as “USEPA
    Organic Methods”). (For methods 504.1, 508.1, and 525.2 only.).
    See NTIS.
    “Procedures for Radiochemical Analysis of Nuclear Reactor
    Aqueous Solutions”. See NTIS.
    USEPA, Science and Technology Branch, Criteria and Standards Division,
    Office of Drinking Water, Washington, D.C. 20460:
    “Guidance Manual for Compliance with the Filtration and
    Disinfection Requirements for Public Water Systems using Surface
    Water Sources”, October, 1989.

    28
    USGS. Books and Open-File Reports Section, United States Geological
    Survey, Federal Center, Box 25425, Denver, CO 80225-0425:
    Methods available upon request by method number from “Methods
    of Analysis by the U.S. Geological Survey National Water Quality
    Laboratory--Determination of Inorganic and Organic Constituents in
    Water and Fluvial Sediments”, Open File Report 93-125 or Book 5,
    Chapter A-1, “Methods for Determination of Inorganic Substances
    in Water and Fluvial Sediments”, 3d3rd ed., Open-File Report 85-
    495, 1989, as appropriate (referred to as “USGS Methods”).
    I-1030-85
    I-1062-85
    I-1601-85
    I-1700-85
    I-2598-85
    I-2601-90
    I-2700-85
    I-3300-85
    Methods available upon request by method number from “Methods
    for Determination of Radioactive Substances in Water and Fluvial
    Sediments”, Chapter A5 in Book 5 of “Techniques of Water-
    Resources Investigations of the United States Geological Survey”,
    1997.
    R-1110-76
    R-1111-76
    R-1120-76
    R-1140-76
    R-1141-76
    R-1142-76
    R-1160-76

    29
    R-1171-76
    R-1180-76
    R-1181-76
    R-1182-76
    c)
    The Board incorporates the following federal regulations by reference:
    40 CFR 136, Appendix B and C (19981999).
    40 CFR 141.40(a)(3) Table 1, Lists 1, 2, and 3 (1999).
    40 CFR 141.40(a)(4) Table 2 (1999).
    40 CFR 141.40(a)(5) Table 3 (1999).
    d)
    This Part incorporates no later amendments or editions.
    (Source: Amended at 24 Ill. Reg. ________, effective ________)
    SUBPART F: MAXIMUM CONTAMINANT LEVELS (MCLs) AND MAXIMUM
    RESIDUAL DISINFECTANT LEVELS (MRDLs)
    Section 611.310
    Old Maximum Contaminant Levels (MCLs) for Organic Chemicals
    The following are the MCLs for organic chemicals. The MCLs for organic chemicals in
    subsections (a) and (b)this Section apply to all CWSs. Compliance with the MCLs in subsections
    (a) and (b) is calculated pursuant to Section 611.641 et seq.Subpart O of this Part. Compliance
    with the MCL for TTHMin subsection (c) is calculated pursuant to Subpart P of this Part.
    Contaminant
    Level
    mg/L
    Additional State
    Requirement (*)
    a)
    Chlorinated hydrocarbons
    Aldrin
    0.001
    *
    DDT
    0.05
    *
    Dieldrin
    0.001
    *
    Heptachlor
    0.0001
    *
    Heptachlor epoxide
    0.0001
    *
    BOARD NOTE: Originally derived from 40 CFR 141.12(a)(1994), USEPA
    removed the last entry in this subsection and marked it reserved at 57 Fed. Reg.

    30
    31838 (July 17, 1992). USEPA added another listing of organic MCLs at 40 CFR
    141.61 (1994). Heptachlor, heptachlor epoxide, and 2,4-D appear in both this
    Section and in Section 611.311, with a different MCL in each Section. The
    heptachlor, heptachlor epoxide, and 2,4-D MCLs in this Section are Illinois
    limitations that are more stringent than the federal requirements. However,
    detection of these contaminants or violation of their federally-derived revised
    Section 611.311 MCLs imposes more stringent monitoring, reporting, and notice
    requirements.
    b)
    Chlorophenoxys:
    2,4-D
    0.01
    *
    BOARD NOTE: Originally derived from 40 CFR 141.12(b) (1994), USEPA
    removed the last entry in this subsection and marked it reserved at 56 Fed. Reg.
    3578 (Jan. 30, 1991). See the preceding Board Note regarding the dual listing of
    MCLs for 2,4-D.
    c)
    TTHM
    0.10
    *
    1)
    The MCL of 0.10 mg/L for TTHM applies to a Subpart B community water
    system that serves 10,000 or more persons, until December 31, 2001.
    2)
    The MCL of 0.10 mg/L for TTHM applies to community water systems that
    use only groundwater not under the direct influence of surface water and
    serve 10,000 or more persons, until December 31, 2003.
    3)
    After December 31, 2003, the MCL for TTHM in this Section is no longer
    applicable.
    BOARD NOTE: Derived in part from 40 CFR 141.12(c) (19981999). This is an
    additional State requirement to the extent that it applies to supplies other than
    CWSs that add a disinfectant at any part of treatment and which provide water to
    10,000 or more persons. Also derived from 40 CFR 141.12 (1998). The new
    MCL for TTHM is listed in Section 611.312.
    (Source: Amended at 24 Ill. Reg. ________, effective __________)
    SUBPART G: LEAD AND COPPER
    Section 611.359
    Analytical Methods
    Analyses for lead, copper, pH, conductivity, calcium, alkalinity, orthophosphate, silica, and
    temperature shall be conducted using the methods set forth in Section 611.611(a).
    a)
    Analyses for lead and copper performed for the purposes of compliance with this
    Subpart shall only be conducted by laboratories that have been certified by USEPA

    31
    or the Agency. To obtain certification to conduct analyses for lead and copper,
    laboratories must:
    1)
    Analyze performance evaluation samples that include lead and copper
    provided by USEPA Environmental Monitoring and Support Laboratory or
    equivalent samples provided by the Agency; and
    2)
    Achieve quantitative acceptance limits as follows:
    A)
    For lead: +30 percent of the actual amount in the performance
    evaluation sample when the actual amount is greater than or equal to
    0.005 mg/L (the PQL for lead is 0.005 mg/L);
    B)
    For copper: +10 percent of the actual amount in the performance
    evaluation sample when the actual amount is greater than or equal to
    0.050 mg/L (the PQL for copper is 0.050 mg/L);
    C)
    Achieve the method detection limits (MDLs) defined in Section
    611.350(a) according to the procedures in 35 Ill. Adm. Code 183
    and 40 CFR 136, Appendix B: "Definition and Procedure for the
    Determination of the Method Detection Limit--Revision 1.11"
    (1999); and
    D)
    Be currently certified by USEPA or the Agency to perform analyses
    to the specifications described in subsection (a)(2) of this
    Sectionbelow.
    b)
    The Agency shall, by a SEP issued pursuant to Section 611.110, allow a supplier to
    use previously collected monitoring data for the purposes of monitoring under this
    Subpart if the data were collected and analyzed in accordance with the
    requirements of this Subpart.
    c)
    Reporting lead and copper levels.
    1)
    All lead and copper levels greater than or equal to the lead and copper PQL
    (Pb > 0.005 mg/L and Cu > 0.050 mg/L) must be reported as measured.
    2)
    All lead and copper levels measured less than the PQL and greater than the
    MDL (0.005 mg/L > Pb > MDL and 0.050 mg/L > Cu > MDL) must be
    either reported as measured or as one-half the PQL set forth in subsection
    (a) of this Sectionabove (i.e., reported as 0.0025 mg/L for lead or 0.025
    mg/L for copper).
    3)
    All lead and copper levels below the lead and copper MDL (MDL > Pb)
    must be reported as zero.

    32
    BOARD NOTE: Derived from 40 CFR 141.89 ( 19941999), as amended at
    59 Fed. Reg. 62470 (December 5, 1994).
    (Source: Amended at 24 Ill. Reg. ________, effective _________)
    SUBPART K: GENERAL MONITORING AND ANALYTICAL REQUIREMENTS
    Section 611.490
    Certified Laboratories
    a)
    For the purpose of determining compliance with Subparts L through Q, samples
    will be considered only if they have been analyzed:
    1)
    By a laboratory certified pursuant to Section 4(o) of the Act; or,
    2)
    By a laboratory certified by U.S. EPAUSEPA; or,
    3)
    Measurements for alkalinity, calcium, conductivity, disinfectant residual,
    orthophosphate, silica, turbidity, free chlorine residual, temperature, and
    pH may be performed under the supervision of a certified operator (35 Ill.
    Adm. Code 603.103).
    b)
    Nothing in this Part shall be construed to preclude the Agency or any duly
    designated representative of the Agency from taking samples or from using the
    results from such samples to determine compliance by a supplier of water with the
    applicable requirements of this Part.
    BOARD NOTE: Derived from 40 CFR 141.28 (19941999).
    c)
    The CWS supplier shall have required analyses performed either at an Agency
    laboratory or a certified laboratory. The Agency may require that some or all of
    the required samples be submitted to its laboratories.
    BOARD NOTE: This is an additional State requirement.
    (Source: Amended at 24 Ill. Reg. _______, effective _________)
    SUBPART L: MICROBIOLOGICAL MONITORING AND ANALYTICAL
    REQUIREMENTS
    Section 611.526
    Analytical Methodology
    a)
    The standard sample volume required for total coliform analysis, regardless of
    analytical method used, is 100 mL.
    b)
    Suppliers need only determine the presence or absence of total coliforms,; a
    determination of total coliform density is not required.

    33
    c)
    Suppliers shall conduct total coliform analyses in accordance with one of the
    following analytical methods, incorporated by reference in Section 611.102 (the
    time from sample collection to initiation of analysis may not exceed 30 hours, and
    the supplier is encouraged but not required to hold samples below 10° C during
    transit):
    1)
    Multiple-TubeTotal Coliform Fermentation (MTF) Technique, as set forth
    in Standard Methods, 18th or 19th ed.: Methods 9221 A and B:
    A)
    Lactose broth, as commercially available, may be used in lieu of
    lauryl tryptose broth if the supplier conducts at least 25 parallel
    tests between this medium and lauryl tryptose broth using the water
    normally tested and this comparison demonstrates that the false-
    positive rate and false-negative rate for total coliforms, using
    lactose broth, is less than 10 percent;
    B)
    If inverted tubes are used to detect gas production, the media should
    cover these tubes at least one-half to two-thirds after the sample is
    added; and
    C)
    No requirement exists to run the completed phase on 10 percent of
    all total coliform-positive confirmed tubes.
    2)
    Total Coliform Membrane Filter (MF) Technique, as set forth in Standard
    Methods, 18th or 19th ed.: Methods 9222 A, B, and C.
    3)
    Presence-Absence (P-A) Coliform Test, as set forth in: Standard Methods,
    18th or 19th ed.: Method 9221 D:
    A)
    No requirement exists to run the completed phase on 10 percent of
    all total coliform-positive confirmed tubes; and
    B)
    Six-times formulation strength may be used if the medium is filter-
    sterilized rather than autoclaved.
    4)
    ONPG-MUG test: Standard Methods, 18th or 19th ed.: Method 9223.
    (The ONPG-MUG test is also known as the Aautoanalysis Ccolilert
    Ssystem.).
    5)
    Colisure Test (Autoanalysis Colilert System)from Millipore Corporation,
    incorporated by reference in Section 611.102. (The Colisure Test must be
    incubated for 28 hours before examining results. If an examination of the
    results at 28 hours is not convenient, then results may be examined at any
    time between 28 hours and 48 hoursmay be read after an incubation time of
    24 hours.)

    34
    BOARD NOTE: USEPA included the P-A Coliform and Colisure Tests for
    testing finished water under the coliform rule, but did not include them for
    the purposes of the surface water treatment rule, under Section 611.531, for
    which quantitation of total coliforms is necessary. For these reasons,
    USEPA included Standard Methods: Method 9221 C for the surface water
    treatment rule, but did not include it for the purposes of the total coliform
    rule, under this Section.
    6) E*Colite
    ®
    Test (Charm Sciences, Inc.).
    7) m-ColiBlue24
    ®
    Test (Hatch Company).
    d)
    This subsection corresponds with 40 CFR 141.21(f)(4), which USEPA has marked
    "reserved". This statement maintains structural consistency with the federal
    regulations.
    e)
    Suppliers shall conduct fecal coliform analysis in accordance with the following
    procedure:
    1)
    When the MTF Technique or P-A Coliform Test is used to test for total
    coliforms, shake the lactose-positive presumptive tube or P-A vigorously
    and transfer the growth with a sterile 3-mm loop or sterile applicator stick
    into brilliant green lactose bile broth and EC medium, defined below, to
    determine the presence of total and fecal coliforms, respectively.
    2)
    For approved methods that use a membrane filter, transfer the total
    coliform-positive culture by one of the following methods: remove the
    membrane containing the total coliform colonies from the substrate with a
    sterile forceps and carefully curl and insert the membrane into a tube of EC
    medium.; (Tthe laboratory may first remove a small portion of selected
    colonies for verification); swab the entire membrane filter surface with a
    sterile cotton swab and transfer the inoculum to EC medium (do not leave
    the cotton swab in the EC medium); or inoculate individual total coliform-
    positive colonies into EC medium. Gently shake the inoculated tubes of EC
    medium to insure adequate mixing and incubate in a waterbath at 44.5±0.2°
    C for 24±2 hours. Gas production of any amount in the inner fermentation
    tube of the EC medium indicates a positive fecal coliform test.
    3)
    EC medium is described in Standard Methods, 18th ed. and 19th ed.:
    Method 9221 E.
    4)
    Suppliers need only determine the presence or absence of fecal coliforms,;
    a determination of fecal coliform density is not required.
    f)
    Suppliers shall conduct analysis of E. coli in accordance with one of the following
    analytical methods, incorporated by reference in Section 611.102:

    35
    1)
    EC medium supplemented with 50 μg/L of MUG (final concentration). EC
    medium is as described in subsection (e) of this Section. MUG may be
    added to EC medium before autoclaving. EC medium supplemented with
    50 μg/L MUG is commercially available. At least 10 mL of EC medium
    supplemented with MUG must be used. The inner inverted fermentation
    tube may be omitted. The procedure for transferring a total coliform-
    positive culture to EC medium supplemented with MUG is as in subsection
    (e) of this Section for transferring a total coliform-positive culture to EC
    medium. Observe fluorescence with an ultraviolet light (366 nm) in the
    dark after incubating tube at 44.5±2° C for 24±2 hours; or
    2)
    Nutrient agar supplemented with 100 μg/L MUG (final concentration).
    Nutrient agar is described in Standard Methods, 18th ed. or 19th ed.:
    Method 9221 B, at pages 9-47 to 9-48. This test is used to determine if a
    total coliform-positive sample, as determined by the MF technique or any
    other method in which a membrane filter is used, contains E. coli. Transfer
    the membrane filter containing a total coliform colony or colonies to
    nutrient agar supplemented with 100 μg/L MUG (final concentration). After
    incubating the agar plate at 35° Celsius for 4 hours, observe the colony or
    colonies under ultraviolet light (366-nm) in the dark for fluorescence. If
    fluorescence is visible, E. coli are present.
    3)
    Minimal Medium ONPG-MUG (MMO-MUG) Test, as set forth in Section
    611.Appendix D. (The Autoanalysis ColiertColilert System is a MMO-
    MUG test.) If the MMO-MUG test is total coliform positive after a 24-hour
    incubation, test the medium for fluorescence with a 366-nm ultraviolet light
    (preferably with a 6-watt lamp) in the dark. If fluorescence is observed, the
    sample is E. coli-positive. If fluorescence is questionable (cannot be
    definitively read) after 24 hours incubation, incubate the culture for an
    additional four hours (but not to exceed 28 hours total), and again test the
    medium for fluorescence. The MMO-MUG test with hepes buffer is the
    only approved formulation for the detection of E. coli.
    4)
    The Colisure Test (Autoanalysis Colilert System), from Millipore
    Corporation, incorporated by reference in Section 611.102.
    5) The membrane filter method with MI agar.
    6) The E*Colite
    ®
    Test.
    7) The m-ColiBlue24
    ®
    Test.
    g)
    As an option to the method set forth in subsection (f)(3) of this Section, a supplier
    with a total coliform-positive, MUG-negative, MMO-MUG test may further analyze
    the culture for the presence of E. coli by transferring a 0.1 mL, 28-hour MMO-

    36
    MUG culture to EC medium + MUG with a pipet. The formulation and incubation
    conditions of the EC medium + MUG, and observation of the results, are described
    in subsection (f)(1) of this Section.
    h)
    This subsection corresponds with 40 CFR 141.21(f)(8), a central listing of all
    documents incorporated by reference into the federal microbiological analytical
    methods. The corresponding Illinois incorporations by reference are located at
    Section 611.102. This statement maintains structural parity with USEPA
    regulations.
    BOARD NOTE: Derived from 40 CFR 141.21(f) (19951999).
    (Source: Amended at 24 Ill. Reg. ________, effective _________)
    Section 611.531 Analytical Requirements
    The analytical method(s) specified in this Section must be used to demonstrate compliance with
    the requirements of only 611.Subpart B; they do not apply to analyses performed for the purposes
    of Sections 611.521 through 611.527 of this Subpart. Measurements for pH, temperature,
    turbidity, and RDCs must be conducted under the supervision of a certified operator.
    Measurements for total coliforms, fecal coliforms and HPC must be conducted by a laboratory
    certified by the Agency to do such analysis. The following procedures must be performed by the
    following methods, incorporated by reference in Section 611.102:
    a)
    A supplier shall:
    1)
    Conduct analysisanalyses of pH in accordance with one of the methods
    listed at Section 611.611; and
    2)
    Conduct analyses of total coliforms, fecal coliforms, heterotrophic bacteria,
    and turbidity in accordance with one of the following methods, and by using
    analytical test procedures contained in USEPA Technical Notes,
    incorporated by reference in Section 611.102:
    A)
    Total Coliforms:
    BOARD NOTE: The time from sample collection to initiation of
    analysis for source (raw) water samples required by Sections
    611.521 and 611.532 and 611.Subpart B only must not exceed 8
    hours. The supplier is encouraged but not required to hold samples
    below 10° C during transit.
    i)
    Total coliform fermentation technique: Standard Methods,
    18th ed. or 19th ed.: Method 9221 A, B, and C.

    37
    BOARD NOTE: Lactose broth, as commercially available,
    may be used in lieu of lauryl tryptose broth if the supplier
    conducts at least 25 parallel tests between this medium and
    lauryl tryptose broth using the water normally tested and this
    comparison demonstrates that the false-positive rate and
    false-negative rate for total coliforms, using lactose broth, is
    less than 10 percent. If inverted tubes are used to detect gas
    production, the media should cover these tubes at least one-
    half to two-thirds after the sample is added. No requirement
    exists to run the completed phase on 10 percent of all total
    coliform-positive confirmed tubes.
    ii)
    Total coliform membrane filter technique: Standard
    Methods, 18th ed. or 19th ed.: Method 9222 A, B, and C.
    iii)
    ONPG-MUG test (also known as the aAutoanalysis cColilert
    sSystem): Standard Methods, 18th ed. or 19th ed.: Method
    9223.
    BOARD NOTE: USEPA included the P-A Coliform and
    Colisure Tests for testing finished water under the coliform
    rule, under Section 611.526, but did not include them for the
    purposes of the surface water treatment rule, under this
    Section, for which quantitation of total coliforms is
    necessary. For these reasons, USEPA included Standard
    Methods: Method 9221 C for the surface water treatment
    rule, but did not include it for the purposes of the total
    coliform rule, under Section 611.526.
    B)
    Fecal Coliforms:
    BOARD NOTE: The time from sample collection to initiation of
    analysis for source (raw) water samples required by Sections
    611.521 and 611.532 and 611.Subpart B only must not exceed 8
    hours. The supplier is encouraged but not required to hold samples
    below 10° C during transit.
    i)
    Fecal coliform procedure: Standard Methods, 18th ed. or
    19th ed.: Method 9221 E.
    BOARD NOTE: A-1 broth may be held up to three months
    in a tightly closed screwcap tube at 4° C (39° F).
    ii)
    Fecal Coliform Membrane Filter Procedure: Standard
    Methods, 18th ed. or 19th ed.: Method 9222 D.

    38
    C)
    Heterotrophic bacteria: Pour plate method: Standard Methods,
    18th ed. or 19th ed.: Method 9215 B.
    BOARD NOTE: The time from sample collection to initiation of
    analysis must not exceed 8 hours. The supplier is encouraged but
    not required to hold samples below 10° C during transit.
    D)
    Turbidity:
    i)
    Nephelometric method: Standard Methods, 18th ed. or 19th
    ed.: Method 2130 B.
    ii)
    Nephelometric method: USEPA Environmental Inorganic
    Methods: Method 180.1
    ii)
    GLI Method 2.
    E)
    Temperature: Standard Methods, 18th ed. or 19th ed.: Method
    2550.
    b)
    A supplier shall measure residual disinfectant concentrations with one of the
    following analytical methods from Standard Methods, 18th ed. or 19
    th
    ed., and by
    using analytical test procedures contained in USEPA Technical Notes, incorporated
    by reference in Section 611.102:
    1)
    Free chlorine:
    A)
    Amperometric Titration: Method 4500-Cl D.
    B)
    DPD Ferrous Titrimetric: Method 4500-Cl F.
    C)
    DPD Colimetric: Method 4500-Cl G.
    D)
    Syringaldazine (FACTS): Method 4500-Cl H.
    2)
    Total chlorine:
    A)
    Amperometric Titration: Method 4500-Cl D.
    B)
    Amperometric Titration (low level measurement): Method 4500-Cl
    E.
    C)
    DPD Ferrous Titrimetric: Method 4500-Cl F.
    D)
    DPD Colimetric: Method 4500-Cl G.

    39
    E)
    Iodometric Electrode: Method 4500-Cl I.
    3)
    Chlorine dioxide:
    A)
    Amperometric Titration: Method 4500-ClO
    2
    C or E.
    B)
    DPD Method: Method 4500-ClO
    2
    D.
    4)
    Ozone: Indigo Method: Method 4500-O
    3
    B.
    5)
    Alternative test methods: The Agency may grant a SEP pursuant to Section
    611.110 that allows a supplier to use alternative chlorine test methods as
    follows:
    A)
    DPD colorimetric test kits: Residual disinfectant concentrations for
    free chlorine and combined chlorine may also be measured by using
    DPD colorimetric test kits.
    B)
    Continuous monitoring for free and total chlorine: Free and total
    chlorine residuals may be measured continuously by adapting a
    specified chlorine residual method for use with a continuous
    monitoring instrument, provided the chemistry, accuracy, and
    precision remain the same. Instruments used for continuous
    monitoring must be calibrated with a grab sample measurement at
    least every five days or as otherwise provided by the Agency.
    BOARD NOTE: Suppliers may use a five-tube test or a ten-tube
    test.
    BOARD NOTE: Derived from 40 CFR 141.74(a) (19951999).
    (Source: Amended at 24 Ill. Reg. _______, effective _________)
    SUBPART N: INORGANIC MONITORING AND ANALYTICAL REQUIREMENTS
    Section 611.611
    Inorganic Analysis
    Analytical methods are from documents incorporated by reference in Section 611.102. These are
    mostly referenced by a short name defined by Section 611.102(a). Other abbreviations are defined
    in Section 611.101.
    a)
    Analysis for the following contaminants must be conducted using the following
    methods or an alternative approved pursuant to Section 611.480. Criteria for
    analyzing arsenic, chromium, copper, lead, nickel, selenium, sodium, and thallium
    with digestion or directly without digestion, and other analytical procedures, are
    contained in USEPA Technical Notes, incorporated by reference in Section
    611.102. (This document also contains approved analytical test methods that

    40
    remain available for compliance monitoring until July 1, 1996. These methods will
    not be available for use after July 1, 1996.)
    1)
    Antimony:
    A)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    B)
    Atomic absorption, hydride technique: ASTM Method D3697-92.
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    D)
    Atomic absorption, furnace technique: Standard Methods, 18th ed.:
    Method 3113 B.
    2)
    Arsenic:
    A)
    Inductively-coupled Pplasma:
    i)
    USEPA Environmental Metals Methods: Method 200.7, or
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    D)
    Atomic Absorption, furnace technique:
    i)
    ASTM Method D2972-93 C, or
    ii)
    Standard Methods, 18th ed.: Method 3113 B.
    E)
    Atomic absorption, hydride technique:
    i)
    ASTM Method D2972-93 B, or
    ii)
    Standard Methods, 18th ed.: Method 3114 B.
    3)
    Asbestos: Transmission electron microscopy: USEPA Asbestos Methods-
    100.1 and USEPA Asbestos Methods-100.2.
    4)
    Barium:

    41
    A)
    Inductively-coupled plasma:
    i)
    USEPA Environmental Metals Methods: Method 200.7, or
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    C)
    Atomic absorption, direct aspiration technique: Standard Methods,
    18th ed.: Method 3111 D.
    D)
    Atomic absorption, furnace technique: Standard Methods, 18th ed.:
    Method 3113 B.
    5)
    Beryllium:
    A)
    Inductively-coupled plasma:
    i)
    USEPA Environmental Metals Methods: Method 200.7, or
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    D)
    Atomic absorption, furnace technique:
    i)
    ASTM Method D3645-93 B, or
    ii)
    Standard Methods, 18th ed.: Method 3113 B.
    6)
    Cadmium:
    A)
    Inductively-coupled plasma arc furnace: USEPA Environmental
    Metals Methods: Method 200.7.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.

    42
    D)
    Atomic absorption, furnace technique: Standard Methods, 18th ed.:
    Method 3113 B.
    7)
    Chromium:
    A)
    Inductively-coupled plasma arc furnace:
    i)
    USEPA Environmental Metals Methods: Method 200.7, or
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    D)
    Atomic absorption, furnace technique: Standard Methods, 18th ed.:
    Method 3113 B.
    8)
    Cyanide:
    A)
    Manual distillation (Standard Methods, 18th ed.: Method 4500-CN
    -
    C), followed by spectrophotometric, amenable:
    i)
    ASTM Method D2036-91 B, or
    ii)
    Standard Methods, 18th ed.: Method 4500-CN
    -
    G.
    B)
    Manual distillation (Standard Methods, 18th ed.: Method 4500-CN
    -
    C), followed by spectrophotometric, manual:
    i)
    ASTM Method D2036-91 A,
    ii)
    Standard Methods, 18th ed.: Method 4500-CN
    -
    E, or
    iii)
    USGS Methods: Method I-3300-85.
    C)
    Manual distillation (Standard Methods, 18th ed.: Method 4500-CN
    -
    C), followed by semiautomated spectrophotometric: USEPA
    Environmental Inorganic Methods: Method 335.4.
    D)
    Selective electrode: Standard Methods, 18th ed.: Method 4500-
    CN
    -
    F.

    43
    9)
    Fluoride:
    A)
    Ion Chromatography:
    i)
    USEPA Environmental Inorganic Methods: Method 300.0,
    ii)
    ASTM Method D4327-91, or
    iii)
    Standard Methods, 18th ed.: Method 4110 B.
    B)
    Manual distillation, colorimetric SPADNS: Standard Methods,
    18th ed.: Method 4500-F
    -
    B and D.
    C)
    Manual electrode:
    i)
    ASTM Method D1179-93B, or
    ii)
    Standard Methods, 18th ed.: Method 4500-F
    -
    C.
    D)
    Automated electrode: Technicon Methods: Method 380-75WE.
    E)
    Automated alizarin:
    i)
    Standard Methods, 18th ed.: Method 4500-F
    -
    E, or
    ii)
    Technicon Methods: Method 129-71W.
    10)
    Mercury:
    A)
    Manual cold vapor technique:
    i)
    USEPA Environmental Metals Methods: Method 245.1,
    ii)
    ASTM Method D3223-91, or
    iii)
    Standard Methods, 18th ed.: Method 3112 B.
    B)
    Automated cold vapor technique: USEPA Inorganic Methods:
    Method 245.2.
    C)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    11)
    Nickel:
    A)
    Inductively-coupled plasma:

    44
    i)
    USEPA Environmental Metals Methods: Method 200.7, or
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9;.
    D)
    Atomic absorption, direct aspiration technique: Standard Methods,
    18th ed.: Method 3111 B;.
    E)
    Atomic absorption, furnace technique: Standard Methods, 18th ed.:
    Method 3113 B;.
    12)
    Nitrate:
    A)
    Ion chromatography:
    i)
    USEPA Environmental Inorganic Methods: Method 300.0,
    ii)
    ASTM Method D4327-91,
    iii)
    Standard Methods, 18th ed.: Method 4110 B, or
    iv)
    Waters Test Method B-1011, available from Millipore
    Corporation.
    B)
    Automated cadmium reduction:
    i)
    USEPA Environmental Inorganic Methods: Method 353.2,
    ii)
    ASTM Method D3867-90 A, or
    iii)
    Standard Methods, 18th ed.: Method 4500-NO
    3
    -
    F.
    C)
    Ion selective electrode:
    i)
    Standard Methods, 18th ed.: Method 4500-NO
    3
    -
    D, or
    ii)
    Technical Bulletin 601.
    D)
    Manual cadmium reduction:

    45
    i)
    ASTM Method D3867-90 B, or
    ii)
    Standard Methods, 18th ed.: Method 4500-NO
    3
    -
    E.
    13)
    Nitrite:
    A)
    Ion chromatography:
    i)
    USEPA Environmental Inorganic Methods: Method 300.0,
    ii)
    ASTM Method D4327-91,
    iii)
    Standard Methods, 18th ed.: Method 4110 B, or
    iv)
    Waters Test Method Method B-1011, available from
    Millipore Corporation.
    B)
    Automated cadmium reduction:
    i)
    USEPA Environmental Inorganic Methods: Method 353.2,
    ii)
    ASTM Method D3867-90 A, or
    iii)
    Standard Methods, 18th ed.: Method 4500-NO
    3
    -
    F.
    C)
    Manual cadmium reduction:
    i)
    ASTM Method D3867-90 B, or
    ii)
    Standard Methods, 18th ed.: Method 4500-NO
    3
    -
    E.
    D)
    Spectrophotometric: Standard Methods, 18th ed.: Method 4500-
    NO
    2
    -
    B.
    14)
    Selenium:
    A)
    Atomic absorption, hydride:
    i)
    ASTM Method D3859-93 A, or
    ii)
    Standard Methods, 18th ed.: Method 3114 B.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.

    46
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    D)
    Atomic absorption, furnace technique:
    i)
    ASTM Method D3859-93 B, or
    ii)
    Standard Methods, 18th ed.: Method 3113 B.
    15)
    Thallium:
    A)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    B)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    16)
    Lead:
    A)
    Atomic absorption, furnace technique:
    i)
    ASTM Method D3559-90 DD3559-95 D, or
    ii)
    Standard Methods, 18th ed.: Method 3113 B.
    B)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    C)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    17)
    Copper:
    A)
    Atomic absorption, furnace technique:
    i)
    ASTM Method D1688-90 CD1688-95 C, or
    ii)
    Standard Methods, 18th ed.: Method 3113 B.
    B)
    Atomic absorption, direct aspiration:
    i)
    ASTM Method D1688-90 A, or
    ii)
    Standard Methods, 18th ed.: Method 3111 B.
    C)
    Inductively-coupled plasma:

    47
    i)
    USEPA Environmental Metals Methods: Method 200.7, or
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    D)
    Inductively-coupled plasma-mass spectrometry: USEPA
    Environmental Metals Methods: Method 200.8.
    E)
    Atomic absorption, platform furnace technique: USEPA
    Environmental Metals Methods: Method 200.9.
    18)
    pH:
    A)
    Electrometric:
    i)
    USEPA Inorganic Methods: Method 150.1,
    ii)
    ASTM Method D1293-84, or
    iii)
    Standard Methods, 18th ed.: Method 4500-H
    +
    B.
    B)
    USEPA Inorganic Methods: Method 150.2.
    19)
    Conductivity;: Conductance:
    A)
    ASTM Method D1125-91 AD1125-95 A, or
    B)
    Standard Methods, 18th ed.: Method 2510 B.
    20)
    Calcium:
    A)
    EDTA titrimetric:
    i)
    ASTM Method D511-93 A, or
    ii)
    Standard Methods, 18th ed.: Method 3500-Ca D.
    B)
    Atomic absorption, direct aspiration:
    i)
    ASTM Method D511-93 B, or
    ii)
    Standard Methods, 18th ed.: Method 3111 B.
    C)
    Inductively-coupled plasma:
    i)
    USEPA Environmental Metals Methods: Method 200.7, or

    48
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    21)
    Alkalinity:
    A)
    Titrimetric:
    i)
    ASTM Method D1067-92 B, or
    ii)
    Standard Methods, 18th ed.: Method 2320 B.
    B)
    Electrometric titration: USGS Methods: Method I-1030-85.
    22)
    Orthophosphate (unfiltered, without digestion or hydrolysis):
    A)
    Automated colorimetric, ascorbic acid:
    i)
    USEPA Environmental Inorganic Methods: Method 365.1,
    or
    ii)
    Standard Methods, 18th ed.: Method 4500-P F.
    B)
    Single reagent colorimetric, ascorbic acid:
    i)
    ASTM Method D515-88 A, or
    ii)
    Standard Methods, 18th ed.: Method 4500-P E.
    C)
    Colorimetric, phosphomolybdate: USGS Methods: Method I-1601-
    85.
    D)
    Colorimetric, phosphomolybdate, automated-segmented flow:
    USGS Methods: Method I-2601-90.
    E)
    Colorimetric, phosphomolybdate, automated discrete: USGS
    Methods: Method I-2598-85.
    F)
    Ion Chromatography:
    i)
    USEPA Environmental Inorganic Methods: Method 300.0,
    ii)
    ASTM Method D4327-91, or
    iii)
    Standard Methods, 18th ed.: Method 41104110 B.
    23)
    Silica:

    49
    A)
    Colorimetric, molybdate blue: USGS Methods: Method I-1700-85.
    B)
    Colorimetric, molybdate blue, automated-segmented flow: USGS
    Methods: Method I-2700-85.
    C)
    Colorimetric: ASTM Method D859-88D859-95.
    D)
    Molybdosilicate: Standard Methods, 18th ed.: Method 4500-Si D.
    E)
    Heteropoly blue: Standard Methods, 18th ed.: Method 4500-Si E.
    F)
    Automated method for molybdate-reactive silica: Standard
    Methods, 18th ed.: Method 4500-Si F.
    G)
    Inductively-coupled plasma:
    i)
    USEPA Environmental Metals Methods: Method 200.7, or
    ii)
    Standard Methods, 18th ed.: Method 3120 B.
    24)
    Temperature:; thermometric: Standard Methods, 18th ed.: Method 2550.
    25)
    Sodium:
    A)
    Inductively-coupled plasma: USEPA Environmental Metals
    Methods: Method 200.7.
    B)
    Atomic absorption, direct aspiration: Standard Methods, 18th ed.:
    Method 3111 B.
    b)
    Sample collection for antimony, asbestos, barium, beryllium, cadmium, chromium,
    cyanide, fluoride, mercury, nickel, nitrate, nitrite, selenium, and thallium pursuant
    to Sections 611.600 through 611.604 must be conducted using the following
    sample preservation, container, and maximum holding time procedures:
    1)
    Antimony:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    2)
    Asbestos:

    50
    A)
    Preservative: Cool to 4° C.
    B)
    Plastic or glass (hard or soft).
    C) Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 48 hours.
    3)
    Barium:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    4)
    Beryllium:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    5)
    Cadmium:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    6)
    Chromium:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    7)
    Cyanide:

    51
    A)
    Preservative: Cool to 4°C. Add sodium hydroxide to pH > 12. See
    the analytical methods for information on sample preservation.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 14 days.
    8)
    Fluoride:
    A)
    Preservative: None.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 1 month.
    9)
    Mercury:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 28 days.
    10)
    Nickel:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    11)
    Nitrate, chlorinated:
    A)
    Preservative: Cool to 4° C.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 2814 days.
    12)
    Nitrate, non-chlorinated:

    52
    A)
    Preservative: Concentrated sulfuric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 14 days.
    13)
    Nitrite:
    A)
    Preservative: Cool to 4° C.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 48 hours.
    14)
    Selenium:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    15)
    Thallium:
    A)
    Preservative: Concentrated nitric acid to pH less than 2.
    B)
    Plastic or glass (hard or soft).
    C)
    Holding time: Samples must be analyzed as soon after collection as
    possible, but in any event within 6 months.
    c)
    Analyses under this Subpart must be conducted by laboratories that received
    approval from USEPA or the Agency. Laboratories may conduct sample analyses
    for antimony, beryllium, cyanide, nickel, and thallium under provisional
    certification granted by the Agency until January 1, 1996. The Agency shall certify
    laboratories to conduct analyses for antimony, asbestos, barium, beryllium,
    cadmium, chromium, cyanide, fluoride, mercury, nickel, nitrate, nitrite, selenium,
    and thallium if the laboratory:
    1)
    Analyzes performance evaluation samples, provided by the Agency
    pursuant to 35 Ill. Adm. Code 183.125(c)35 Ill. Adm. Code Part 186, that
    include those substances at levels not in excess of levels expected in
    drinking water; and

    53
    2)
    Achieves quantitative results on the analyses within the following
    acceptance limits:
    A)
    Antimony: ± 30% at greater than or equal to 0.006 mg/L.
    B)
    Asbestos: 2 standard deviations based on study statistics.
    C)
    Barium: ± 15% at greater than or equal to 0.15 mg/L.
    D)
    Beryllium: ± 15% at greater than or equal to 0.001 mg/L.
    E)
    Cadmium: ± 20% at greater than or equal to 0.002 mg/L.
    F)
    Chromium: ± 15% at greater than or equal to 0.01 mg/L.
    G)
    Cyanide: ± 25% at greater than or equal to 0.1 mg/L.
    H)
    Fluoride: ± 10% at 1 to 10 mg/L.
    I)
    Mercury: ± 30% at greater than or equal to 0.0005 mg/L.
    J)
    Nickel: ± 15% at greater than or equal to 0.01 mg/L.
    K)
    Nitrate: ± 10% at greater than or equal to 0.4 mg/L.
    L)
    Nitrite: ± 15% at greater than or equal to 0.4 mg/L.
    M)
    Selenium: ± 20% at greater than or equal to 0.01 mg/L.
    N)
    Thallium: ± 30% at greater than or equal to 0.002 mg/L.
    BOARD NOTE: Subsection (e) is derived from the table to 40 CFR 141.23(k)(2) (1995) and the
    discussion at 57 Fed. Reg. 31809 (July 17, 1992). Section 611.609 is derived from 40 CFR
    141.23(k) (1995).
    (Source: Amended at 24 Ill. Reg. _______, effective ________)
    Section 611.612
    Monitoring Requirements for Old Inorganic MCLs
    a)
    Analyses for the purpose of determining compliance with the old inorganic MCLs
    of Section 611.300 are required as follows:
    1)
    Analyses for all CWSs utilizing surface water sources must be repeated at
    yearly intervals.

    54
    2)
    Analyses for all CWSs utilizing only groundwater sources must be repeated
    at three-year intervals.
    3)
    This subsection (a)(3) corresponds with 40 CFR 141.23(1)(3) (19941999),
    which requires monitoring for the repealed old MCL for nitrate at a
    frequency specified by the state. The Board has followed the U.S.
    EPAUSEPA lead and repealed that old MCL. This statement maintains
    structural consistency with U.S. EPAUSEPA rules.
    4)
    This subsection (a)(4) corresponds with 40 CFR 141.23(1)(4) (19941999),
    which authorizes the state to determine compliance and initiate enforcement
    action. This authority exists through the authorization of the Act, not through
    federal rules. This statement maintains structural consistency with U.S.
    EPAUSEPA rules.
    b)
    If the result of an analysis made under subsection (a) of this Section above
    indicates that the level of any contaminant listed in Section 611.300 exceeds the old
    MCL, the supplier shall report to the Agency within 7 days and initiate three
    additional analyses at the same sampling point within one month.
    c)
    When the average of four analyses made pursuant to subsection (b) of this Section
    above, rounded to the same number of significant figures as the old MCL for the
    substance in question, exceeds the old MCL, the supplier shall notify the Agency
    and give notice to the public pursuant to Subpart T of this Part. Monitoring after
    public notification must be at a frequency designated by the Agency by a SEP
    granted pursuant to Section 611.110 and must continue until the old MCL has not
    been exceeded in two successive samples or until a different monitoring schedule
    becomes effective as a condition to a variance, an adjusted standard, a site specific
    rule, an enforcement action, or another SEP granted pursuant to Section 611.110.
    d)
    This subsection corresponds with 40 CFR 141.23(o) (19941999), which pertains
    to monitoring for the repealed old MCL for nitrate. The Board has followed the
    U.S. EPAUSEPA action and repealed that old MCL. This statement maintains
    structural consistency with U.S. EPAUSEPA rules.
    e)
    This subsection corresponds with 40 CFR 141.23(p) (19941999), which pertains
    to the use of existing data up until a date long since expired. The Board did not
    adopt the original provision in R88-26. This statement maintains structural
    consistency with U.S. EPAUSEPA rules.
    f)
    Except for arsenic, for which analyses must be made in accordance with Section
    611.611, analyses conducted to determine compliance with the old MCLs of
    Section 611.300 must be made in accordance with the following methods,
    incorporated by reference in Section 611.102.

    55
    1)
    Fluoride: The methods specified in Section 611.611(c) shall apply for the
    purposes of this Section.
    2)
    Iron:
    A)
    Standard Methods, 18th ed.:
    i)
    Method 3111 B, or
    ii)
    Method 3113 B, or
    iii)
    Method 3120 B.
    B)
    USEPAU.S. EPA Environmental Metals Methods:
    i)
    Method 200.7, or
    ii)
    Method 200.9.
    3)
    Manganese:
    A)
    Standard Methods, 18th ed.:
    i)
    Method 3111 B,
    ii)
    Method 3113 B, or
    iii)
    Method 3120 B.
    B)
    USEPAU.S. EPA Environmental Metals Methods:
    i)
    Method 200.7,
    ii)
    Method 200.8, or
    iii)
    Method 200.9.
    4)
    Zinc:
    A)
    Standard Methods, 18th ed.:
    i)
    Method 3111 B, or
    ii)
    Method 3120 B.
    B)
    USEPAU.S. EPA Environmental Metals Methods:

    56
    i)
    Method 200.7, or
    ii)
    Method 200.8.
    BOARD NOTE: The provisions of subsections (a) through (f) of this Section above
    derive from 40 CFR 141.23(l) through (p) (19941999), as amended at 59 Fed.
    Reg. 62466 (Dec. 5, 1994). USEPAU.S. EPA removed and reserved 40 CFR
    141.23(q) (formerly 40 CFR 141.23(f)) at 59 Fed. Reg. 62466 (Dec. 5, 1994).
    Subsection (f)(2) of this Section above relates to a contaminant for which
    USEPAU.S. EPA specifies an MCL, but for which it repealed the analytical
    method. Subsections (f)(2) through (f)(4) of this Sectionabove relate exclusively to
    additional state requirements. The Board retained subsections (f)(1), (f)(3), and
    (f)(4) of this Section to set forth methods for the inorganic contaminants for which
    there is a state-only MCL. The methods specified are those set forth in 40 CFR
    143.4(b) (1999), as amended at 59 Fed. Reg. 62471 (Dec. 5, 1994), for
    secondary MCLs. The predecessor to subsections (a) through (e) of this
    Sectionabove were formerly codified as Section 611.601. The predecessor to
    subsection (f) of this Sectionabove was formerly codified as Section 611.606.
    (Source: Amended at 24 Ill. Reg. _______, effective _______)
    SUBPART O: ORGANIC MONITORING AND ANALYTICAL REQUIREMENTS
    Section 611.645
    Analytical Methods for Organic Chemical Contaminants
    Analysis for the Section 611.311(a) VOCs under Section 611.646; the Section 611.311(c) SOCs
    under Section 611.648; the Section 611.310 old organic MCLs under Section 611.641; and for
    THMs, TTHMs, and TTHM potential shall be conducted using the methods listed in this Section or
    by equivalent methods as approved by the Agency pursuant to Section 611.480. All methods are
    from USEPA Organic Methods unless otherwise indicated.
    Volatile Organic Chemical Contaminants (VOCs):
    Contaminant
    Analytical Methods
    Benzene
    502.2, 524.2
    Carbon tetrachloride
    502.2, 524.2, 551.1
    Chlorobenzene
    502.2, 524.2
    1,2-Dichlorobenzene
    502.2, 524.2
    1,4-Dichlorobenzene
    502.2, 524.2

    57
    1,2-Dichloroethane
    502.2, 524.2
    cis-Dichloroethylene
    502.2, 524.2
    trans-Dichloroethylene
    502.2, 524.2
    Dichloromethane
    502.2, 524.2
    1,2-Dichloropropane
    502.2, 524.2
    Ethylbenzene
    502.2, 524.2
    Styrene
    502.2, 524.2
    Tetrachloroethylene
    502.2, 524.2, 551.1
    1,1,1-Trichloroethane
    502.2, 524.2, 551.1
    Trichloroethylene
    502.2, 524.2, 551.1
    Toluene
    502.2, 524.2
    1,2,4-Trichlorobenzene
    502.2, 524.2
    1,1-Dichloroethylene
    502.2, 524.2
    1,1,2-Trichloroethane
    502.2, 524.2
    Vinyl chloride
    502.2, 524.2
    Xylenes (total)
    502.2, 524.2

    58
    Synthetic Organic Chemical Contaminants (SOCs):
    Contaminant
    Analytical Methods
    2,3,7,8-Tetrachlorodibenzodioxin (2,3,7,8-TCDD or dioxin)
    Dioxin and Furan Method
    1613
    2,4-D
    515.1, 515.2, 555, 515.1,
    515.3, D5317-93
    2,4,5-TP (Silvex)
    515.1, 515.2, 555, 515.1,
    515.3, D5317-93
    Alachlor
    505*, 507, 508.1, 525.2,
    551.1
    Atrazine
    505*, 507, 508.1, 525.2,
    551.1
    Benzo(a)pyrene
    525.2, 550, 550.1
    Carbofuran
    531.1, Standard Methods,
    18th ed.: Method 6610
    Chlordane
    505, 508, 508.1, 525.2
    Dalapon
    515.1, 552.1, 552.2, 515.3
    Di(2-ethylhexyl)adipate
    506, 525.2
    Di(2-ethylhexyl)phthalate
    506, 525.2
    Dibromochloropropane (DBCP)
    504.1, 551.1
    Dinoseb
    515.1, 515.2, 515.3, 555
    Diquat
    549.1
    Endothall
    548.1
    Endrin
    505, 508, 508.1, 525.2,
    551.1
    Ethylene Dibromide (EDB)
    504.1, 551.1

    59
    Glyphosate
    547, Standard Methods,
    18th ed.: Method 6651
    Heptachlor
    505, 508, 508.1, 525.2,
    551.1
    Heptachlor Epoxide
    505, 508, 508.1, 525.2,
    551.1
    Hexachlorobenzene
    505, 508, 508.1, 525.2,
    551.1
    Hexachlorocyclopentadiene
    505, 508, 508.1, 525.2,
    551.1
    Lindane
    505, 508, 508.1, 525.2,
    551.1
    Methoxychlor
    505, 508, 508.1, 525.2,
    551.1
    Oxamyl
    531.1, Standard Methods,
    18th ed.: Method 6610
    PCBs (measured for compliance purposes as
    decchlorobiphenyl)
    508A
    PCBs (qualitatively identified as Araclors)
    505, 508, 508.1, 525.2
    Pentachlorophenol
    515.1, 515.2, 525.2, 555,
    515.3, D5317-93
    Picloram
    515.1, 515.2, 555, 515.3,
    D5317-93
    Simazine
    505*, 507, 508.1, 525.2,
    551.2
    Toxaphene
    505, 508, 525.2, 508.1
    Total Trihalomethanes (TTHMs):
    Contaminant
    Analytical Methods

    60
    Total Trihalomethanes (TTHMs), Trihalomethanes (THMs),
    and Maximum Total Trihalomethane Potential
    502.2, 524.2, 551551.1
    State-Only MCLs (for which a method is not listed above):
    Contaminant
    Analytical Methods
    Aldrin
    505, 508, 508.1, 525.2
    DDT
    505, 508
    Dieldrin
    505, 508, 508.1, 525.2
    * denotes that, for the particular contaminant, a nitrogen-phosphorus detector should be substituted
    for the electron capture detector in method 505 (or another approved method should be used) to
    determine alachlor, atrazine, and simazine if lower detection limits are required.
    BOARD NOTE: Derived from 40 CFR 141.24(e) (19951999).
    (Source: Amended at 24 Ill. Reg. ______, effective _______)
    Section 611.646 Phase I, Phase II, and Phase V Volatile Organic Contaminants
    Monitoring of the Phase I, Phase II, and Phase V VOCs for the purpose of determining compliance
    with the MCL must be conducted as follows:
    a)
    Definitions. As used in this Section:
    "Detect" and "detection" means that the contaminant of interest is present at a level
    greater than or equal to the "detection limit".
    "Detection limit" means 0.0005 mg/L.
    BOARD NOTE: Derived from 40 CFR 141.24(f)(7), (f)(11), (f)(14)(i), and
    (f)(20) (19941999). This is a "trigger level" for Phase I, Phase II, and Phase V
    VOCs inasmuch as it prompts further action. The use of the term "detect" in this
    sectionSection is not intended to include any analytical capability of quantifying
    lower levels of any contaminant, or the "method detection limit". Note, however,
    that certain language at the end of federal paragraph (f)(20) is capable of meaning
    that the "method detection limit" is used to derive the "detection limit". The Board
    has chosen to disregard that language at the end of paragraph (f)(20) in favor of the
    more direct language of paragraphs (f)(7) and (f)(11).
    "Method detection limit", as used in subsections (q) and (t) of this Sectionbelow
    means the minimum concentration of a substance that can be measured and reported
    with 99 percent confidence that the analyte concentration is greater than zero and is
    determined from analysis of a sample in a given matrix containing the analyte.

    61
    BOARD NOTE: Derived from 40 CFR 136, Appendix B (19941999). The
    method detection limit is determined by the procedure set forth in 40 CFR 136,
    Appendix B. See subsection (t) of this Sectionbelow.
    b)
    Required sampling. Each supplier shall take a minimum of one sample at each
    sampling point at the times required in subsection (u) of this Sectionbelow.
    c)
    Sampling points.
    1)
    Sampling points for GWSs. Unless otherwise provided by SEP, a GWS
    supplier shall take at least one sample from each of the following points:
    each entry point that is representative of each well after treatment.
    2)
    Sampling points for SWSs and mixed systems. Unless otherwise provided
    by SEP, a SWS or mixed system supplier shall sample from each of the
    following points:
    A)
    Each entry point after treatment; or
    B)
    Points in the distribution system that are representative of each
    source.
    3)
    The supplier shall take each sample at the same sampling point unless the
    Agency has granted a SEP that designates another location as more
    representative of each source, or treatment plant, or within the distribution
    system.
    4)
    If a system draws water from more than one source, and the sources are
    combined before distribution, the supplier shall sample at an entry point
    during periods of normal operating conditions when water is representative
    of all sources being used.
    BOARD NOTE: Subsections (b) and (c) of this Sectionabove derived from
    40 CFR 141.24(f)(1) through (f)(3) (19941999).
    d)
    Each CWS and NTNCWS supplier shall take four consecutive quarterly samples
    for each of the Phase I VOCs, excluding vinyl chloride, and Phase II VOCs during
    each compliance period, beginning in the compliance period starting in the initial
    compliance period.
    e)
    Reduction to annual monitoring frequency. If the initial monitoring for the Phase I,
    Phase II, and Phase V VOCs as allowed in subsection (r)(1) of this Sectionbelow
    has been completed by December 31, 1992, and the supplier did not detect any of
    the Phase I VOCs, including vinyl chloride, Phase II, or Phase V VOCs, then the
    supplier shall take one sample annually beginning in the initial compliance period.

    62
    f)
    GWS reduction to triennial monitoring frequency. After a minimum of three years
    of annual sampling, GWS suppliers that have not previously detected any of the
    Phase I VOCs, including vinyl chloride, Phase II, or Phase V VOCs, shall take one
    sample during each three-year compliance period.
    g)
    A CWS or NTNCWS supplier that has completed the initial round of monitoring
    required by subsection (d) of this Sectionabove and which did not detect any of the
    Phase I VOCs, including vinyl chloride, Phase II, and Phase V VOCs, may apply to
    the Agency for a SEP pursuant to Section 611.110 that releases it from the
    requirements of subsection (e) or (f) of this Sectionabove. A supplier that serves
    fewer than 3300 service connections may apply to the Agency for a SEP pursuant to
    Section 611.110 that releases it from the requirements of subsection (d) of this
    Sectionabove as to 1,2,4-trichlorobenzene.
    BOARD NOTE: Derived from 40 CFR 141.24(f)(7) and (f)(10) (19941999), and
    the discussion at 57 Fed. Reg. 31825 (July 17, 1992). Provisions concerning the
    term of the waiver appear below in subsections (i) and (j) of this Sectionbelow.
    The definition of "detect", parenthetically added to the federal counterpart
    paragraph, is in subsection (a) of this Sectionabove.
    h)
    Vulnerability Assessment. The Agency shall consider the factors of Section
    611.110(e) in granting a SEP from the requirements of subsections (d), (e), or (f) of
    this Sectionabove sought pursuant to subsection (g) of this Sectionabove.
    i)
    A SEP issued to a GWS pursuant to subsection (g) of this Sectionabove is for a
    maximum of six years, except that a SEP as to the subsection (d) of this
    Sectionabove monitoring for 1,2,4-trichlorobenzene shall apply only to the initial
    round of monitoring. As a condition of a SEP, except as to a SEP from the initial
    round of subsection (d) of this Sectionabove monitoring for 1,2,4-trichlorobenzene,
    the supplier shall, within 30 months after the beginning of the period for which the
    waiver was issued, reconfirm its vulnerability assessment required by subsection
    (h) of this Sectionabove and submitted pursuant to subsection (g) of this
    Sectionabove, by taking one sample at each sampling point and reapplying for a
    SEP pursuant to subsection (g) of this Sectionabove. Based on this application, the
    Agency shall either:
    1)
    If it determines that the PWS meets the standard of Section 611.610(e),
    issue a SEP that reconfirms the prior SEP for the remaining three-year
    compliance period of the six-year maximum term; or,
    2)
    Issue a new SEP requiring the supplier to sample annually.
    BOARD NOTE: This provision does not apply to SWSs and mixed
    systems.

    63
    j)
    Special considerations for SEPs for SWSs and mixed systems.
    1)
    The Agency must determine that a SWS is not vulnerable before issuing a
    SEP pursuant to a SWS supplier. A SEP issued to a SWS or mixed system
    supplier pursuant to subsection (g) of this Sectionabove is for a maximum of
    one compliance period; and
    2)
    The Agency may require, as a condition to a SEP issued to a SWS or mixed
    supplier, that the supplier take such samples for Phase I, Phase II, and Phase
    V VOCs at such a frequency as the Agency determines are necessary, based
    on the vulnerability assessment.
    BOARD NOTE: There is a great degree of similarity between 40 CFR
    141.24(f)(7), the provision applicable to GWSs, and 40 CFR
    141.24(f)(10), the provision for SWSs. The Board has consolidated the
    common requirements of both paragraphs into subsection (g) of this
    Sectionabove. Subsection (j) of this Sectionabove represents the elements
    unique to SWSs and mixed systems, and subsection (i) of this Sectionabove
    relates to GWSs. Although 40 CFR 141.24(f)(7) and (f)(10) are silent as to
    mixed systems, the Board has included mixed systems with SWSs because
    this best follows the federal scheme for all other contaminants.
    k)
    If one of the Phase I VOCs, excluding vinyl chloride, Phase II, or Phase V VOCs is
    detected in any sample, then:
    1)
    The supplier shall monitor quarterly for that contaminant at each sampling
    point that resulted in a detection.
    2)
    Annual monitoring shall be required.
    A)
    The Agency shall grant a SEP pursuant to Section 611.110 that
    allows a supplier to reduce the monitoring frequency to annually at a
    sampling point if it determines that the sampling point is reliably
    and consistently below the MCL.
    B)
    A request for a SEP must include the following minimal
    information:
    i)
    For a GWS, two quarterly samples.
    ii)
    For a SWS or mixed system, four quarterly samples.
    C)
    In issuing a SEP, the Agency shall specify the level of the
    contaminant upon which the "reliably and consistently"
    determination was based. All SEPs that allow less frequent
    monitoring based on an Agency "reliably and consistently"

    64
    determination shall include a condition requiring the supplier to
    resume quarterly monitoring pursuant to subsection (k)(1) of this
    Sectionabove if it violates the MCL specified by Section 611.311.
    3)
    Suppliers that monitor annually shall monitor during the quarter(s) that
    previously yielded the highest analytical result.
    4)
    Suppliers that do not detect a contaminant at a sampling point in three
    consecutive annual samples may apply to the Agency for a SEP pursuant to
    Section 611.110 that allows it to discontinue monitoring for that
    contaminant at that point, as specified in subsection (g) of this
    Sectionabove.
    5)
    A GWS supplier that has detected one or more of the two-carbon
    contaminants listed in subsection (k)(5)(A) of this Sectionbelow shall
    monitor quarterly for vinyl chloride as described in subsection (k)(5)(B) of
    this Sectionbelow, subject to the limitation of subsection (k)(5)(C) of this
    Sectionbelow.
    A)
    Two-carbon contaminants (Phase I or II VOC):
    1,2-Dichloroethane (Phase I)
    1,1-Dichloroethylene (Phase I)
    cis-1,2-Dichloroethylene (Phase II)
    trans-1,2-Dichloroethylene (Phase II)
    Tetrachloroethylene (Phase II)
    1,1,1-Trichloroethylene (Phase I)
    Trichloroethylene (Phase I)
    B)
    The supplier shall sample quarterly for vinyl chloride at each
    sampling point at which it detected one or more of the two-carbon
    contaminants listed in subsection (k)(5)(A) of this Sectionabove.
    C)
    The Agency shall grant a SEP pursuant to Section 611.110 that
    allows the supplier to reduce the monitoring frequency for vinyl
    chloride at any sampling point to once in each three-year
    compliance period if it determines that the supplier has not detected
    vinyl chloride in the first sample required by subsection (k)(5)(B)
    of this Sectionabove.
    l)
    Quarterly monitoring following MCL violations.
    1)
    Suppliers that violate an MCL for one of the Phase I VOCs, including vinyl
    chloride, Phase II, or Phase V VOCs, as determined by subsection (o) of
    this Sectionbelow, shall monitor quarterly for that contaminant, at the

    65
    sampling point where the violation occurred, beginning the next quarter
    after the violation.
    2)
    Annual monitoring.
    A)
    The Agency shall grant a SEP pursuant to Section 611.110 that
    allows a supplier to reduce the monitoring frequency to annually if it
    determines that the sampling point is reliably and consistently below
    the MCL.
    B)
    A request for a SEP must include the following minimal
    information: four quarterly samples.
    C)
    In issuing a SEP, the Agency shall specify the level of the
    contaminant upon which the "reliably and consistently"
    determination was based. All SEPs that allow less frequent
    monitoring based on an Agency "reliably and consistently"
    determination shall include a condition requiring the supplier to
    resume quarterly monitoring pursuant to subsection (l)(1) of this
    Sectionabove if it violates the MCL specified by Section 611.311.
    D)
    The supplier shall monitor during the quarter(s) that previously
    yielded the highest analytical result.
    m)
    Confirmation samples. The Agency may issue a SEP pursuant to Section 610.110
    to require a supplier to use a confirmation sample for results that it finds dubious
    for whatever reason. The Agency must state its reasons for issuing the SEP if the
    SEP is Agency-initiated.
    1)
    If a supplier detects any of the Phase I, Phase II, or Phase V VOCs in a
    sample, the supplier shall take a confirmation sample as soon as possible,
    but no later than 14 days after the supplier receives notice of the detection.
    2)
    Averaging is as specified in subsection (o) of this Sectionbelow.
    3)
    The Agency shall delete the original or confirmation sample if it determines
    that a sampling error occurred, in which case the confirmation sample will
    replace the original or confirmation sample.
    n)
    This subsection corresponds with 40 CFR 141.24(f)(14), an optional USEPA
    provision relating to compositing of samples that USEPA does not require for state
    programs. This statement maintains structural consistency with USEPA rules.
    o)
    Compliance with the MCLs for the Phase I, Phase II, and Phase V VOCs must be
    determined based on the analytical results obtained at each sampling point.

    66
    1)
    For suppliers that conduct monitoring at a frequency greater than annual,
    compliance is determined by a running annual average of all samples taken
    at each sampling point.
    A)
    If the annual average of any sampling point is greater than the MCL,
    then the supplier is out of compliance.
    B)
    If the initial sample or a subsequent sample would cause the annual
    average to exceed the MCL, then the supplier is out of compliance
    immediately.
    C)
    Any samples below the detection limit shall be deemed as zero for
    purposes of determining the annual average.
    2)
    If monitoring is conducted annually, or less frequently, the supplier is out of
    compliance if the level of a contaminant at any sampling point is greater
    than the MCL. If a confirmation sample is taken, the determination of
    compliance is based on the average of two samples.
    3)
    When the portion of the distribution system that is out of compliance is
    separable from other parts of the distribution system and has no
    interconnections, the supplier may issue the public notice required by
    Subpart T of this Part only to persons served by that portion of the
    distribution system that is not in compliance.
    p)
    This provision corresponds with 40 CFR 141.24(f)(16) (19941999), which
    USEPA removed and reserved at 59 Fed. Reg. 62468 (Dec. 5, 1994). This
    statement maintains structural consistency with the federal regulations.
    q)
    Analysis under this Section must only be conducted by laboratories that have
    received certification by USEPA or the Agency according to the following
    conditions:
    1)
    To receive certification to conduct analyses for the Phase I VOCs,
    excluding vinyl chloride, Phase II VOCs, and Phase V VOCs, the laboratory
    must:
    A)
    Analyze performance evaluation samples that include these
    substances provided by the Agency pursuant to 35 Ill. Adm. Code
    183.125(c)35 Ill. Adm. Code Part 186;
    B)
    Achieve the quantitative acceptance limits under subsections
    (q)(1)(C) and (q)(1)(D) of this Sectionbelow for at least 80 percent
    of the Phase I VOCs, excluding vinyl chloride, Phase II VOCs,
    except vinyl chloride, or Phase V VOCs;

    67
    C)
    Achieve quantitative results on the analyses performed under
    subsection (q)(1)(A) of this Sectionabove that are within ± 20
    percent of the actual amount of the substances in the performance
    evaluation sample when the actual amount is greater than or equal to
    0.010 mg/L;
    D)
    Achieve quantitative results on the analyses performed under
    subsection (q)(1)(A) of this Sectionabove that are within ± 40
    percent of the actual amount of the substances in the performance
    evaluation sample when the actual amount is less than 0.010 mg/L;
    and
    E)
    Achieve a method detection limit of 0.0005 mg/L, according to the
    procedures in 40 CFR 136, appendix B, incorporated by reference
    in Section 611.102.
    2)
    To receive certification to conduct analyses for vinyl chloride the
    laboratory must:
    A)
    Analyze performance evaluation samples provided by the Agency
    pursuant to 35 Ill. Adm. Code 183.125(c)35 Ill. Adm. Code Part
    186;
    B)
    Achieve quantitative results on the analyses performed under
    subsection (q)(2)(A) of this Sectionabove that are within ± 40
    percent of the actual amount of vinyl chloride in the performance
    evaluation sample;
    C)
    Achieve a method detection limit of 0.0005 mg/L, according to the
    procedures in 40 CFR 136, appendix B, incorporated by reference
    in Section 611.102; and
    D)
    Obtain certification pursuant to subsection (q)(1) of this
    Sectionabove for Phase I VOCs, excluding vinyl chloride, Phase II
    VOCs, and Phase V VOCs.
    r)
    Use of existing data.
    1)
    The Agency shall allow the use of data collected after January 1, 1988 but
    prior to the effective date of this Section, pursuant to Agency sample
    request letters, if it determines that the data are generally consistent with the
    requirements of this Section.
    2)
    The Agency shall grant a SEP pursuant to Section 611.110 that allows a
    supplier to monitor annually beginning in the initial compliance period if it
    determines that the supplier did not detect any Phase I, Phase II, or Phase V

    68
    VOCs using existing data allowed pursuant to subsection (r)(1) of this
    Sectionabove.
    s)
    The Agency shall, by SEP, increase the number of sampling points or the frequency
    of monitoring if it determines that it is necessary to detect variations within the
    PWS.
    t)
    Each laboratory certified for the analysis of Phase I, Phase II, or Phase V VOCs
    pursuant to subsection (q)(1) or (q)(2) of this Sectionabove shall:
    1)
    Determine the method detection limit (MDL), as defined in 40 CFR 136,
    Appendix B, incorporated by reference in Section 611.102, at which it is
    capable of detecting the Phase I, Phase II, and Phase V VOCs; and,
    2)
    Achieve an MDL for each Phase I, Phase II, and Phase V VOC that is less
    than or equal to 0.0005 mg/L.
    u)
    Each supplier shall monitor, within each compliance period, at the time designated
    by the Agency by SEP pursuant to Section 611.110.
    BOARD NOTE: Derived from 40 CFR 141.24(f) (19941999).
    (Source: Amended at 24 Ill. Reg. _______, effective _________)
    IT IS SO ORDERED.
    I, Dorothy M. Gunn, Clerk of the Illinois Pollution Control Board, hereby certify that the above opinion and
    order was adopted on the 24th day of August 2000 by a vote of 7-0.
    Dorothy M. Gunn, Clerk
    Illinois Pollution Control Board

    Back to top